Equipment and CalibrationView History
Equipment Entry
Instrument: | CO2 Probe |
Manufacture: | Forerunner-Eosense |
Model No: | GP-001 |
Serial No: | GP20140018 |
Tag No: | NULL |
Acquisition Date: | 2017-05-31 |
Project: | NULL |
Location on 2020-10-08: | Factory, NULL |
Notes: |
Provided by Eosense to replace one of the 8 sensors they disposed of at the factory. dual range 0-2% and 0-6% |
|
|
Documents
Calibration: cal_Eosense_GP-001_snVarious_20170526.pdf
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
4854 |
2020-10-08 |
Lab none |
Shipped |
Factory none |
Sent to Eosense for inspection |
3794 |
2018-11-06 |
Lab none |
Tested |
Lab none |
Connected to a CR6 and recorded 10s data. The sensor was maxing out (reading ~5V) on both 0-2% and 0-6% ranges. The connection between sensor and cable seemed secure and clean. The sensor should go back to the factory. |
3790 |
2018-10-25 |
West Pond CO2 Probe B |
Removed |
Lab none |
Removed from WP because of high readings. It was replaced by sn GP20140026 two weeks later. |
3161 |
2017-08-16 |
Lab none |
Installed |
West Pond CO2 Probe B |
Installed at WP to replace a previous bad eosGP |
|
|
|