LI-7500A 75H-2180 2021-03-08 Lab Calibration Lab Last calibration: 2020-11-24
AGC=99.7 SDM= DAC1= DAC2=
Ambient P=100.6kPa, Tair=25.7C, Tdew=3.7C
Zero Air at 400psi, AMP CA3923 474.19 at 550 Tdewspan=20.00
CO2 zero 0.9138 -> .9138 Offset: .18ppm
CO2 span 1.0295 -> 1.0231 Offset: 3.95ppm
H2O zero 0.8911 -> .8923 Offset: 7.9ppm
H2O span 1.0245 -> 1.0243 Offset: 0ppm
Notes: Larger H2O zero offset than I expected
PS-30M 11400080 2021-02-26 Sherman Wetland-Temp1 Solar Controller Removed Lab none Removed from the decommissioned SW temporary location and brought back to the lab to be cleaned up.
LI-7500A 75H-2669 2021-03-04 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed at Bouldin Alfalfa to replace sn 75H-2182 for calibration.
LI-7500A 75H-2182 2021-03-04 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Alfalfa and brought back to the lab for calibration. Replaced by sn 75H-2669.
LI-7500A 75H-2514 2021-03-04 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at Bouldin Corn to replace sn 1594 for calibration.
LI-7500A 1594 2021-03-04 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn and brought back to the lab for calibration. Replaced by sn 75H-2514.
S2-411 1054 2021-03-04 East End Narrow Band NDVI Reflected B Removed Lab none Removed from East End and brought back to the lab for testing. It had some NANs, but might've been influenced by the misbehaving NDVI_in sensor that was wired to the same control port.
S2-411 1053 2021-03-04 East End Narrow Band NDVI Incoming B Removed Lab none Removed from East End because it had a lot of NANs. Brought back to the lab for testing.
LI-7500A 0418 2021-03-04 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at Mayberry to replace sn 0073 for calibration.
LI-7500A 0073 2021-03-04 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry and brought back to the lab for calibration. Replaced by sn 0418.
LI-7500A 0035 2021-03-04 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed at Sherman Wetland to replace sn 0065 for calibration.
LI-7500A 0065 2021-03-04 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed from Sherman Wetland and brought back to the lab for calibration. Replaced by sn 0035.
PAR-LITE 010159 2004-04-30 Lab none Installed Tonzi PAR Sensor Incoming Cal Installed on the radiation boom in the incoming position for inter-comparison with the existing Licor Quantum sensor.
LI-7500A 0418 2021-03-03 Lab Calibration Lab Last calibration: 2020-11-24
AGC=97.7 SDM= DAC1= DAC2=
Ambient P=99.3kPa, Tair=25.7C, Tdew=8.2C
Zero Air at 450psi, AMP CA3923 474.19 at 550 Tdewspan=20.00
CO2 zero 0.8995 -> .9003 Offset: 3.01ppm
CO2 span 1.0034 -> .9978 Offset: 3.49ppm
H2O zero 0.8465 -> .8461 Offset: -2.0ppm
H2O span 0.9847 -> .9906 Offset: -.14ppm
Notes: I had to add more water the dewpt generator to get it to stay at the set dewpoint, otherwise it would just slowly decline forever. Calibrated earlier in the day, but I had forgotten to update the pressure coefficients in the config file for LI-7550 sn AIU-614. With the incorrect coefficients, offsets were (CO2 zero = 0.78, CO2 span = 0.23, H2O zero = -4.6, H2O span = 0.04).
LI-7500A 0035 2021-03-03 Lab Calibration Lab Last calibration: 2020-11-23
AGC=101.3 SDM= DAC1= DAC2=
Ambient P=99.4kPa, Tair=25.1C, Tdew=7.6C
Zero Air at 450psi, AMP CA3923 474.19 at 550 Tdewspan=20.00
CO2 zero 0.9375 -> .9376 Offset: .7ppm
CO2 span 1.019 -> 1.0151 Offset: 2.44ppm
H2O zero .8983 -> .8983 Offset: 1.9ppm
H2O span 1.035 -> 1.0319 Offset: .07ppm
Notes:
miniDOT-usb 931528 2021-02-26 Mayberry Removed Lab none Removed from MB wetland, where it was buried in the mud, and brought back to the lab to download its data
miniDOT-usb 931528 2020-05-29 Lab none Installed Mayberry none Buried in the mud at 30 cm at the 0.5 footprint site at Mayberry
LI-7500A 75H-2667 2021-03-02 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0418 for calibration.
LI-7500A 0418 2021-03-02 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 75H-2667.
