Instrument:PAR Sensor
Manufacture:Kipp & Zonen
Model No:PAR-LITE
Serial No:990031
Tag No:
Acquisition Date:0000-00-00
Project:
Location on 2022-02-02:Gilbert Tract,
Notes:

Documents

Calibration: cal_KippZonen_PARLite_sn990031_19990618.pdf
Report: Vaira PAR Cross Compare - 2020-06-12

History

Rec No.DateLoc FromTitleLoc ToNotes
5670 2022-02-02 East End PAR Sensor Undercanopy C Moved Gilbert Tract PAR Sensor Incoming Removed from lower location south of the tower at East End. This sensor was disconnected from the data logger two weeks before. This sensor was moved to GT to replace the incoming PAR there that was having negative values at night.
5231 2021-03-30 Lab none Installed East End none Installed in EE wetland to measure undercanopy PAR. This sensor is S of the tower mounted on a grounding rod. The sensor is sitting in a heterogeneous patch of dried reeds about 30cm above the water level
5114 2021-02-10 Tonzi PAR Sensor Incoming Cal Removed Lab none Removed from Tonzi Tower after three weeks of inter comparison with the AMP standard
5058 2021-01-21 Lab none Installed Tonzi none Installed as part of the inter-comparison kit to test PAR, AirT and RH on the top of the Tower.
4936 2020-10-26 Lab none Repair Lab none The cable insulation on this sensor was rotted. I cut the cable off near the sensor and spliced on a new cable. I also added an aluminum bracket to provide strain relief and used liquid electrical tape to seal the hole where the cable enters the sensor.
4805 2020-09-29 Vaira PAR Sensor Reflected Cal Removed Lab none Removed from the tower (reflected cal) for repair - the cable is all rotted
4801 2020-09-16 Vaira PAR Sensor Reflected Moved Vaira PAR Sensor Reflected Cal Moved from the main reflected position to the reflected cal position in preparation to being removed.
31 2010-01-08 Vaira PAR Sensor Reflected Location Vaira PAR Sensor Reflected This instrument is installed in the downward diretion collecting PAR reflected from the surface. Calibration factor on instrument is 5.08 µV/µmol/s.m2
2344 2000-10-07 Lab none Installed Vaira PAR Sensor Reflected Installed facing down with the initial set up of the Vaira site.
 
 