Equipment and Calibration

View History

Equipment Entry

Instrument:Water Level
Manufacture:Campbell Sci.
Model No:CS451
Serial No:70011236
Tag No:
Acquisition Date:2015-09-28
Project:CDFW
Location on 2023-10-27:Lab,
Notes: 0-7.25psi = 0-5.1m of water

Documents

Calibration: Campbell_CS451_70011236.pdf
PackingSlip: ps_Campbell_20150928.pdf
Report: Water Level sensor calibration with staff gauge - 2018-04-24

History

Rec No.DateLoc FromTitleLoc ToNotes
6457 2023-10-27 Mayberry Water Level Removed Lab none Removed from MB because not working
4528 2020-03-11 Mayberry Water Level Notes Mayberry Water Level Before we moved the water level sensor:
Top of well to WL = 71cm
Top of well to probe = 86cm
After we moved the water level sensor:
Top of well to WL = 69cm
Top of well to probe = 91cm (therefore, water level = 22 cm)
Sensor reading = 0.31 psi (x 70.41 psi/cm = 21.8 cm – good match with physical reading!)
Staff gauge = 27cm
3597 2018-07-12 Lab none Installed Mayberry Water Level Installed at MB to replace sn 70011229, which was removed b/c of suspect diurnal fluctuations.
3103 2017-09-26 East End-Temp2 Water Level Removed Lab none Removed from the East End Temporary Tower 2 - Closed Canopy location with the decommissioning of the site.
2867 2017-06-14 Lab none Installed East End-Temp2 Water Level Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up.
2710 2017-04-04 Twitchell Rice Water Level Removed Lab none Removed with Twitchell Rice tower decommissioning.
2326 2016-06-24 Lab none Installed Twitchell Rice Water Level Installed on the Twitchell Rice tower to replace the existing sensor that had a diurnal wave in its data.
1968 2015-09-04 Factory Calibration Factory Initial factory calibration. Recommended interval: 2 years of service.
 
 