Equipment and Calibration

View History

Equipment Entry

Instrument:Forced Diffusion
Manufacture:Forerunner-Eosense
Model No:eosFD
Serial No:FD20180048
Tag No:
Acquisition Date:2019-11-12
Project:Borrowed
Location on 2023-11-13:Lab,
Notes: One of five FD chambers loaned to us from Eosense for experiments. This is the second batch or borrowed chambers.

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
6478 2023-11-13 Tonzi-SoilCO2 Forced Diffusion Open Removed Lab none Removed because CO2 readings seem to be too low. We tried a performance check but missed being able to reset the offset. Brought to the lab for testing.
5924 2022-04-27 Lab none Installed Tonzi-SoilCO2 Forced Diffusion Open Reinstalled at Tree 1 after testing in the lab. I had removed it because I thought it was broken but it might have been reading funny because it was lying on its side. It seemed okay in the lab. I installed it in a slightly different location with more litter on the soil surface.
5786 2022-04-12 Tonzi-SoilCO2 Forced Diffusion Tree 1 Removed Lab none Removed from Tonzi Tree 1 for testing. Identical CO2 readings (zero flux). Chamber was on its side which would cause this symptom - brought in for testing.
5404 2021-07-29 Tonzi-SoilCO2 Forced Diffusion Tree 1 Maintenance Tonzi-SoilCO2 Forced Diffusion Tree 1 I unfastened this chamber and laid it on its side for a performance check.
Average Flux: 0.0, 0.04 STDEV
Average ATM: 242.5ppm, 51.9C
Average Soil: 243.0ppm, 51.7C
I reset the baseline CO2 concentration to 400ppm. This chamber does not have a sun shield and even though it was in the shade under the trees its temperatures were very high.
4933 2020-10-22 Lab none Installed Tonzi-SoilCO2 Forced Diffusion Tree 1 Installed at the under tree location at Tonzi to replace sn FD20180068 which has been reading zero CO2 concentration.
4829 2020-10-05 Bouldin Alfalfa Forced Diffusion A Removed Lab none Removed from Bouldin Alfalfa to check on it's low CO2 concentration readings. Replaced by sn FD20180047
4775 2020-09-23 Bouldin Corn Forced Diffusion B Moved Bouldin Alfalfa Forced Diffusion A Moved from BC to BA to replace sn FD20160001, which was brought back to the lab for testing
4796 2020-09-01 Bouldin Corn Forced Diffusion B Moved Bouldin Corn Forced Diffusion B Moved tower and soil sensors back from field in preparation for corn harvest. Chambers were reinstalled under the tower.
4512 2020-05-29 Lab none Installed Bouldin Corn Forced Diffusion B Reinstalled in normal density corn just in front of tower
4422 2020-04-17 Bouldin Corn Forced Diffusion B Removed Lab none Removed from BC and brought back to the lab because there will be tractor work on the field soon.
4324 2020-01-07 Lab none Installed Bouldin Corn Forced Diffusion B Reinstalled at Bouldin Corn after being removed when the field was flooded. Installed near previous location on 12 inch collars.
4733 2019-12-20 Bouldin Corn Forced Diffusion B Removed Lab none Removed from the field by Tyler due to being partially submerged by winter flooding.
4730 2019-11-14 Lab none Installed Bouldin Corn Forced Diffusion B Installed in the corn field near Tyler's southern soil gas ports