Equipment and Calibration History

View Equipment |

View history entry 3758 for Licor LI-7500A sn: 75H-2182

Date: 2018-10-24
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-07-10 AGC=87.3 SDM=0 DAC1=0 DAC2=0 Ambient P=100.55kPa, Tair=26.87C, Tdew=12.61C Zero Air at 1500psi, AMP Air 408.97 at 1100 Tdewspan=16.00 CO2 zero 0.8841 -> .8844 Offset: 1.11ppm CO2 span 1.0168 -> 1.0188 Offset: -1.04ppm H2O zero 0.845 -> .8449 Offset: -.6ppm H2O span 1.0202 -> 1.0143 Offset: .13ppm Notes:
Entered by: