Equipment and Calibration

View History

Equipment Entry

Instrument:Interface
Manufacture:YSI
Model No:EXO2
Serial No:22C103416
Tag No:
Acquisition Date:2022-04-29
Project:DeltaCouncil-Dutch
Location on 2024-04-10:Hill Slough,
Notes: Part number 599502-01, seven ports, 4D cell batteries, 10m depth. Integrated non-vented depth sensor sn 22A107586.

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
6647 2024-04-10 Lab none Installed Hill Slough Interface EXO Reinstalled at the HS channel after regular lab cleaning and calibration
6639 2024-04-04 Hill Slough Water Level Removed Lab none Removed from the Hill Slough channel and brought back to the lab for calibration (and to install the new auto wiper that arrived)
6600 2024-03-06 Factory none Received Lab none Received wiper sn 13A100277 (firmware 1.0.86) from YSI after repair and calibration. Came with an extra brush and a small allen key for installing the brush. The factory inspection reported that water had leaked into the probe housing and all internal components were corroded. They replaced the wiper under warranty. YSI recommends sending the wiper back to the factory every year to have wiper motor o-rings replaced and lubed to prevent water leakage.
6617 2024-01-25 Lab none Installed Hill Slough Water Level Installed in the Hill Slough channel after calibration
6555 2024-01-23 Lab none Tested Lab none The wiper hasn't worked properly since being installed in mid-December. When we pulled the sonde last week, the brush was parked over the pH sensor and all the sensors were dirty. I brought the sonde back to the lab for routine calibration and tested the brush. The brush did work smoothly as first, but then it became intermittent. Sometimes it didn't work at all and sometimes it worked only in spurts (and the EXO would disconnect from the laptop right before the brush moved). The wiper did not show up under the Sensors or Calibration tab. Nothing changed when I tried plugging the wiper into a different port. I talked to Matt at YSI who thinks something is messed up electronically inside the wiper--it can receive commands (to wipe) but is not talking back to the sonde. I will send the wiper back to YSI for repair.
6553 2024-01-22 Lab none Calibration Lab none Depth(m),Last cal: 2023-10-18
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.1.3
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.000 m,-0.076 m,0.000 m, 0.000 m, 17.888 °C
6542 2024-01-18 Hill Slough Water Level Removed Lab none Removed from HS channel and brought back to the lab for routine calibration
6446 2023-10-18 Lab none Installed Hill Slough Water Level Installed in the Hill Slough channel after calibration in the lab. The whole sonde was raised about 20cm to keep it out of the mud.
6437 2023-10-11 Hill Slough Removed Lab none Removed for inspection in the lab because turbidy/fDOM/ph/ORP look weird
6390 2023-08-24 Lab none Installed Hill Slough Water Level Reinstalled after quarterly cleaning and calibration
6377 2023-08-17 Lab none Calibration Lab none Depth(m),Last cal: 5/8/2023 15:15:13
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.1.3
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.000 m,-0.074 m,0.000 m, 0.000 m, 19.959 °C
6369 2023-05-09 Lab none Installed Hill Slough Water Level Installed in the channel at Hill Slough (same height as before)
6279 2023-04-25 Hill Slough Removed Lab none Removed from HS channel for routine calibration
6155 2023-01-06 Lab none Installed Hill Slough Water Level Installed in the channel at Hill Slough (same height as before) after calibrating in the lab yesterday. Installed without batteries in the sonde.
6150 2023-01-03 Hill Slough Removed Lab none Removed from Hill Slough channel and brought back to the lab for calibration.
6554 2022-09-29 Lab none Calibration Lab none Depth (m), Last cal: 9/23/2022 11:38:38 AM
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.1.3
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.000 m,0.109 m,0.000 m,0.000 m,22.000 °C
6005 2022-09-29 Lab none Installed Hill Slough Water Level Installed in the channel at Hill Slough. The sonde is now wired to the CR23X, which queries it for data every 10sec. The sonde is measureing and storing data internally every 5min.
6053 2022-09-23 Lab none Calibration Lab none Depth (m), Last cal: 6/27/2022 1:49:10 PM
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.1.3
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.000 m,-0.008 m,0.000 m,0.000 m,22.000 °C
6002 2022-09-16 Hill Slough Removed Lab none Removed from the channel at Hill Slough. It was brought back to the lab for cleaning, calibration, and to convert its output to SDI-12 so we can wire it to the channel CR23X.
5921 2022-07-13 Lab none Installed Hill Slough Water Level Installed at HS in the channel off of the boardwalk, about 30m east of the tower. We installed the sonde near low tide to ensure it would stay underwater at all times.
6051 2022-06-27 Lab none Calibration Lab none Depth (m), Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.1.3
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.000 m,10.300 m,0.000 m,0.000 m,18.800 °C