Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 [35] 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71
Ser No.DateLoc FromTitleLoc ToNotes
LI-7500A 75H-2176 2019-01-04 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace sn 0418 for calibration.
LI-7500A 0418 2019-01-04 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira tower for calibration. Replaced by sn 75H-2176.
GP-001 GP20140019 2019-01-03 East End CO2 probe Removed Lab none Removed and brought back to the lab for testing.
LI-7500A 0041 2019-01-02 Lab none Calibration Lab none Last calibration: 2018-07-03
AGC=98 SDM=0 DAC1=0 DAC2=0
Ambient P=101.2kPa, Tair=20.1C, Tdew=3.7C
Zero Air at 1400psi, AMP Air 408.97 at 875 Tdewspan=16.00
CO2 zero 0.9966 -> .9953 Offset: -4.19ppm
CO2 span 1.0173 -> 1.0194 Offset: -1.25ppm
H2O zero 0.8192 -> .8248 Offset: 34.0ppm
H2O span 1.0285 -> 1.0312 Offset: -.07ppm
Notes: H2O zero offset larger than I expected, but H2O span seems fine.
LI-7500A 0042 2019-01-02 Lab none Calibration Lab none Last calibration: 2018-07-10
AGC=98.7 SDM=0 DAC1=0 DAC2=0
Ambient P=101.3kPa, Tair=19.2C, Tdew=.7C
Zero Air at 1400psi, AMP Air 408.97 at 875 Tdewspan=16.00
CO2 zero 0.9485 -> .9489 Offset: 1.45ppm
CO2 span 1.0039 -> 1.0187 Offset: -7.39ppm
H2O zero 0.8406 -> .8434 Offset: 18.2ppm
H2O span 1.0158 -> 1.0334 Offset: -.37ppm
Notes: Dewpoint generator was set at 16.03C and I adjusted it to 16.00 before beginning calibration. H2O zero offset larger than I expected, but H2O span seems fine.
LI-7500A 75H-2176 2019-01-02 Lab none Calibration Lab none Last calibration: 2018-07-10
AGC=102 SDM=0 DAC1=0 DAC2=0
Ambient P=101.2kPa, Tair=20.5C, Tdew=3.3C
Zero Air at 1250psi, AMP Air 408.97 at 850 Tdewspan=16.00
CO2 zero 0.8671 -> .8676 Offset: 2.45ppm
CO2 span 1.0029 -> 1.0114 Offset: -4.48ppm
H2O zero 0.848 -> .8525 Offset: 27.5ppm
H2O span 0.9088 -> .8206 Offset: -.27ppm
Notes: H2O zero offset larger than I expected, but H2O span seems fine.
BB-Fab BB-TZ_soilco2 2018-12-13 Tonzi Upgrade Tonzi This system has been failing overnight partly because one set of Vaisala probes have not been cycling off. I also added to old batter most recently used on East End.
BB-Fab BB-TF 2018-12-13 Tonzi-Floor Upgrade Tonzi-Floor Nine old G27 flooded batteries, a mixture of ages, were removed.
Most tested bad. Two tested ok at first but weak later. They were replaced by four G31 flooded batteries wired in pairs for a 24V system. The panels and tower ends were also modified for 24V.
RF401 35785 2018-12-19 Tonzi Removed Lab none Removed from the OSU SP1 system to test out data collection by LattePanda in the lab.
RF401 35754 2018-12-19 Tonzi Removed Lab none Removed from the OSU SP3 system to test out data collection by LattePanda in the lab.
Mark I A000251 2018-12-18 Lab none Installed Bouldin Alfalfa Weather Station A The Arable Mark was reinstalled after charging in the lab.
Mark I A000251 2018-12-13 Bouldin Alfalfa Weather Station A Removed Lab none Brought back to the lab because it did not seem to be charging in the field. Seemed to charge up fine in the lab.
SI-121 1282 2010-12-09 Tonzi Removed Lab none Removed from Tonzi with the end of the surface temperature experiment.
SI-121 1281 2010-12-09 Tonzi Removed Lab none Removed from Tonzi with the end of the surface temperature experiment.
LI-7500A 0041 2002-02-08 Lab none Installed Tonzi-Floor Installed on the Tonzi Tower to replace sn 0035 for calibration
LI-7500A 0035 2002-02-08 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor for calibration. Replaced by sn 0041.
