Equipment and Calibration

View History

Equipment Entry

Instrument:pH ORP
Manufacture:YSI
Model No:EXO pH-ORP
Serial No:22D102641
Tag No:
Acquisition Date:2022-05-18
Project:Dutch Slough
Location on 2024-04-10:Hill Slough,
Notes: Electrolyte in the sensor module will be depleted in 12-18 months at which point the module will need to be replaced.
Ideally, delta slope > 165 mV, and the mV per decade = 59.16 +/- 5.

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
6651 2024-04-10 Lab none Installed Hill Slough pH-ORP Reinstalled at the HS channel after regular lab cleaning and calibration
6699 2024-04-08 Lab none Calibration Lab none ORP (mV), Last Cal: 2024-01-22
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,226.7 mV,228.5 mV,226.7 mV,0.0 mV,23.000 °C
Note: Used Zobell solution (mixed 2022-09) in small brown plastic bottle. Waited to calibrate until change <0.1mV/10sec (a few min past the point when "Unstable" changes to "Stable").
6698 2024-04-08 Lab none Calibration Lab none pH, Last cal: 2024-01-22
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.01 pH,7.06 pH,7.01 pH,-8.2 mV,22.000 °C
Pt2,10.06 pH,10.15 pH,10.08 pH,-179.0 mV,22.000 °C
Pt3,4.00 pH,4.14 pH,4.02 pH,154.6 mV,22.000 °C
4-7Slope,4-7mVPerDecade,7-10Slope,7-10mVPerDecade
162.8 mV,54.09 mV,170.7mV,55.98
Note: Waited for each point to change <0.1mV/10sec for calibration (a few minutes past point where "Unstable" changes to "Stable"). Used opened standards from May 2023.
6643 2024-04-04 Hill Slough pH-ORP Removed Lab none Removed from the Hill Slough channel and brought back to the lab for calibration (and to install the new auto wiper that arrived)
6621 2024-01-25 Lab none Installed Hill Slough pH-ORP Installed in the Hill Slough channel after calibration
6552 2024-01-22 Lab none Calibration Lab none pH, Last cal: 2023-10-16
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.01 pH,7.01 pH,7.01 pH,-4.8 mV,19.428 °C
Pt2,10.06 pH,10.04 pH,10.06 pH,-174.7 mV,20.116 °C
Pt3,4.00 pH,4.05 pH,4.01 pH,161.2 mV,21.033 °C
4-7Slope,4-7mVPerDecade,7-10Slope,7-10mVPerDecade
166.0 mV,55.16,169.9 mV,55.71
Note: Calibration pretty close and delta slope and mV per decade are ok. Used opened standards from May 2023.
6551 2024-01-22 Lab none Calibration Lab none ORP (mV), Last Cal: 2023-10-16
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,230.4 mV,227.6 mV,223.3 mV,0.0 mV,21.275 °C
Note: Poured Zobell solution (mixed 2022-09) into small brown plastic bottle for this calibration.
6546 2024-01-18 Hill Slough pH-ORP Removed Lab none Removed from HS channel and brought back to the lab for routine calibration
6450 2023-10-18 Lab none Installed Hill Slough pH-ORP Installed in the Hill Slough channel after calibration in the lab. The whole sonde was raised about 20cm to keep it out of the mud.
6441 2023-10-11 Hill Slough pH-ORP Removed Lab none Removed for inspection in the lab because turbidy/fDOM/ph/ORP look weird
6394 2023-08-24 Lab none Installed Hill Slough pH-ORP Reinstalled after quarterly cleaning and calibration
6374 2023-08-17 Lab none Calibration Lab none pH, Last cal: 5/5/2023 15:26
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.00 pH,1.33 pH,7.00 pH,-10.9 mV,23.727 °C
Pt2,10.10 pH,5.28 pH,10.10 pH,-181.5 mV,24.030 °C
Pt3,4.01 pH,-2.66 pH,3.98 pH,163.0 mV,24.696 °C
4-7Slope,4-7mVPerDecade,7-10Slope,7-10mVPerDecade
173.8 mV,58.14,170.7 mV,55.06
Note: Very far off but delta slope and mV per decade are ok. Used opened standards from May 2023.
6373 2023-08-17 Lab none Calibration Lab none ORP (mV), Last Cal: 5/5/2023 15:04
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,223.0 mV,239.5 mV,223.0 mV,0.0 mV,22.877 °C
6442 2023-06-14 Lab none Installed Hill Slough pH-ORP Installed new tip on pH/ORP probe
6278 2023-04-25 Hill Slough pH-ORP Removed Lab none Removed from HS channel for routine calibration
6154 2023-01-06 Lab none Installed Hill Slough pH-ORP Installed in the channel at Hill Slough (same height as before) after calibrating in the lab yesterday.
6149 2023-01-03 Hill Slough pH-ORP Removed Lab none Removed from Hill Slough channel and brought back to the lab for calibration.
6015 2022-09-29 Lab none Installed Hill Slough pH-ORP Channel Installed as part of the EXO2 sonde in the channel at Hill Slough.
6057 2022-09-23 Lab none Calibration Lab none ORP (mV), Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,223.0 mV,220.6 mV,223.0 mV,0.0 mV,25.000 °C
6056 2022-09-23 Lab none Calibration Lab none pH, Last cal: 7/12/2022 2:26:48 PM
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.00 pH,6.96 pH,7.00 pH,-4.1 mV,24.500 °C
Pt2,10.01 pH,10.15 pH,10.01 pH,-181.2 mV,24.500 °C
Pt3,4.01 pH,3.83 pH,4.02 pH,169.1 mV,24.500 °C
6012 2022-09-16 Hill Slough pH-ORP Channel Removed Lab none Removed as part of the EXO2 channel deployed in the channel at HS. Brought back to the lab for cleaning and calibration.
6009 2022-07-13 Lab none Installed Hill Slough pH-ORP Channel Installed as part of the EXO2 sonde at HS in the channel off of the boardwalk, about 30m east of the tower.
6052 2022-07-12 Lab none Calibration Lab none pH, Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
Slope is lower than what it should be for a new sensor,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.00 pH,7.02 pH,7.00 pH,-1.4 mV,20.000 °C
Pt2,10.00 pH,9.89 pH,10.00 pH,-167.8 mV,20.000 °C
Pt3,4.00 pH,4.39 pH,4.02 pH,151.1 mV,20.000 °C