Equipment and Calibration History
Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 [21] 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70Ser No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
LI-7500A 75H-2180 | 2021-03-08 | Lab | Calibration | Lab | Last calibration: 2020-11-24 AGC=99.7 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=25.7C, Tdew=3.7C Zero Air at 400psi, AMP CA3923 474.19 at 550 Tdewspan=20.00 CO2 zero 0.9138 -> .9138 Offset: .18ppm CO2 span 1.0295 -> 1.0231 Offset: 3.95ppm H2O zero 0.8911 -> .8923 Offset: 7.9ppm H2O span 1.0245 -> 1.0243 Offset: 0ppm Notes: Larger H2O zero offset than I expected |
PS-30M 11400080 | 2021-02-26 | Sherman Wetland-Temp1 Solar Controller | Removed | Lab none | Removed from the decommissioned SW temporary location and brought back to the lab to be cleaned up. |
LI-7500A 75H-2669 | 2021-03-04 | Lab none | Installed | Bouldin Alfalfa Gas Analyzer CO2/H2O | Installed at Bouldin Alfalfa to replace sn 75H-2182 for calibration. |
LI-7500A 75H-2182 | 2021-03-04 | Bouldin Alfalfa Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Alfalfa and brought back to the lab for calibration. Replaced by sn 75H-2669. |
LI-7500A 75H-2514 | 2021-03-04 | Lab none | Installed | Bouldin Corn Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 1594 for calibration. |
LI-7500A 1594 | 2021-03-04 | Bouldin Corn Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Corn and brought back to the lab for calibration. Replaced by sn 75H-2514. |
S2-411 1054 | 2021-03-04 | East End Narrow Band NDVI Reflected B | Removed | Lab none | Removed from East End and brought back to the lab for testing. It had some NANs, but might've been influenced by the misbehaving NDVI_in sensor that was wired to the same control port. |
S2-411 1053 | 2021-03-04 | East End Narrow Band NDVI Incoming B | Removed | Lab none | Removed from East End because it had a lot of NANs. Brought back to the lab for testing. |
LI-7500A 0418 | 2021-03-04 | Lab none | Installed | Mayberry Gas Analyzer CO2/H2O | Installed at Mayberry to replace sn 0073 for calibration. |
LI-7500A 0073 | 2021-03-04 | Mayberry Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Mayberry and brought back to the lab for calibration. Replaced by sn 0418. |
LI-7500A 0035 | 2021-03-04 | Lab none | Installed | Sherman Wetland Gas Analyzer CO2/H2O | Installed at Sherman Wetland to replace sn 0065 for calibration. |
LI-7500A 0065 | 2021-03-04 | Sherman Wetland Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Sherman Wetland and brought back to the lab for calibration. Replaced by sn 0035. |
PAR-LITE 010159 | 2004-04-30 | Lab none | Installed | Tonzi PAR Sensor Incoming Cal | Installed on the radiation boom in the incoming position for inter-comparison with the existing Licor Quantum sensor. |
LI-7500A 0418 | 2021-03-03 | Lab | Calibration | Lab | Last calibration: 2020-11-24 AGC=97.7 SDM= DAC1= DAC2= Ambient P=99.3kPa, Tair=25.7C, Tdew=8.2C Zero Air at 450psi, AMP CA3923 474.19 at 550 Tdewspan=20.00 CO2 zero 0.8995 -> .9003 Offset: 3.01ppm CO2 span 1.0034 -> .9978 Offset: 3.49ppm H2O zero 0.8465 -> .8461 Offset: -2.0ppm H2O span 0.9847 -> .9906 Offset: -.14ppm Notes: I had to add more water the dewpt generator to get it to stay at the set dewpoint, otherwise it would just slowly decline forever. Calibrated earlier in the day, but I had forgotten to update the pressure coefficients in the config file for LI-7550 sn AIU-614. With the incorrect coefficients, offsets were (CO2 zero = 0.78, CO2 span = 0.23, H2O zero = -4.6, H2O span = 0.04). |
LI-7500A 0035 | 2021-03-03 | Lab | Calibration | Lab | Last calibration: 2020-11-23 AGC=101.3 SDM= DAC1= DAC2= Ambient P=99.4kPa, Tair=25.1C, Tdew=7.6C Zero Air at 450psi, AMP CA3923 474.19 at 550 Tdewspan=20.00 CO2 zero 0.9375 -> .9376 Offset: .7ppm CO2 span 1.019 -> 1.0151 Offset: 2.44ppm H2O zero .8983 -> .8983 Offset: 1.9ppm H2O span 1.035 -> 1.0319 Offset: .07ppm Notes: |
miniDOT-usb 931528 | 2021-02-26 | Mayberry | Removed | Lab none | Removed from MB wetland, where it was buried in the mud, and brought back to the lab to download its data |
miniDOT-usb 931528 | 2020-05-29 | Lab none | Installed | Mayberry none | Buried in the mud at 30 cm at the 0.5 footprint site at Mayberry |
LI-7500A 75H-2667 | 2021-03-02 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0418 for calibration. |
LI-7500A 0418 | 2021-03-02 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 75H-2667. |
WM 1590 153805 | 2021-03-02 | Cal Kit Anemometer | Removed | Lab none | Removed from TZ Tower top with the calibration kit |
CR1000 4190 | 2021-03-02 | Cal Kit Data Logger | Removed | Lab none | Removed from TZ Tower top with the calibration kit |
PAR-LITE 060831 | 2021-03-02 | Cal Kit PAR Sensor Reflected | Removed | Lab none | Removed from TZ Tower top with the calibration kit |
HMP155 S3410851 | 2021-03-02 | Cal Kit RH & Temp | Removed | Lab none | Removed from TZ Tower top with the calibration kit |
LI-7500A 75H-2176 | 2021-03-02 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0035 for calibration. |
LI-7500A 0035 | 2021-03-02 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 75H-2176. |
LI-7500A 75H-2150 | 2021-03-02 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 75H-2180 for calibration. |
LI-7500A 75H-2180 | 2021-03-02 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 75H-2150. |
LI-7500A 75H-2669 | 2021-03-01 | Lab | Calibration | Lab | Last calibration: 2020-08-27 AGC=101.8 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=29.0C, Tdew=1.3C Zero Air at 500psi, AMP CA04396 408.97 at 300 Tdewspan=20.0 CO2 zero 0.9184 -> 0.9190 Offset: 2.3ppm CO2 span 1.0118 -> 1.0059 Offset: 3.03ppm H2O zero 0.9012 -> 0.8984 Offset: -17.2ppm H2O span 1.0121 -> 1.0033 Offset: 0.19ppm Notes: Flow rated larger than usual - 2lpm instead of 2cfpm. Dew point generator problems before cal - dew point stuck at 34 instead of 20C. The dewpt gen was fixed an ran at 20C before starting H2O span but dewpt initially jumped to 25C but steadily declined to 20C. I double checked CO2 zero/span and H2O zero afterwards and they seemed fine. |
LI-7500A 75H-2667 | 2021-03-01 | Lab | Calibration | Lab | Last calibration: 2020-08-25 AGC=102.6 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=27.6C, Tdew=2.4C Zero Air at 500psi, AMP CA04396 408.97 at 250 Tdewspan=20.0 CO2 zero 0.942 -> 0.9423 Offset: 1.2ppm CO2 span 1.0172 -> 1.0092 Offset: 4.13ppm H2O zero 0.8403 -> 0.8385 Offset: -10.8ppm H2O span 1.0244 -> 1.0147 Offset: 0.2ppm Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm |
LI-7500A 75H-2514 | 2021-03-01 | Lab | Calibration | Lab | Last calibration: 2020-12-11 AGC=101.1 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=27.2C, Tdew=0.92C Zero Air at 500psi, AMP CA04396 408.97 at 250 Tdewspan=20.0 CO2 zero 0.9049 -> 0.9048 Offset: -0.73ppm CO2 span 1.0082 -> 0.9984 Offset: 5.03ppm H2O zero 0.8398 -> 0.8389 Offset: -6.7ppm H2O span 1.0093 -> 1.0061 Offset: 0.07ppm Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm |
LI-7500A 75H-2176 | 2021-03-01 | Lab | Calibration | Lab | Last calibration: 2020-12-02 AGC=102.5 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=22.9C, Tdew=4.6C Zero Air at 450psi, AMP CA04396 408.97 at 250 Tdewspan=20.0 CO2 zero 0.8673 -> 0.8671 Offset: -0.25ppm CO2 span 1.009 -> 1.000 Offset: 2.53ppm H2O zero 0.8531 -> 0.8524 Offset: -1.3ppm H2O span 0.9238 -> 0.9092 Offset: 0.37ppm Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm. |
LI-7500A 75H-2150 | 2021-03-01 | Lab | Calibration | Lab | Last calibration: 2020-08-20 AGC=100.8 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=27.4C, Tdew=2.08C Zero Air at 500psi, AMP CA04396 408.97 at 250 Tdewspan=20.0 CO2 zero 0.9019 -> 0.9019 Offset: -0.3ppm CO2 span 1.0014 -> 0.9975 Offset: 1.93ppm H2O zero 0.8665 -> 0.8675 Offset: 6.8ppm H2O span 1.011 -> 1.0040 Offset: 0.15ppm Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm. |
NR01 2420 | 2021-02-26 | Lab none | Installed | Sherman Wetland Net Radiometer 4-way Cal A | Installed on the Sherman Wetland tower for inter-comparison after repair in the lab. It is clamped near the end of the square tube of the radiation boom pointing south. |
CNR4 121038 | 2021-02-26 | Factory none | Installed | Sherman Wetland Net Radiometer 4-way Cal B | Installed on the Sherman Wetland tower for inter-comparison. This is the AMP standard. It is clamped near the end of the square tube of the radiation boom pointing south. |
CR1000 23865 | 2021-02-26 | Lab none | Installed | Sherman Wetland Data Logger Cal | Installed on the Sherman Wetland tower to collect data from two 4-way net radiometers for inter-comparison |
BB-Fab BB-SW_Temp1 | 2021-02-26 | Sherman Wetland-Temp1 Battery Bank | Disposed | Disposed none | Removed from the decommissioned temporary location. The batteries were brought back to the lab and will be used at Hill Slough but this battery bank has ceased to exist as an equipment entry |
KD140SX-UFBS 11ZV0M0829 | 2021-02-26 | Sherman Wetland-Temp1 Solar Panel A | Moved | Hill Slough Solar Panel A | Removed from the decommissioned temporary location for use at Hill Slough |
KD140SX-UFBS 11ZV0M0836 | 2021-02-26 | Sherman Wetland-Temp1 Solar Panel B | Moved | Hill Slough Solar Panel B | Removed from the decommissioned temporary location for use at Hill Slough |
KD140SX-UFBS 11ZV0M0855 | 2021-02-26 | Sherman Wetland-Temp1 Solar Panel C | Moved | Hill Slough Solar Panel C | Removed from the decommissioned temporary location for use at Hill Slough |
PAR-LITE 050597 | 2021-02-26 | Sherman Wetland-Temp1 PAR Sensor Incoming | Removed | Lab none | Removed from the decommissioned temporary location and brought back to the lab to be cleaned up. |
LI-610 DPG478B | 2021-03-01 | Lab none | Maintenance | Lab none | The dew point was stuck at about 34C - not dropping to the set point of 20C. I found a crack in the "from cooler" tubing. I replaced this piece of tubing but this didn't solve the problem. There was an air bubble in this piece of tubing and the machine was making an unusual bubbling sound. Refilling the reservoir with DI water immediately change the sound, got rid of the air bubble in the tubing and the dew point began to drop towards the set point. |
Mark II C003800 | 2021-02-25 | Lab none | Disposed | Disposed none | Returned to Arable |
Mark II C003789 | 2021-02-25 | Lab none | Disposed | Disposed none | Returned to Arable |
Mark II C002469 | 2020-04-23 | Lab none | Disposed | Disposed none | Returned to Arable |
Mark II C002445 | 2020-04-23 | Lab none | Disposed | Disposed none | Returned to Arable |
Mark II C002714 | 2020-04-23 | Lab none | Disposed | Disposed none | Returned to Arable |
Mark II C006322 | 2021-02-04 | Lab none | Installed | Bouldin Alfalfa Weather Station PL_N-B | Installed on the northern Planet Labs post to replace one of the two units eaten by the sheep |
Mark II C006282 | 2021-02-04 | Lab none | Installed | Bouldin Alfalfa Weather Station PL_N-A | Installed on the northern Planet Labs post to replace one of the two units eaten by the sheep |
Mark I A001311 | 2021-02-25 | Lab none | Disposed | Disposed none | Returned to Arable due to end of life. The was a cal val unit and was replaced by a new Mark II |
Mark I A001359 | 2021-02-25 | Lab none | Disposed | Disposed none | Returned to Arable due to end of life. The was a cal val unit and was replaced by a new Mark II |
Mark I A001353 | 2021-02-25 | Lab none | Disposed | Disposed none | Returned to Arable due to end of life. The was a cal val unit and was replaced by a new Mark II |