NR-LITE 990350 2002-02-08 Lab none Installed Tonzi Gas Analyzer CO2/H2O Installed inside the orange fence on the floor met CR10x data logger.
LI-7500A 0073 2002-02-08 Lab Gas Analyzer CO2/H2O Installed Tonzi-Tower Installed on the Tonzi Tower to replace sn 0042 for calibration
LI-7500A 0042 2002-02-08 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from the Tonzi Tower for calibration. Replaced by sn 0073.
DO6400/T DO-06 2018-12-06 East End DO Sensor Removed Lab none Removed from East End because it was not working and has been made obsolete by MiniDot sensor.
Mark I A000251 2018-11-28 Lab none Installed Bouldin Alfalfa Weather Station A Installed at BA to replace sn A000591, which will be sent back to Arable for inspection.
Mark I A000591 2018-11-28 Bouldin Alfalfa Weather Station A Removed Lab none Removed from BA and brought back to the lab. We will send this back to Arable b/c they noticed a faulty thermocouple. It was replaced by A000251.
HMP60 L1740420 2018-11-14 Factory Received Lab none Received back from Campbell Sci. They tested it against their reference for temp and RH at 90%, 50%, and 20% and could not find any issues. We should test it again in the lab.
HMP60 L1740398 2018-11-14 Factory Received Lab none Received back from Campbell Sci. They tested it against their reference for temp and RH at 90%, 50%, and 20% and could not find any issues. We should test it again in the lab.
HMP60 L2310736 2018-09-14 Lab none Disposed Factory Shipped back to Campbell Sci for testing (low readings). They found the sensor to be irreparably bad and disposed of it for us.
HMP60 L1740420 2018-09-14 Lab none Shipped Factory Shipped back to Campbell Sci for testing (drifting downward).
HMP60 L1740398 2018-09-14 Lab none Shipped Factory Shipped back to Campbell Sci for testing (drifting downward).
BB-Fab BB-TF 2018-10-29 Tonzi-Floor Maintenance Tonzi-Floor One of the batteries in the battery bank seems to have an internal short and was sapping power. The bad battery was removed. Power started to be bad on October 20.
GP-001 GP20140016 2018-11-06 Lab none Tested Lab none Connected to a CR6 and recorded 10s data. The sensor was maxing out (reading ~5V) on both 0-2% and 0-6% ranges. The connection between sensor and cable seemed secure and clean. The sensor should go back to the factory.
GP-001 GP20140018 2018-11-06 Lab none Tested Lab none Connected to a CR6 and recorded 10s data. The sensor was maxing out (reading ~5V) on both 0-2% and 0-6% ranges. The connection between sensor and cable seemed secure and clean. The sensor should go back to the factory.
LI-7700 TG1-0421 2018-11-06 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2018-05-14
RSSI=81%, Optical RH=13%
Ambient P=100.2kPa, Tair=26.7C
Zero Air at 1350psi, AMP Air 1.9497 at 950
CH4 Zero Offset: -0.015ppm
CH4 Span Offset: 0.0083ppm
Notes: First calibration after power board replacement
LI-7700 TG1-0421 2018-11-05 Lab none Upgrade Lab none I swapped out power board pn 9677-079G per Licor's instructions. The new power board is pn 9977-079. According to Licor, "the original board needs a new resistor for the "under-voltage" problem (potential to drop out when heater and washers run simultaneously while incoming power is lower than usual, say 11-10 VDC)". The new board Rev 0A (V3) should not have that problem.
GP-001 GP20140026 2018-11-01 Lab none Installed West Pond CO2 Probe B Installed at WP to replace sn GP20140018, which was brought back to the lab for too-high readings.
GP-001 GP20140018 2018-10-25 West Pond CO2 Probe B Removed Lab none Removed from WP because of high readings. It was replaced by sn GP20140026 two weeks later.
LI-7700 TG1-0423 2018-11-01 Lab none Installed East End Gas Analyzer CH4 Installed at EE to replace sn TG1-0421, which was brought back to the lab for a power board replacement and calibration.
LI-7700 TG1-0421 2018-11-01 East End Gas Analyzer CH4 Removed Lab none Removed from EE and brought back to the lab for a power board replacement and calibration. It was replaced by sn TG1-0423
GP-001 GP20140084 2018-11-01 Lab none Installed Sherman Wetland CO2 Probe Installed at SW to replace sn GP20140016, which was removed b/c it was maxing out.
Mark I A000591 2018-11-02 Bouldin Alfalfa Weather Station A Notes Bouldin Alfalfa Weather Station A Arable e-mailed to say sn A000591 which has a bad backup temperature sensor and requested we send it back. Not knowing what the exact source of the failure is, and having checked and found no similar behavior in any other device, they want to find out why. We planned to remove this Arable on 2018-10-04, but instead removed sn A000251 from BC and sent that Arable back to the factory. The Arable at BC was replaced by sn A001911. sn A000591 is still at BA for now.
Mark I A000251 2018-10-04 Bouldin Corn Weather Station A Removed Lab none Removed from Bouldin Corn to be sent back to Arable for inspection of back up temperature sensor. Replaced by sn A001911. Later we realized we removed the incorrect Arable and they had requested instead sn A000591.
LI-7700 TG1-0423 2018-10-30 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2018-05-08
RSSI=80%, Optical RH=8%
Ambient P=100.8kPa, Tair=27.97C
Zero Air at 1500psi, AMP Air 1.9497 at 1000
CH4 Zero Offset: -0.015ppm
CH4 Span Offset: -0.0027ppm
Notes: First calibration after power board was replaced
LI-7700 TG1-0423 2018-10-29 Lab none Upgrade Lab none I swapped out power board pn 9677-073G per Licor's instructions. The new power board is pn 9977-079. According to Licor the original boards Rev 0, could be damaged by exposure to voltage above 24VDC. The new board Rev 0A (V3) should not have that problem.

Note: This instrument was replaced with 2 original screws and 2 long screws. The long screws were used to help re-assemble the instrument, and I forgot to swap them for the original, shorter, screws.
LI-7500A 75H-2150 2018-10-28 Lab none Installed East End Gas Analyzer CO2/H2O Installed at EE to replace sn 75H-2176, which was brought back to the lab for calibration.
LI-7500A 75H-2176 2018-10-28 East End Gas Analyzer CO2/H2O Removed Lab none Removed from EE and brought back to the lab for calibration. It was replaced by sn 75H-2150.
NR-LITE 990354 2018-10-30 Lab none Location Lab none This sensor is on the tram and has probably been there the last 10 years.
PAR-LITE 030381 2018-10-30 Lab none Location Lab none This sensor has been on the tram in the outgoing position for at least the last 10 years.
PAR-LITE 030380 2018-10-30 Lab none Location Lab none This sensor has been on the tram in the incoming position for at least the last 10 years.
LI-7700 TG1-0424 2018-10-25 Lab none Installed Sherman Wetland Gas Analyzer CH4 Installed at SW to replace sn TG1-0423, which was brought back to the lab for power board replacement and for calibration. Heater settings changed from Bouldin settings (top: 25%, 2 degC, 18:00-11:00; bottom: 25%, 30%, 18:00-11:00) to Sherman/Twitchell settings (top: 5%, 2 degC, 18:00-9:00; bottom: 5%, 30%, 18:00-9:00).
LI-7700 TG1-0423 2018-10-25 Sherman Wetland Gas Analyzer CH4 Removed Lab Gas Analyzer CH4 Removed from SW and brought back to the lab for a power board swap and calibration. It was replaced by sn TG1-0424.
LI-7500A 75H-2182 2018-10-25 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at WP to replace sn 0042, which was brought back to the lab for calibration.
LI-7500A 0042 2018-10-25 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from WP and brought back to the lab for calibration. it was replaced by sn 75H-2182.
TR-5251 40415-1006 2018-10-25 Mayberry Precipitation Tested Mayberry Precipitation Repaired chewed cable and tested with 475ml water. Got 100 tips. Did not adjust set screws.
PAR-LITE 040412 2018-10-25 Lab none Installed Mayberry PAR Sensor Reflected Water Installed at MB (measuring reflected water) to replace sn 010161, which was brought back to the lab for repair/calibration.
PAR-LITE 010161 2018-10-25 Mayberry PAR Sensor Reflected Water Removed Lab none Removed from MB (measuring reflected water) and brought back to the lab to replace the rotted cable and for calibration. It was replaced by sn 040412.
PAR-LITE 010159 2018-10-25 Lab none Installed Mayberry PAR Sensor Incoming Installed at MB in the incoming position to replace sn 10162, which was brought back to the lab for repair/calibration.
PAR-LITE 10162 2018-10-25 Mayberry PAR Sensor Incoming Removed Lab none Removed from MB and brought back to the lab to replace the rotted cable and for calibration. It was replaced by sn 010159.
LI-7700 TG1-0211 2018-10-25 East Pond Gas Analyzer CH4 Calibration East Pond Gas Analyzer CH4 Calibration Location: East Pond
Last calibration: 2018-04-13
RSSI=74%, Optical RH=10%
Ambient P=101.6kPa, Tair=26.4C
Zero Air at 1950psi, AMP Air 3.121 at 2100
CH4 Zero Offset: .029ppm
CH4 Span Offset: .006ppm
Notes: Calm day
LI-7700 TG1-0155 2018-10-25 Bouldin Corn Gas Analyzer CH4 Calibration Bouldin Corn Gas Analyzer CH4 Calibration Location: Bouldin Corn
Last calibration: 2018-04-04
RSSI=91%, Optical RH=16%
Ambient P=101.5kPa, Tair=27.2C
Zero Air at 1925psi, AMP Air 3.121 at 1975
CH4 Zero Offset: .015ppm
CH4 Span Offset: -.013ppm
Notes: Gently breezy day
LI-7700 TG1-0224 2018-10-25 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration: 2018-04-04
RSSI=70%, Optical RH=22%
Ambient P=101.7kPa, Tair=25.5C
Zero Air at 1950psi, AMP Air 3.121 at 2100
CH4 Zero Offset: -.004ppm
CH4 Span Offset: -.002ppm
Notes: Calm day. Before calibrating, we flushed head spaced with zero air. Optics RH dried down from 22% to 13% before calibration started.
LI-7700 TG1-0223 2018-10-25 Mayberry Gas Analyzer CH4 Calibration Mayberry Gas Analyzer CH4 Calibration Location: MB
Last calibration: 2018-04-25
RSSI=81%, Optical RH=8%
Ambient P=101.6kPa, Tair=20.5C
Zero Air at 2000psi, AMP Air 3.121 at 2100
CH4 Zero Offset: 0.021ppm
CH4 Span Offset: 0.018ppm
Notes: Calm day
LI-7500A 75H-2182 2018-10-24 Lab none Calibration Lab none Last calibration: 2018-07-10
AGC=87.3 SDM=0 DAC1=0 DAC2=0
Ambient P=100.55kPa, Tair=26.87C, Tdew=12.61C
Zero Air at 1500psi, AMP Air 408.97 at 1100 Tdewspan=16.00
CO2 zero 0.8841 -> .8844 Offset: 1.11ppm
CO2 span 1.0168 -> 1.0188 Offset: -1.04ppm
H2O zero 0.845 -> .8449 Offset: -.6ppm
H2O span 1.0202 -> 1.0143 Offset: .13ppm
Notes:
LI-7500A 75H-2150 2018-10-24 Lab none Calibration Lab none Last calibration: 2018-07-10
AGC=97.5 SDM=0 DAC1=0 DAC2=0
Ambient P=100.6kPa, Tair=26.7C, Tdew=11.95C
Zero Air at 1500psi, AMP Air 408.97 at 1100 Tdewspan=16.00
CO2 zero 0.9017 -> .9021 Offset: 1.86ppm
CO2 span 1.002 -> 1.0050 Offset: -1.74ppm
H2O zero 0.8662 -> .8660 Offset: -1.9ppm
H2O span 1.0069 -> 1.0069 Offset: 0.00ppm
Notes: No change in H2O span.
LI-7700 TG1-0424 2018-10-24 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2018-04-13
RSSI=82.9%, Optical RH=9%
Ambient P=100.6kPa, Tair=27.8C
Zero Air at 1500psi, AMP CH4 1.9497 at 1150
CH4 Zero Offset: -0.015ppm
CH4 Span Offset: 0.043ppm
Notes: First calibration after replacing power board - no problems found.
LI-7700 TG1-0424 2018-10-22 Lab none Upgrade Lab none I swapped out power board pn 9677-079 per Licor's instructions. According to Licor the original boards Rev 0, could be damaged by exposure to voltage above 24VDC. The new board Rev 0A (V3) should not have that problem.
TE525MM 49187-109 2018-10-18 Sherman Barn Precipitation Calibration Sherman Barn Precipitation Got 94 of 100 tips - adjusted by a total of 3.5 turns CCW (screws were loose, so this total is suspicious; should have been a CW adjustment when all was said and done)

2018-10-18: Got 94 of 100 tips - adjusted by 0.5 turns CW.
2018-12-18: Got 117 of 100 tips - adjusted by 1 turn CCW, but screws feel very loose.
2019-01-03: Got 118 of 100 tips - adjusted by 3 turns CCW. (previous adjustment of 1 turn CCW did not seem to make a difference).
2019-01-24: Got 100 of 100 tips! No adjustments made.
LI-7500A 75H-2669 2018-10-18 Lab none Installed Sherman Barn Gas Analyzer CO2/H2O Installed on Sherman Barn to replace sn 0041 for calibration
LI-7500A 0041 2018-10-18 Sherman Barn Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from Sherman Barn for calibration. Replaced by sn 75H-2669.
GP-001 GP20140016 2018-10-18 Sherman Wetland CO2 Probe Removed Lab none Removed from Sherman Wetland due to railed readings not responding the atmospheric CO2.
TE525MM 65552-615 2018-10-18 Sherman Wetland Precipitation Calibration Sherman Wetland Precipitation Got 94 of 100 tips - adjusted screws a total of 1.5 turns CW

2018-10-18: Got 94 of 100 tips - adjusted screws 0.5 turns CW
2018-10-25: Got 96 of 100 tips - adjusted set screws 1 turn CW.
2018-12-18: 102 tips - no adjustment.
TR-525M 50229-112 2018-10-18 East End Precipitation Calibration East End Precipitation Got 94 of 100 tips - adjusted screws a total of 0.5 turns CW.

2018-10-28: Got 98 of 100 tips and adjusted set screws 0.5 turn CW.
2018-11-01: Got 93 of 100 tips - no adjustment and need to retest
2018-12-18: 98 of 100 tips - no adjustment.
LI-7700 TG1-0424 2018-10-18 Bouldin Alfalfa Gas Analyzer CH4 Removed Lab none Removed from Bouldin Alfalfa for power board upgrade and to swap in for other two 7700s needing upgrade.
LI-7500A 75H-2667 2018-10-18 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed on Bouldin Alfalfa to replace sn 74H-2150 for calibration.
LI-7500A 75H-2150 2018-10-18 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Alfalfa for calibration. Replaced by sn 75H-2667.
TE525MM 65560-615 2018-10-18 Bouldin Alfalfa Precipitation Calibration Bouldin Alfalfa Precipitation Got 90 of 100 tips. Adjusted screws a total of 2.5 turns CW.

2018-10-18: Got 90 tips and adjusted screws 1 turn CW.
2018-11-01: Got 91 tips and adjusted screws 2 turns CCW accidentally (should have been CW).
2018-12-13: Got 86 tips and adjusted screws 4.5 turns CW.
2018-12-18: Got 107 tips and adjusted screws 1 turn CCW.
LI-7500A 75H-2668 2018-10-18 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed on Bouldin Corn to replace sn 75H-2182 for calibration
LI-7500A 75H-2182 2018-10-18 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 75H-2668.
TE525MM 65548-615 2018-10-18 Bouldin Corn Precipitation Calibration Bouldin Corn Precipitation Got 94 of 100 tips. Adjusted a total of 1 turn CW.

2018-10-18: Got 94 of 100 tips - adjusted screws 0.5 turns CW
2018-10-25: Got 97 tips and adjusted set screws 0.5 turn CW.
2018-11-01: 99 of 100 tips - no adjustment.
LI-7500A 0418 2018-10-17 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on Vaira to replace sn 0065 that had no CO2, H2O or signal strength.
LI-7500A 0065 2018-10-17 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira due to no CO2, H2O or Signal Strength - maybe bad source or detector. Replaced by sn 0418
TR-525I-R1 24875-999 2018-10-17 Vaira Precipitation Calibration Vaira Precipitation Got 78 of 100 tips - adjusted screws 3 turns CW
TE525 48132-611 2018-10-17 Tonzi Calibration Tonzi Got 134 of 100 tips - adjusted screws 4 turns CCW
TE525 48130-611 2018-10-17 Tonzi Calibration Tonzi Got 88 of 100 tips - adjusted screws 2.5 turns CW
TR-525I-R1 24872-999 2018-10-17 Tonzi Calibration Tonzi Got 88 of 100 tips - adjusted screws 2.5 turns CW
LI-7500A 75H-2669 2018-10-10 Lab none Calibration Lab none Last calibration: 2018-10-09
AGC=102 SDM=0 DAC1=0 DAC2=0
Ambient P=99.9kPa, Tair=26.6C, Tdew=12.9C
Zero Air at 1400psi, AMP Air 408.97 at 1100 Tdewspan=16.00
CO2 zero 0.918 -> .9188 Offset: 3.4ppm
CO2 span 1.0125 -> 1.0106 Offset: .9ppm
H2O zero 0.8989 -> .8992 Offset: 1.6ppm
H2O span 1.0049 -> 1.0034 Offset: .04ppm
Notes: Re-calibrated to check Katrina's calibration yesterday, which looks good.