WM 1590 153805 2021-03-02 Cal Kit Anemometer Removed Lab none Removed from TZ Tower top with the calibration kit
CR1000 4190 2021-03-02 Cal Kit Data Logger Removed Lab none Removed from TZ Tower top with the calibration kit
PAR-LITE 060831 2021-03-02 Cal Kit PAR Sensor Reflected Removed Lab none Removed from TZ Tower top with the calibration kit
HMP155 S3410851 2021-03-02 Cal Kit RH & Temp Removed Lab none Removed from TZ Tower top with the calibration kit
LI-7500A 75H-2176 2021-03-02 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed at Tonzi-Floor to replace sn 0035 for calibration.
LI-7500A 0035 2021-03-02 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Floor for calibration. Replaced by sn 75H-2176.
LI-7500A 75H-2150 2021-03-02 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed at Tonzi-Tower to replace sn 75H-2180 for calibration.
LI-7500A 75H-2180 2021-03-02 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Tower for calibration. Replaced by sn 75H-2150.
LI-7500A 75H-2669 2021-03-01 Lab Calibration Lab Last calibration: 2020-08-27
AGC=101.8 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=29.0C, Tdew=1.3C
Zero Air at 500psi, AMP CA04396 408.97 at 300 Tdewspan=20.0
CO2 zero 0.9184 -> 0.9190 Offset: 2.3ppm
CO2 span 1.0118 -> 1.0059 Offset: 3.03ppm
H2O zero 0.9012 -> 0.8984 Offset: -17.2ppm
H2O span 1.0121 -> 1.0033 Offset: 0.19ppm
Notes: Flow rated larger than usual - 2lpm instead of 2cfpm. Dew point generator problems before cal - dew point stuck at 34 instead of 20C. The dewpt gen was fixed an ran at 20C before starting H2O span but dewpt initially jumped to 25C but steadily declined to 20C. I double checked CO2 zero/span and H2O zero afterwards and they seemed fine.
LI-7500A 75H-2667 2021-03-01 Lab Calibration Lab Last calibration: 2020-08-25
AGC=102.6 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=27.6C, Tdew=2.4C
Zero Air at 500psi, AMP CA04396 408.97 at 250 Tdewspan=20.0
CO2 zero 0.942 -> 0.9423 Offset: 1.2ppm
CO2 span 1.0172 -> 1.0092 Offset: 4.13ppm
H2O zero 0.8403 -> 0.8385 Offset: -10.8ppm
H2O span 1.0244 -> 1.0147 Offset: 0.2ppm
Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm
LI-7500A 75H-2514 2021-03-01 Lab Calibration Lab Last calibration: 2020-12-11
AGC=101.1 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=27.2C, Tdew=0.92C
Zero Air at 500psi, AMP CA04396 408.97 at 250 Tdewspan=20.0
CO2 zero 0.9049 -> 0.9048 Offset: -0.73ppm
CO2 span 1.0082 -> 0.9984 Offset: 5.03ppm
H2O zero 0.8398 -> 0.8389 Offset: -6.7ppm
H2O span 1.0093 -> 1.0061 Offset: 0.07ppm
Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm
LI-7500A 75H-2176 2021-03-01 Lab Calibration Lab Last calibration: 2020-12-02
AGC=102.5 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=22.9C, Tdew=4.6C
Zero Air at 450psi, AMP CA04396 408.97 at 250 Tdewspan=20.0
CO2 zero 0.8673 -> 0.8671 Offset: -0.25ppm
CO2 span 1.009 -> 1.000 Offset: 2.53ppm
H2O zero 0.8531 -> 0.8524 Offset: -1.3ppm
H2O span 0.9238 -> 0.9092 Offset: 0.37ppm
Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm.
LI-7500A 75H-2150 2021-03-01 Lab Calibration Lab Last calibration: 2020-08-20
AGC=100.8 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=27.4C, Tdew=2.08C
Zero Air at 500psi, AMP CA04396 408.97 at 250 Tdewspan=20.0
CO2 zero 0.9019 -> 0.9019 Offset: -0.3ppm
CO2 span 1.0014 -> 0.9975 Offset: 1.93ppm
H2O zero 0.8665 -> 0.8675 Offset: 6.8ppm
H2O span 1.011 -> 1.0040 Offset: 0.15ppm
Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm.
NR01 2420 2021-02-26 Lab none Installed Sherman Wetland Net Radiometer 4-way Cal A Installed on the Sherman Wetland tower for inter-comparison after repair in the lab. It is clamped near the end of the square tube of the radiation boom pointing south.
