Equipment and Calibration

View History

Equipment Entry

Instrument:Narrow Band
Manufacture:Decagon-METER
Model No:SRS-Ni NDVI
Serial No:2046503376
Tag No:
Acquisition Date:2017-06-03
Project:Kearney
Location on 2023-09-18:Loaned,
Notes: Cosine correcting Teflon diffuser, hemispherical field of view

Documents

No documents found

Report: Mayberry - NDVI (SRS, Planet) - 2017-06-06

History

Rec No.DateLoc FromTitleLoc ToNotes
6416 2023-09-18 Lab none Loaned Loaned none Loaned to Gracie Pearsall at UCSC for a informational display at Elkhorn Reserve
6179 2023-01-25 Lab none Tested Disposed none Removed the spliced cable and tested in the lab. Still no response on the terminal emulator. This sensor is probably dead. It's weird that both incoming and outgoing sensors went bad at the the same time. Both sensors had spliced cables and were connected to the same C port and the same pluggable power block. One of the spliced wires for data (red) had a few cracks in the plastic sheath; maybe it shorted against the bare G wire and killed both sensors.
6172 2023-01-18 Mayberry Narrow Band NDVI Incoming Removed Lab none Removed from MB for testing in the lab. It was not responding on the terminal emulator.
2826 2017-06-05 Lab none Installed Mayberry Narrow Band NDVI Incoming Installed in incoming position at Mayberry.
2987 2017-06-04 Lab none Maintenance Lab none The SDI-12 address was set to E indicating a hemispherical outgoing orientation and the head was labeled with sn, address, and 'out'.
 
 