Equipment and Calibration

View History

Equipment Entry

Instrument:Conductivity Sensor
Manufacture:YSI
Model No:EXO Wiped Cond/Temp
Serial No:23D104030
Tag No:
Acquisition Date:2023-04-04
Project:UCSC
Location on 2024-04-10:Hill Slough,
Notes: Wiped conductivity/temperature sensor for EXO sensor. Came with an extra wiper brush, spacing o-rings, and zip ties.

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
6650 2024-04-10 Lab none Installed Hill Slough Conductivity Sensor Reinstalled at the HS channel after regular lab cleaning and calibration
6697 2024-04-08 Lab none Calibration Lab none Sp Cond (uS/cm),Last cal: 2024-01-22
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.0.5
CellConstant 0.48
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,5000.0 µS/cm,4849.8 µS/cm,5000.0 µS/cm, 0.0 µS/cm,21.490 °C
Note: Used USGS standard from Fondriest that expired Aug 2023. It was stored in fridge w/ cap taped and I brought it to room temperature before calibration.
6642 2024-04-04 Hill Slough Conductivity Sensor Removed Lab none Removed from the Hill Slough channel and brought back to the lab for calibration (and to install the new auto wiper that arrived)
6620 2024-01-25 Lab none Installed Hill Slough Conductivity Sensor Installed in the Hill Slough channel after calibration
6550 2024-01-22 Lab none Calibration Lab none Sp Cond (uS/cm),Last cal: 2023-08-17
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.0.5
CellConstant 0.46
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,8420.0 µS/cm,8507.6 µS/cm,8420.0 µS/cm, 0.0 µS/cm,20.226 °C
Note: Used 10,000 uS/cm standard from fridge (brought it to room temperature first). Measured temp-compensated conductivity with handheld YSI probe
6545 2024-01-18 Hill Slough Conductivity Sensor Removed Lab none Removed from HS channel and brought back to the lab for routine calibration
6449 2023-10-18 Lab none Installed Hill Slough Conductivity Sensor Installed in the Hill Slough channel after calibration in the lab. The whole sonde was raised about 20cm to keep it out of the mud.
6440 2023-10-11 Hill Slough Conductivity Sensor Removed Lab none Removed for inspection in the lab because turbidy/fDOM/ph/ORP look weird
6393 2023-08-24 Lab none Installed Hill Slough Conductivity Sensor Reinstalled after quarterly cleaning and calibration
6372 2023-08-17 Lab none Calibration Lab none Sp Cond (uS/cm),Last cal: 8/17/2023 13:16:20
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.0.5
CellConstant 0.46
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,974.0 µS/cm,970.4 µS/cm,974.0 µS/cm, 0.0 µS/cm,24.379 °C
Note: Checked calibration from earlier today. Calibration looks good and value stabilized quickly. 10,000 uS/cm standard from fridge and temp-compensated conductivity measured with handheld YSI probe. (Note 2024-01-24: Probably used 1,000 uS/cm standard, not 10,000 uS/cm)
6371 2023-08-17 Lab none Calibration Lab none Sp Cond (uS/cm),Last cal: 5/5/2023 13:09:53
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.0.5
CellConstant 0.46
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,958.0 µS/cm,967.3 µS/cm,958.0 µS/cm, 0.0 µS/cm,19.770 °C
Note: 10,000 uS/cm standard from fridge and temp-compensated conductivity measured with handheld YSI probe (Note 2024-01-24: Probably used 1,000 uS/cm standard, not 10,000 uS/cm)
6370 2023-05-09 Lab none Installed Hill Slough Conductivity Sensor Installed in the channel at Hill Slough (same height as before) after calibration