Equipment and Calibration History

View Equipment |

View history entry 3755 for Licor LI-7700 sn: TG1-0424

Date: 2018-10-22
From: Lab - none
To: Lab - none
Title: Upgrade
Notes: I swapped out power board pn 9677-079 per Licor's instructions. According to Licor the original boards Rev 0, could be damaged by exposure to voltage above 24VDC. The new board Rev 0A (V3) should not have that problem.
Entered by: Joe Verfaillie