Equipment and Calibration History

View Equipment |

View history entry 6424 for Licor LI-7500A sn: 1594

Date: 2023-10-03
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2023-06-12 AGC=100.7 SDM= DAC1= DAC2= AIU sn AIU-1569, Presure A0=58.004, A1=15.479 Ambient P=100.4kPa, Tair=23.5C, Tdew=13.3C Zero Air at 750psi, AMP CA04119 524.77 at 700 Tdewspan=20.00 CO2 zero 0.8777 -> .8777 Offset: .4ppm CO2 span 1.0098 -> 1.0092 Offset: .31ppm H2O zero 0.811 -> .8117 Offset: 4.0ppm H2O span 1.0126 -> 1.0074 Offset: .12ppm Notes: Vertical orientation
Entered by: