Equipment and Calibration

View History

Equipment Entry

Instrument:Precipitation
Manufacture:Texas Electronics
Model No:TR-525I-R1
Serial No:24875-999
Tag No:
Acquisition Date:0000-00-00
Project:
Location on 2021-10-06:Vaira,
Notes:

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
5482 2021-10-06 Vaira Precipitation Calibration Vaira Precipitation Tested using our new method with the peristaltic pump and 237ml of water which should result in 50 tips. I got 54 tips. I adjusted the screws two full turns CCW which should decrease the number of tips by 8-12% (4-6 tips).
4123 2019-10-24 Vaira Precipitation Calibration Vaira Precipitation I connected the tipping bucket to a CR23x and dripped 475ml of water through the bucket which should result in 100tips. We got 103 tips. I did not adjust it.
3739 2018-10-17 Vaira Precipitation Calibration Vaira Precipitation Got 78 of 100 tips - adjusted screws 3 turns CW
2309 2010-09-10 Lab none Installed Vaira Precipitation Reinstalled on the Vaira tower after testing in the lab. The testing revealed that the problem was a bug in the data logger program that was recording instantaneous values instead of 30min averages.
205 2010-09-09 Lab none Calibration Lab none During the troubleshooting, I ran 16oz (473.3ml) of water through the tipping bucket. According to the manual, this should have resulted in 100 tips. Instead I got 88 tips. I then ran 23.7ml of water that should have resulted in 5 tips and got about 4.5. I then turned the adjustment screws on the bottom two full turns clockwise and got better than five tips. I backed off the screws counter-clockwise a half turn. Now running 473.3ml of water through the tipping bucket results in 99 tips.
204 2010-09-07 Lab none Troubleshooting Lab none Troubleshooting the bad final storage values led to the discovery of a bug in the Campbell program on the data logger. If the Totalize function is burried in a "If time is" function, it doesn't work. The follow instructions do not work.

58: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 1 Interval (same units as above)
3: 30 Then Do

59: Do (P86)
1: 10 Set Output Flag High (Flag 0)

60: Totalize (P72)^12888
1: 1 Reps
2: 57 Loc [ Precip_mm ]

61: End (P95)

Instead the code should read:

58: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 1 Interval (same units as above)
3: 10 Then Do

59: Totalize (P72)^12888
1: 1 Reps
2: 57 Loc [ Precip_mm ]
203 2010-09-02 Vaira Precipitation Removed Lab none This tipping bucket was removed from the Vaira tower to troubleshoot a problem that was causing the total number of tips not to be saved in the final storage location.
 
 