Equipment and Calibration

View History

Equipment Entry

Instrument:Anemometer
Manufacture:Gill
Model No:WM 1590
Serial No:153804
Tag No:NULL
Acquisition Date:2015-09-24
Project:CDFW
Location on 2023-11-07:Lab, NULL
Notes: 4x14-bit analog outputs, 20Hz, Firmware version 2326-701-01

Documents

No documents found

Report: Sonic Comparison - 2022-04-07

History

Rec No.DateLoc FromTitleLoc ToNotes
6462 2023-11-07 Loaned none Returned Lab none Returned from Yuankun Xu (Roland Burgmann Lab) and Yifang Cheng at the conclusion of their field project at the Slumgullion Landslide site in Colorado.
6379 2023-08-22 Lab none Loaned Loaned none Loaned to Yuankun Xu (Roland Burgmann Lab) and Yifang Cheng (Berkeley Seismo Lab) for use at Slumgullion Landslide site in Colorado for about one month mid-Sept to mid-Oct. Loaned with sonic sn 152603 and CR6s sn 2114 and sn 2115.
6106 2022-11-30 Bouldin Alfalfa-ADEX Anemometer Bouncer Lower Removed Lab none Removed from the main tower and brought back to the lab at the end of the ADEX intercomparison
5841 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Anemometer Bouncer Lower Installed on a cross arm at the Bouncer tower at 2.42m alg for the advection experiment at BA
5771 2022-04-06 Lab none Tested Lab none I tested this sensor briefly in the lab against sn 152601. U and V had a much smaller magnitude. I checked the analog settings and they had been set to 5m/s range. I changed U and V from 5m/s to 30m/s. I left W as 5m/s. U, V, W, SOS seem reasonable. I will be moving this intercomparison to my backyard tomorrow to run for a few days outdoors.
5660 2022-01-13 Lab none Tested Lab none This sonic was giving bad SOS analog output at Mayberry after a bunch or messy power cycles. Today in the lab the digital SOS was 350m/s at 34% humidity and 102.5kPa and 25C. Sonic temp reading about 5C high - seem too much with high pressure and low humidity. I check the settings and the fourth analog output that should be SOS was set to Status (good is high, bad is low) which makes sense with the data we were seeing. I reset the fourth Analog output to SOS, saved the config. All winds were reading zero in the lab. I check the SOS output on the calibration Licor box and it read 350, same as the digital. Might need to do an inter-comparison before I trust the SOS.
5647 2022-01-02 Mayberry Anemometer Removed Lab none Removed from the MB tower and brought back to the lab for testing. The SOS was bad. Before removing, I measured a heading of 26 deg true N on my phone compass.
2597 2017-01-26 Lab none Installed Mayberry Anemometer Installed at Mayberry tower. It replaced sn 074417.
2600 2017-01-25 Lab none Upgrade Lab none Upgraded firmware to correct for W bug.

Old SN: Y153804
New SN: X153804
Upgrade file: Y074417-Y124215-m.gcu
Old firmware: 2329-601-03
New firmware: 2329-701-01

All Checksum fields report PASS.

Log file: .txt

Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting.

-------

THE FOLLOWING FIELDS HAVE CHANGED

Output rate : P1 1 Hz
Output rate : P8 20 Hz

THE NEW CONFIGURATION IN FULL

==== DETAILS OF DETECTED ANEMOMETER ========================

Serial Number : X153804

---- SW VERSION--------------------------------------------
Anemometer Firmware Version: 2329-701

---- HW DETAILS OF ANEMOMETER------------------------------
Detected Anemometer Class : WindMaster 20Mhz
Number of AXIS on this unit : 3
DAC resolution of attached unit : 14
Number of DAC channels : 4
Maximum sampling rate : (8) P8 20 Hz

---- USER CONFIGURATION OF ANEMOMETER----------------------
Baud Rate : 19200
Output Message Format : M1 UV(W)
Output Message Velocity Units : m/s
Output Message Terminator :
Retries : ON
Instantaneous Sampling : OFF
Factory Calibration : ON
Physical Communications : Auto
Resolution settings : High resolution
Alignment settings : North to spar
Averaging settings : Averaging inactive
Output rate : P1 1 Hz
Output rate : P8 20 Hz
Polar angle minimum magnitude: 0.050 m/s
Power-On Message : Display Power On Message
ASCII Format : Comma Separated (CSV)
SOS and Sonic Temperature display : SOS
Analogue output mode : -5V to 5V
Unit Identifier : 'Q'
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)
Output Channel 4 : Speed of Sound
Synchronised Polling : Off

2582 2017-01-17 East End Anemometer Removed Lab none Returned to lab for firmware upgrade. It was replaced by sn 124215.
2146 2016-01-12 Lab none Installed East End Anemometer Installed at East End, brand new
 
 