LI-7500A 75H-2669 2018-10-09 Lab none Calibration Lab none Last calibration: 2018-06-25
AGC=103 SDM=0 DAC1=0 DAC2=0
Ambient P=99.88kPa, Tair=27.1C, Tdew=12.9C
Zero Air at 1425psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.9182 -> .9180 Offset: -.8ppm
CO2 span 1.0125 -> 1.0125 Offset: 2.2ppm
H2O zero 0.8994 -> .8989 Offset: 52.7ppm
H2O span 1.0062 -> 1.0049 Offset: 16.02ppm
Notes: Calibrated by Katrina. The H2O offset is large, but the change is coefficients is small, so perhaps Katrina wrote down the wrong offset.
LI-7500A 75H-2667 2018-10-10 Lab none Calibration Lab none Last calibration: 2018-10-09
AGC=104 SDM=0 DAC1=0 DAC2=0
Ambient P=99.9kPa, Tair=16.7C, Tdew=11.5C
Zero Air at 1400psi, AMP Air 408.97 at 1100 Tdewspan=16.00
CO2 zero 0.9413 -> .9412 Offset: -.5ppm
CO2 span 1.0127 -> 1.0145 Offset: -1.57ppm
H2O zero 0.8362 -> .8367 Offset: 3.3ppm
H2O span 1.0328 -> 1.0218 Offset: .22ppm
Notes: Re-calibrated to check Katrina's calibration yesterday, which looks good.
LI-7500A 75H-2667 2018-10-09 Lab none Calibration Lab none Last calibration: 2018-06-25
AGC=104 SDM=0 DAC1=0 DAC2=0
Ambient P=99.9kPa, Tair=27.7C, Tdew=9.5C
Zero Air at 1450psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.9414 -> .9413 Offset: -.42ppm
CO2 span 1.0162 -> 1.0127 Offset: -87.77ppm
H2O zero 0.8375 -> .8362 Offset: -7.4ppm
H2O span 1.0237 -> 1.0328 Offset: -.19ppm
Notes: Calibrated by Katrina. The CO2 span offset is large, but the change is coefficients is small, so perhaps Katrina wrote down the wrong offset.
LI-7500A 0418 2018-10-10 Lab none Calibration Lab none Last calibration: 2018-04-13
AGC=93.2 SDM=0 DAC1=0 DAC2=0
Ambient P=99.98kPa, Tair=29.11C, Tdew=15.00C
Zero Air at 1500psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.9001 -> .8982 Offset: -7.43ppm
CO2 span 1.0057 -> .9965 Offset: 4.86ppm
H2O zero 0.8426 -> .8454 Offset: 16.8ppm
H2O span 0.9892 -> .9813 Offset: 0.16ppm
Notes: calibrated by Katrina
LI-7500A 75H-2668 2018-10-08 Lab none Calibration Lab none Last calibration: 2018-10-02
AGC=101 SDM=0 DAC1=0 DAC2=0
Ambient P=99.9kPa, Tair=29.3C, Tdew=5.8C
Zero Air at 1500psi, AMP Air 408.97 at 1200 Tdewspan=16.00
CO2 zero 0.9389 -> .9394 Offset: 2.06ppm
CO2 span 1.0172 -> 1.0164 Offset: .38ppm
H2O zero 0.9072 -> .9057 Offset: -8.2ppm
H2O span 1.0238 -> 1.0157 Offset: .17ppm
Notes: Second calibration since chemical swap. Sensor had been running for 5 days, since 10/3 when it had a huge H2O zero offset when analyzing zero gas. I'm not sure why it was reading weird values on 10/3, but the numbers seem ok today, consistent with a chemical swap. In the future, we should let the sensor run for 1 week after swapping internal chemicals, rather than calibrating after only 24 hours.
LI-7500A 75H-2180 2018-10-04 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed on Sherman Wetland to replace sn 75H-2667 for calibration.
LI-7500A 75H-2667 2018-10-04 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed from Sherman Wetland for calibration. Replaced by sn 75H-2180.
LI-7500A 0035 2018-10-04 Lab none Installed East Pond Gas Analyzer CO2/H2O Installed on the East Pond tower to replace sn 0418 for calibration.
LI-7500A 0418 2018-10-04 East Pond Gas Analyzer CO2/H2O Removed Lab none Removed from the East Pond tower for calibration. Replaced by sn 0035.
Mark I A001911 2018-10-04 Lab none Installed Bouldin Corn Weather Station A Installed on Bouldin Corn to replace sn A000251 which is to be sent back to Arable for inspection of backup temperature sensor. Later we realized we had sent back the incorrect Arable; they had requested sn A000591.
LI-7500A 1594 2018-10-04 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed on the Mayberry tower to replace sn 75H-2669 for calibration.
LI-7500A 75H-2669 2018-10-04 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry for calibration. Replaced by sn 1594
LI-7500A 75H-2668 2018-10-03 Lab none Tested Lab none Tested with zero air and the most recent configuration file from 2018-10-02, but CO2 was reading 7.2 umol/mol and H2O was reading -50 mmol/m3. I'm going to let it run a few more days before checking again.
LI-7500A 75H-2668 2018-10-02 Lab none Calibration Lab none Last calibration: 2018-09-26
AGC=101 SDM=0 DAC1=0 DAC2=0
Ambient P=99.9kPa, Tair=26.8C, Tdew=15.9C
Zero Air at 1600psi, AMP Air 408.97 at 1200 Tdewspan=20.00
CO2 zero 0.9392 -> .9389 Offset: -1.47ppm
CO2 span 1.0153 -> 1.0172 Offset: -1.09ppm
H2O zero 0.9056 -> .9072 Offset: 9.2ppm
H2O span 1.0166 -> 1.0238 Offset: -.16ppm
Notes: First calibration after swapping out internal chemicals. Kept the sensor on for 3 days after the swap before the calibration.
LI-7500A 1594 2018-10-03 Lab none Calibration Lab none Last calibration: 2018-06-25
AGC=98.6 SDM=0 DAC1=0 DAC2=0
Ambient P=100.16kPa, Tair=28.4C, Tdew=15.18C
Zero Air at 1500psi, AMP Air 408.97 at 1200 Tdewspan=20.00
CO2 zero 0.8648 -> .8644 Offset: -1.74ppm
CO2 span 1.0181 -> 1.0150 Offset: 1.68ppm
H2O zero 0.8869 -> .8870 Offset: 0.8ppm
H2O span 1.0106 -> 1.0056 Offset: .11ppm
Notes: Spanned H2O at 20.00 because initial dew point temperature in the room was > 16.00 degC.
LI-7500A 0035 2018-10-02 Lab none Calibration Lab none Last calibration: 2018-06-25
AGC=98.9 SDM=0 DAC1=0 DAC2=0
Ambient P=99.95kPa, Tair=26.66C, Tdew=16.36C
Zero Air at 1550psi, AMP Air 408.96 at 1200 Tdewspan=20.00
CO2 zero 0.9374 -> .9375 Offset: .45ppm
CO2 span 1.0296 -> 1.0142 Offset: 1.57ppm
H2O zero 0.8979 -> .8981 Offset: 2.3ppm
H2O span 1.0261 -> 1.0277 Offset: -.04ppm
Notes: Spanned H2O at 20.00 because initial dew point temperature in the room was > 16.00 degC.
LI-7500A 75H-2180 2018-10-02 Lab none Calibration Lab none Last calibration: 2018-06-25
AGC=98.6 SDM=0 DAC1=0 DAC2=0
Ambient P=99.86kPa, Tair=27.09C, Tdew=17.03C
Zero Air at 1500psi, AMP Air 408.96 at 1200 Tdewspan=20.00
CO2 zero 0.9141 -> .9137 Offset: -1.7ppm
CO2 span 1.0267 -> 1.023 Offset: 1.86ppm
H2O zero 0.8865 -> .883 Offset: 11.6ppm
H2O span 1.0274 -> .9893 Offset: 4.48ppm
Notes: Calibrated by Katrina. Spanned H2O at 20.00 because initial dew point temperature in the room was > 16.00 degC.