CNR4 121038 2021-02-26 Factory none Installed Sherman Wetland Net Radiometer 4-way Cal B Installed on the Sherman Wetland tower for inter-comparison. This is the AMP standard. It is clamped near the end of the square tube of the radiation boom pointing south.
CR1000 23865 2021-02-26 Lab none Installed Sherman Wetland Data Logger Cal Installed on the Sherman Wetland tower to collect data from two 4-way net radiometers for inter-comparison
BB-Fab BB-SW_Temp1 2021-02-26 Sherman Wetland-Temp1 Battery Bank Disposed Disposed none Removed from the decommissioned temporary location. The batteries were brought back to the lab and will be used at Hill Slough but this battery bank has ceased to exist as an equipment entry
KD140SX-UFBS 11ZV0M0829 2021-02-26 Sherman Wetland-Temp1 Solar Panel A Moved Hill Slough Solar Panel A Removed from the decommissioned temporary location for use at Hill Slough
KD140SX-UFBS 11ZV0M0836 2021-02-26 Sherman Wetland-Temp1 Solar Panel B Moved Hill Slough Solar Panel B Removed from the decommissioned temporary location for use at Hill Slough
KD140SX-UFBS 11ZV0M0855 2021-02-26 Sherman Wetland-Temp1 Solar Panel C Moved Hill Slough Solar Panel C Removed from the decommissioned temporary location for use at Hill Slough
PAR-LITE 050597 2021-02-26 Sherman Wetland-Temp1 PAR Sensor Incoming Removed Lab none Removed from the decommissioned temporary location and brought back to the lab to be cleaned up.
LI-610 DPG478B 2021-03-01 Lab none Maintenance Lab none The dew point was stuck at about 34C - not dropping to the set point of 20C. I found a crack in the "from cooler" tubing. I replaced this piece of tubing but this didn't solve the problem. There was an air bubble in this piece of tubing and the machine was making an unusual bubbling sound. Refilling the reservoir with DI water immediately change the sound, got rid of the air bubble in the tubing and the dew point began to drop towards the set point.
Mark II C003800 2021-02-25 Lab none Disposed Disposed none Returned to Arable
Mark II C003789 2021-02-25 Lab none Disposed Disposed none Returned to Arable
Mark II C002469 2020-04-23 Lab none Disposed Disposed none Returned to Arable
Mark II C002445 2020-04-23 Lab none Disposed Disposed none Returned to Arable
Mark II C002714 2020-04-23 Lab none Disposed Disposed none Returned to Arable
Mark II C006322 2021-02-04 Lab none Installed Bouldin Alfalfa Weather Station PL_N-B Installed on the northern Planet Labs post to replace one of the two units eaten by the sheep
Mark II C006282 2021-02-04 Lab none Installed Bouldin Alfalfa Weather Station PL_N-A Installed on the northern Planet Labs post to replace one of the two units eaten by the sheep
Mark I A001311 2021-02-25 Lab none Disposed Disposed none Returned to Arable due to end of life. The was a cal val unit and was replaced by a new Mark II
Mark I A001359 2021-02-25 Lab none Disposed Disposed none Returned to Arable due to end of life. The was a cal val unit and was replaced by a new Mark II
Mark I A001353 2021-02-25 Lab none Disposed Disposed none Returned to Arable due to end of life. The was a cal val unit and was replaced by a new Mark II
CR1000 23865 2021-02-25 Lab none Maintenance Lab none The internal lithium battery was nearly dead 1.6V. I couldn't find a new one but swapped it for one reading 2.6V instead. Be sure the check the clock when using this logger.
CNR4 121038 2019-11-17 Factory none Calibration Factory none Initial calibration by Kipp and Zonen. LW on 2019-11-17: LW_in = 6.83 uV/(W/m2), LW_out = 5.48 uV/(W/m2). SW on 2019-11-15: SW_in = 14.56 uV/(W/m2), SW_out = 14.17 uV/(W/m2)
NR01 2420 2021-02-18 Lab none Repair Lab none Over the last week or so I disassembled and cleaned this sensor. I don't see any water damage but I can't get the LW junction board loose as it seems glued to the case. I totally replaced the four wires running between the LW and SW sides. These wires were broken after having been previously broken and spliced. I soldered new wires directly to the junction boards. I collected about two days of data on the window sill with the sensor partially disassembled so that all four sensors could point up together. The relationship between the paired sensors was very tight (r-squared >0.99) - excluding a portion of the LW data that might have been affected by solar heating. LW had a sloped of 1.01 but SW had a slope of 1.20 - new calibration coefficients are probably needed. Today I completely reassembled the sensor and put it out the window on a boom to collect some data to check the assembly.
