View history entry 5252 for Decagon-METER CTD-10 sn: 1260403156

Date: 2021-04-29
From: Lab - none
To: Lab - none
Title: Tested
Notes: I couldn't get this sensor to work in the field. Seem to test okay on a CR1000 terminal using the SDI-12 pass through. Will try again in the field again next time. It's using address 0 (zero): 0I! - id, 0M! - Measure, 0D0! - Data. Note CRBasic program automatically issues 0D0! command after 0M! command.
Entered by: Joe Verfaillie