Equipment and Calibration

View History

Equipment Entry

Instrument:RH & Temp
Manufacture:Vaisala
Model No:HMP60
Serial No:L1740420
Tag No:NULL
Acquisition Date:2015-09-28
Project:CDFW
Location on 2019-01-31:East End, NULL
Notes: NULL

Documents

No documents found

Report: HMP RH and AirT Comparison 12-2017 - 2017-12-18

History

Rec No.DateLoc FromTitleLoc ToNotes
3866 2019-01-31 Lab none Installed East End RH & Temp Installed at EE to replace HMP45 sn U3030046, which was removed for bad RH readings.
3801 2018-11-14 Factory Received Lab none Received back from Campbell Sci. They tested it against their reference for temp and RH at 90%, 50%, and 20% and could not find any issues. We should test it again in the lab.
3798 2018-09-14 Lab none Shipped Factory Shipped back to Campbell Sci for testing (drifting downward).
3499 2018-06-22 Lab none Tested Lab none Temperature reading funky when tested in the lab while testing a CR1000 set up for Sherman Barn site. This HMP60 read 18C, and an HMP45 read 24C, which makes more sense. Current outside temperature is 22-24C.
3527 2018-06-13 Twitchell Alfalfa RH & Temp Removed Lab none Removed with Twitchell Alfalfa site decommissioning.
3264 2017-12-19 Twitchell Alfalfa RH & Temp Moved Twitchell Alfalfa RH & Temp Moved back up to its usual position near the 7500 from the HMP inter-comparison height.
3269 2017-12-18 Twitchell Alfalfa 0 Moved Twitchell Alfalfa 0 Installed at Twitchell Alfalfa for an inter-comparison of HMPs
2627 2017-02-14 Lab none Installed Twitchell Alfalfa RH & Temp Installed at Twitchell Alfalfa
 
 