PAR-LITE 990017 1999-02-09 Factory none Calibration Factory none Initial factory calibration: 5.16uV/(umol/m2/s)
PAR-LITE 990019 1999-02-09 Factory none Calibration Factory none Initial factory calibration; 5.26uV/(umol/m2/s)
PRB-K Type #8 000000402FF3 2021-02-10 Tonzi Soil Moisture Probe F -0.50m Removed Lab none Removed from Tonzi floor due to fixed readings for a long while. This sensor was out of the ground at the edge of the trees near the path to the tower. It was buried vertically with the top of the sensor 31cm below the surface - the time were 44-50cm below the surface.
PAR-LITE 990031 2021-02-10 Tonzi PAR Sensor Incoming Cal Removed Lab none Removed from Tonzi Tower after three weeks of inter comparison with the AMP standard
PAR-LITE 990025 2021-02-10 Cal Kit PAR Sensor Incoming Removed Lab none Removed from TZ Tower after three weeks of inter comparison with the AMP standard
WM 1590 153805 2021-02-10 Lab none Installed Cal Kit Anemometer Installed on the TZ Tower for inter-comparison. This sensor is the one that got wet when the BC tower fell over. It's mounted on the post just north of the main eddy sensors, at the same height, and oriented to true north.
eosFD FD20200001 2021-02-10 Lab none Installed Vaira Forced Diffusion North Installed near the old north soil co2 profile. This unit has a homemade solar shield to try to keep it cooler. Could use a soil collar.
eosFD FD20200001 2021-02-04 Factory none Returned Lab none Returned from Eosense where its pump relay board was replaced under warranty repair.
GP-001 GP20140026 2021-02-04 West Pond CO2 Probe B Removed Lab none Removed from WP and brought back to the lab b/c this sensor has been reading noise since late Nov
LI-7500A 1594 2021-02-01 Lab Calibration Lab Last calibration: 2020-08-20
AGC=99.3 SDM=0 DAC1=- DAC2=-
Ambient P=100.3kPa, Tair=28.4C, Tdew=11.0C
Zero Air at 0psi, AMP CA04396 408.97 at 0 Tdewspan=20.0
CO2 zero 0.8651 -> 0.8646 Offset: -2.1ppm
CO2 span 1.02124 -> 1.0131 Offset: -0.32ppm
H2O zero 0.8838 -> 0.8890 Offset: 29.85ppm
H2O span 1.015 -> 1.000 Offset: 0.35ppm
Notes: The usual 7550 box in the lab is running the newest firmware v8.8 and is complaining that the v8.5 config file has bad number formats for GPS locations. I used 7550 sn AIU-1866 for calibration instead. This box and head will go to replace equipment on the BC tower that fell in the water.
NR01 2420 2021-01-30 Lab none Damaged Lab none This sensor was removed from a submerged position on the fallen BC tower. The shortwave side had been torn loose breaking the wires similar to last time it was damaged. After returning to the lab on the 29th I pulled it apart and tried to air it out. On the 30th it was still damp inside and I ran dry air through the holes under the circuit boards for two days.
WM 1590 153805 2021-02-01 Lab none Tested Lab none After being removed from a partially submerged position on the fallen BC tower, I tested this sensor in the lab. There was no evidence that water had leaked inside. I connected it to a Licor system and used fans to blow air through in on the floor. The response sign and magnitude looked about right. I plan to run it on the Tonzi tower for a couple weeks to check its performance.
LI-7500A 75H-2176 2021-02-01 Lab none Tested Lab none On the 29th after being retrieved from a partially submerged position on the fallen BC tower, I removed the desiccant bottles and saw no evidence that water had gotten inside. I flowed dry air into the desiccant bottle openings over night. On the 1st, after consultation with Licor, I powered up this head on the calibration rig in the lab. It seemed to operate fine. I flooded the base with zero air before sealing it back up but it needs to run for a while to equilibrate and be calibrated.
LI-7700 TG1-0155 2021-02-04 Lab none Shipped Factory none Shipped to Licor and RMA 00225226 for evaluation after being submerged when the BC tower fell in the storm last week.
LI-7700 TG1-0155 2021-02-01 Lab none Tested Lab none On the 29th I removed this sensor from being submerged after the BC tower fell just after midnight on the morning of the 27th. On the evening of the 29th, about 1/2 cup of water came out of the bottom when I removed the base. I removed the desiccant bottle in the head and saw no evidence that water had leaked in there. I left dry air flowing inside the base over night. On the 30th I removed the bottom case and only saw minor water damage on the pressure sensor. On the 1st, after consultation with Licor, I powered up the analyzer and got green power and status lights but no communications on the Ethernet. It will need to be returned to Licor for further evaluation.
SRS-Pi PRI 897103002 2021-02-03 Lab none Installed Bouldin Corn Narrow Band PRI Reflected Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
SRS-Ni NDVI 2046503357 2021-02-03 Lab none Installed Bouldin Corn Narrow Band NDVI Reflected Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
SRS-Ni NDVI 882103020 2021-02-03 Lab none Installed Bouldin Corn Narrow Band NDVI Incoming Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
CNR1 990181 2021-02-03 Lab none Installed Bouldin Corn Net Radiometer 4-way Installed on the BC tower to replace the Hukseflux NR01 that broke when the tower fell and still needs repair and cross comparison.
Mark II C006123 2021-02-03 Lab none Installed Bouldin Corn Weather Station B Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
Mark II C003251 2021-02-03 Lab none Installed Bouldin Corn Weather Station C Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
PA-II-SD 68c63acc07a9 2021-02-03 Lab none Installed Bouldin Corn Air Quality Sensor Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
SD130BN 0030F4D21491 2021-02-03 Lab none Installed Bouldin Corn Camera Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
PQS 1 140441 2021-02-03 Lab none Installed Bouldin Corn PAR Sensor Reflected Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
PQS 1 140453 2021-02-03 Lab none Installed Bouldin Corn PAR Sensor Incoming Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
SRS-Pi PRI 913003004 2021-02-03 Lab none Installed Bouldin Corn Narrow Band PRI Incoming Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week
LI-7500A 1594 2021-02-03 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed on the BC tower to replace sensors damaged when the tower fell last week
LI-7550 AIU-1866 2021-02-03 Lab none Installed Bouldin Corn Interface LI-7550 Installed on the BC tower to replace sensors damaged when the tower fell last week
LI-7700 TG1-0224 2021-02-03 Lab none Installed Bouldin Corn Gas Analyzer CH4 Installed on the BC tower to replace sensors damaged when the tower fell last week
WM 1590 153803 2021-02-03 Lab none Installed Bouldin Corn Anemometer Installed on the BC tower to replace sensors damaged when the tower fell last week
WL16U 57278 2018-05-08 Lab none Repair Lab none I stripped off the old rotting insulation and then slid on a long continuous piece of heat shrink. I shrunk the whole length of the cable. I disassembled the data logging enclosure and kept only the data logger board and USB connector in the end cap. I cut off the twin 9V battery plugs and will power the sensor from solar connected rechargeable battery.
WL16U 57278 2018-03-27 Tonzi Water Level Floor Removed Lab none Removed from the Floor Well. The cable insulation is rotting and needs to be replaced. Replaced by the new METER water level sensor a couple of weeks ago.
CMTD-As4s0 3317052234210 2018-05-11 Lab none Installed Tonzi none Installed at the Road Well to power Global Water sensor.
WL16U 57278 2018-05-11 Lab none Installed Tonzi Water Level Road Installed at the Road Well to measure water level at 5min intervals storing 30min averages
Mark II C006123 2021-01-29 Bouldin Corn Weather Station B Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Maybe some rain in the Tair bulge
Mark II C003251 2021-01-29 Bouldin Corn Weather Station C Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Maybe some rain in the Tair bulge
PA-II-SD 84f3ebd5ffb1 2021-01-29 Bouldin Corn Air Quality Sensor Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Got rained on and the power connector is corroded
SD130BN 0030F4D21491 2021-01-29 Bouldin Corn Camera Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
SRS-Pi PRI 897103002 2021-01-29 Bouldin Corn Narrow Band PRI Reflected Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
SRS-Pi PRI 913003004 2021-01-29 Bouldin Corn Narrow Band PRI Incoming Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27.
SRS-Ni NDVI 2046503357 2021-01-29 Bouldin Corn Narrow Band NDVI Reflected Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
SRS-Ni NDVI 882103020 2021-01-29 Bouldin Corn Narrow Band NDVI Incoming Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
PQS 1 140441 2021-01-29 Bouldin Corn PAR Sensor Reflected Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
PQS 1 140453 2021-01-29 Bouldin Corn PAR Sensor Incoming Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged
NR01 2420 2021-01-29 Bouldin Corn Net Radiometer 4-way Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged and broken in two pieces
LI-7550 AIU-0614 2021-01-29 Bouldin Corn Interface LI-7550 Removed Lab none Removed due to the tower falling over in the storm near midnight on 2021-01-26/27.