Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Forerunner-Eosense
Model No:GP-001
Serial No:GP20210002
Tag No:
Acquisition Date:2021-03-15
Project:CDFW
Location on 2023-11-28:Lab,
Notes: This is a replacement for sn GP20140017. A new sensor board and filter were installed in the casing and calibrated to 0-2% range.

Documents

Calibration: cal_Eosense_GP-001_snGP20210002_20210309.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
6475 2023-11-28 Gilbert Tract CO2 Probe Removed Lab none Removed from GT because of noisy data. Replaced by GP20160117.
6225 2023-03-23 Lab none Installed Gilbert Tract CO2 Probe Installed at Gilbert Tract on a post (just off the far end of the boardwalk) with the other water quality sensors. Should be at the same height as before (about 5cm above the soil)
5201 2021-03-09 Factory none Calibration Factory none Factory calibration y=mV/5000*1.2*2%