Equipment and Calibration History

View Equipment |

View history entry 3735 for Licor LI-7500A sn: 75H-2669

Date: 2018-10-10
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-10-09 AGC=102 SDM=0 DAC1=0 DAC2=0 Ambient P=99.9kPa, Tair=26.6C, Tdew=12.9C Zero Air at 1400psi, AMP Air 408.97 at 1100 Tdewspan=16.00 CO2 zero 0.918 -> .9188 Offset: 3.4ppm CO2 span 1.0125 -> 1.0106 Offset: .9ppm H2O zero 0.8989 -> .8992 Offset: 1.6ppm H2O span 1.0049 -> 1.0034 Offset: .04ppm Notes: Re-calibrated to check Katrina's calibration yesterday, which looks good.
Entered by: