Date: |
2018-10-29 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Upgrade |
Notes: |
I swapped out power board pn 9677-073G per Licor's instructions. The new power board is pn 9977-079. According to Licor the original boards Rev 0, could be damaged by exposure to voltage above 24VDC. The new board Rev 0A (V3) should not have that problem.
Note: This instrument was replaced with 2 original screws and 2 long screws. The long screws were used to help re-assemble the instrument, and I forgot to swap them for the original, shorter, screws. |
Entered by: |
Daphne Szutu |
|
|