CR1000 23865 | 2021-02-25 | Lab none | Maintenance | Lab none | The internal lithium battery was nearly dead 1.6V. I couldn't find a new one but swapped it for one reading 2.6V instead. Be sure the check the clock when using this logger. |
CNR4 121038 | 2019-11-17 | Factory none | Calibration | Factory none | Initial calibration by Kipp and Zonen. LW on 2019-11-17: LW_in = 6.83 uV/(W/m2), LW_out = 5.48 uV/(W/m2). SW on 2019-11-15: SW_in = 14.56 uV/(W/m2), SW_out = 14.17 uV/(W/m2) |
NR01 2420 | 2021-02-18 | Lab none | Repair | Lab none | Over the last week or so I disassembled and cleaned this sensor. I don't see any water damage but I can't get the LW junction board loose as it seems glued to the case. I totally replaced the four wires running between the LW and SW sides. These wires were broken after having been previously broken and spliced. I soldered new wires directly to the junction boards. I collected about two days of data on the window sill with the sensor partially disassembled so that all four sensors could point up together. The relationship between the paired sensors was very tight (r-squared >0.99) - excluding a portion of the LW data that might have been affected by solar heating. LW had a sloped of 1.01 but SW had a slope of 1.20 - new calibration coefficients are probably needed. Today I completely reassembled the sensor and put it out the window on a boom to collect some data to check the assembly. |
PAR-LITE 990017 | 1999-02-09 | Factory none | Calibration | Factory none | Initial factory calibration: 5.16uV/(umol/m2/s) |
PAR-LITE 990019 | 1999-02-09 | Factory none | Calibration | Factory none | Initial factory calibration; 5.26uV/(umol/m2/s) |
PRB-K Type #8 000000402FF3 | 2021-02-10 | Tonzi Soil Moisture Probe F -0.50m | Removed | Lab none | Removed from Tonzi floor due to fixed readings for a long while. This sensor was out of the ground at the edge of the trees near the path to the tower. It was buried vertically with the top of the sensor 31cm below the surface - the time were 44-50cm below the surface. |
PAR-LITE 990031 | 2021-02-10 | Tonzi PAR Sensor Incoming Cal | Removed | Lab none | Removed from Tonzi Tower after three weeks of inter comparison with the AMP standard |
PAR-LITE 990025 | 2021-02-10 | Cal Kit PAR Sensor Incoming | Removed | Lab none | Removed from TZ Tower after three weeks of inter comparison with the AMP standard |
WM 1590 153805 | 2021-02-10 | Lab none | Installed | Cal Kit Anemometer | Installed on the TZ Tower for inter-comparison. This sensor is the one that got wet when the BC tower fell over. It's mounted on the post just north of the main eddy sensors, at the same height, and oriented to true north. |
eosFD FD20200001 | 2021-02-10 | Lab none | Installed | Vaira Forced Diffusion North | Installed near the old north soil co2 profile. This unit has a homemade solar shield to try to keep it cooler. Could use a soil collar. |
eosFD FD20200001 | 2021-02-04 | Factory none | Returned | Lab none | Returned from Eosense where its pump relay board was replaced under warranty repair. |
GP-001 GP20140026 | 2021-02-04 | West Pond CO2 Probe B | Removed | Lab none | Removed from WP and brought back to the lab b/c this sensor has been reading noise since late Nov |
LI-7500A 1594 | 2021-02-01 | Lab | Calibration | Lab | Last calibration: 2020-08-20 AGC=99.3 SDM=0 DAC1=- DAC2=- Ambient P=100.3kPa, Tair=28.4C, Tdew=11.0C Zero Air at 0psi, AMP CA04396 408.97 at 0 Tdewspan=20.0 CO2 zero 0.8651 -> 0.8646 Offset: -2.1ppm CO2 span 1.02124 -> 1.0131 Offset: -0.32ppm H2O zero 0.8838 -> 0.8890 Offset: 29.85ppm H2O span 1.015 -> 1.000 Offset: 0.35ppm Notes: The usual 7550 box in the lab is running the newest firmware v8.8 and is complaining that the v8.5 config file has bad number formats for GPS locations. I used 7550 sn AIU-1866 for calibration instead. This box and head will go to replace equipment on the BC tower that fell in the water. |
NR01 2420 | 2021-01-30 | Lab none | Damaged | Lab none | This sensor was removed from a submerged position on the fallen BC tower. The shortwave side had been torn loose breaking the wires similar to last time it was damaged. After returning to the lab on the 29th I pulled it apart and tried to air it out. On the 30th it was still damp inside and I ran dry air through the holes under the circuit boards for two days. |
WM 1590 153805 | 2021-02-01 | Lab none | Tested | Lab none | After being removed from a partially submerged position on the fallen BC tower, I tested this sensor in the lab. There was no evidence that water had leaked inside. I connected it to a Licor system and used fans to blow air through in on the floor. The response sign and magnitude looked about right. I plan to run it on the Tonzi tower for a couple weeks to check its performance. |
LI-7500A 75H-2176 | 2021-02-01 | Lab none | Tested | Lab none | On the 29th after being retrieved from a partially submerged position on the fallen BC tower, I removed the desiccant bottles and saw no evidence that water had gotten inside. I flowed dry air into the desiccant bottle openings over night. On the 1st, after consultation with Licor, I powered up this head on the calibration rig in the lab. It seemed to operate fine. I flooded the base with zero air before sealing it back up but it needs to run for a while to equilibrate and be calibrated. |
LI-7700 TG1-0155 | 2021-02-04 | Lab none | Shipped | Factory none | Shipped to Licor and RMA 00225226 for evaluation after being submerged when the BC tower fell in the storm last week. |
LI-7700 TG1-0155 | 2021-02-01 | Lab none | Tested | Lab none | On the 29th I removed this sensor from being submerged after the BC tower fell just after midnight on the morning of the 27th. On the evening of the 29th, about 1/2 cup of water came out of the bottom when I removed the base. I removed the desiccant bottle in the head and saw no evidence that water had leaked in there. I left dry air flowing inside the base over night. On the 30th I removed the bottom case and only saw minor water damage on the pressure sensor. On the 1st, after consultation with Licor, I powered up the analyzer and got green power and status lights but no communications on the Ethernet. It will need to be returned to Licor for further evaluation. |
SRS-Pi PRI 897103002 | 2021-02-03 | Lab none | Installed | Bouldin Corn Narrow Band PRI Reflected | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
SRS-Ni NDVI 2046503357 | 2021-02-03 | Lab none | Installed | Bouldin Corn Narrow Band NDVI Reflected | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
SRS-Ni NDVI 882103020 | 2021-02-03 | Lab none | Installed | Bouldin Corn Narrow Band NDVI Incoming | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
CNR1 990181 | 2021-02-03 | Lab none | Installed | Bouldin Corn Net Radiometer 4-way | Installed on the BC tower to replace the Hukseflux NR01 that broke when the tower fell and still needs repair and cross comparison. |
Mark II C006123 | 2021-02-03 | Lab none | Installed | Bouldin Corn Weather Station B | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
Mark II C003251 | 2021-02-03 | Lab none | Installed | Bouldin Corn Weather Station C | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
PA-II-SD 68c63acc07a9 | 2021-02-03 | Lab none | Installed | Bouldin Corn Air Quality Sensor | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
SD130BN 0030F4D21491 | 2021-02-03 | Lab none | Installed | Bouldin Corn Camera | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
PQS 1 140441 | 2021-02-03 | Lab none | Installed | Bouldin Corn PAR Sensor Reflected | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
PQS 1 140453 | 2021-02-03 | Lab none | Installed | Bouldin Corn PAR Sensor Incoming | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
SRS-Pi PRI 913003004 | 2021-02-03 | Lab none | Installed | Bouldin Corn Narrow Band PRI Incoming | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
LI-7500A 1594 | 2021-02-03 | Lab none | Installed | Bouldin Corn Gas Analyzer CO2/H2O | Installed on the BC tower to replace sensors damaged when the tower fell last week |
LI-7550 AIU-1866 | 2021-02-03 | Lab none | Installed | Bouldin Corn Interface LI-7550 | Installed on the BC tower to replace sensors damaged when the tower fell last week |
LI-7700 TG1-0224 | 2021-02-03 | Lab none | Installed | Bouldin Corn Gas Analyzer CH4 | Installed on the BC tower to replace sensors damaged when the tower fell last week |
WM 1590 153803 | 2021-02-03 | Lab none | Installed | Bouldin Corn Anemometer | Installed on the BC tower to replace sensors damaged when the tower fell last week |
WL16U 57278 | 2018-05-08 | Lab none | Repair | Lab none | I stripped off the old rotting insulation and then slid on a long continuous piece of heat shrink. I shrunk the whole length of the cable. I disassembled the data logging enclosure and kept only the data logger board and USB connector in the end cap. I cut off the twin 9V battery plugs and will power the sensor from solar connected rechargeable battery. |
WL16U 57278 | 2018-03-27 | Tonzi Water Level Floor | Removed | Lab none | Removed from the Floor Well. The cable insulation is rotting and needs to be replaced. Replaced by the new METER water level sensor a couple of weeks ago. |
CMTD-As4s0 3317052234210 | 2018-05-11 | Lab none | Installed | Tonzi none | Installed at the Road Well to power Global Water sensor. |
WL16U 57278 | 2018-05-11 | Lab none | Installed | Tonzi Water Level Road | Installed at the Road Well to measure water level at 5min intervals storing 30min averages |
Mark II C006123 | 2021-01-29 | Bouldin Corn Weather Station B | Removed | Lab none | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Maybe some rain in the Tair bulge |
Mark II C003251 | 2021-01-29 | Bouldin Corn Weather Station C | Removed | Lab none | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Maybe some rain in the Tair bulge |
PA-II-SD 84f3ebd5ffb1 | 2021-01-29 | Bouldin Corn Air Quality Sensor | Removed | Lab none | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Got rained on and the power connector is corroded |
SD130BN 0030F4D21491 | 2021-01-29 | Bouldin Corn Camera | Removed | Lab none | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
SRS-Pi PRI 897103002 | 2021-01-29 | Bouldin Corn Narrow Band PRI Reflected | Removed | Lab none | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
SRS-Pi PRI 913003004 | 2021-01-29 | Bouldin Corn Narrow Band PRI Incoming | Removed | Lab none | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. |
SRS-Ni NDVI 2046503357 | 2021-01-29 | Bouldin Corn Narrow Band NDVI Reflected | Removed | Lab none | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
SRS-Ni NDVI 882103020 | 2021-01-29 | Bouldin Corn Narrow Band NDVI Incoming | Removed | Lab none | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
PQS 1 140441 | 2021-01-29 | Bouldin Corn PAR Sensor Reflected | Removed | Lab none | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
PQS 1 140453 | 2021-01-29 | Bouldin Corn PAR Sensor Incoming | Removed | Lab none | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
NR01 2420 | 2021-01-29 | Bouldin Corn Net Radiometer 4-way | Removed | Lab none | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged and broken in two pieces |
LI-7550 AIU-0614 | 2021-01-29 | Bouldin Corn Interface LI-7550 | Removed | Lab none | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. |