Date: 2023-09-05
From: Lab -
To: Lab -
Title: Calibration
Notes: Calibration Location: Lab Last calibration: 2023-01-19 RSSI=80%, Optical RH=23% Ambient P=100.0kPa, Tair=25.0C Zero Air at 800psi, AMP CA04119 4.2238 at 850 CH4 Zero Offset: .028ppm CH4 Span Offset: -.0458ppm Notes: Calibrated zero on 2023-09-05 and span on 2023-09-07 (P=100.3 kPa, T=24.1C). After calibration, swapped drierite and flushed headspace with zero air until optics RH=10%.