Equipment and Calibration

View History

Equipment Entry

Instrument:Forced Diffusion
Manufacture:Forerunner-Eosense
Model No:eosFD
Serial No:FD20180068
Tag No:
Acquisition Date:2019-11-12
Project:Borrowed
Location on 2022-08-23:Disposed,
Notes: One of five FD chambers loaned to us from Eosense for experiments. This is the second batch or borrowed chambers.

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
6264 2022-08-24 Factory none Disposed Disposed none This unit was originally borrowed from Eosense. Since this model is no longer in production they can not repair it ahead of other purchased units so we will not be getting it back.
6263 2022-08-23 Lab none Shipped Factory none Sent to Eosense on RMA #220809 for repair.
5579 2021-11-17 Vaira Forced Diffusion West Removed Lab none Removed from Vaira West due to bad CO2 concentration readings
5373 2021-07-13 Vaira Forced Diffusion West Maintenance Vaira Forced Diffusion West Performance check done. Referenced to 400ppm.
Average Flux: -0.0, 0.08 STDEV
Average ATM: 610.7ppm, 49.6C
Average Soil: 609.8ppm, 49.8C
5218 2021-03-17 Lab none Installed Vaira Forced Diffusion West Installed at Vaira to replace sn FD20190028 for possible use at Hill Slough.
5200 2021-03-15 Factory none Returned Lab none Returned from Eosense. It was sent in for zero CO2 readings but Eosense reported that it seemed to be reading correctly. It was calibrated from CO2 concentration and CO2 flux.
5199 2021-01-12 Lab none Shipped Factory none Sent to Eosense for repair due to zero CO2 readings
4932 2020-10-22 Tonzi-SoilCO2 Forced Diffusion Tree 1 Removed Lab none Removed from the under tree location at Tonzi because its co2 sensor has been reading zero. Replaced by sn FD20180048
4173 2019-11-19 Lab none Installed Tonzi Forced Diffusion Tree 1 Installed under the tree at the Tonzi soil co2 profiles