View history entry 3760 for Licor LI-7700 sn: TG1-0224

Date: 2018-10-25
From: West Pond - Gas Analyzer CH4
To: West Pond - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: West Pond Last calibration: 2018-04-04 RSSI=70%, Optical RH=22% Ambient P=101.7kPa, Tair=25.5C Zero Air at 1950psi, AMP Air 3.121 at 2100 CH4 Zero Offset: -.004ppm CH4 Span Offset: -.002ppm Notes: Calm day. Before calibrating, we flushed head spaced with zero air. Optics RH dried down from 22% to 13% before calibration started.
Entered by:
Field notes