Equipment and Calibration

View History

Equipment Entry

Instrument:Data Logger
Manufacture:Campbell Sci.
Model No:CR6
Serial No:2114
Tag No:
Acquisition Date:2015-09-28
Project:CDFW
Location on 2023-11-07:Lab,
Notes:

Documents

Calibration: cal_Campbell_CR6_sn2114_20150821.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
6463 2023-11-07 Loaned none Returned Lab none Returned from Yuankun Xu (Roland Burgmann Lab) and Yifang Cheng at the conclusion of their field project at the Slumgullion Landslide site in Colorado.
6380 2023-08-22 Lab none Loaned Loaned none Loaned to Yuankun Xu (Roland Burgmann Lab) and Yifang Cheng (Berkeley Seismo Lab) for use at Slumgullion Landslide site in Colorado for about one month mid-Sept to mid-Oct. Loaned with sonics sn 152603 and 153804 and CR6 sn 2115.
5857 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Data Logger Bouncer Flux Installed on the Bouncer tower to collect 20Hz data from the lower sonic and 10sec data from the two RH/Temp sensors for the advection experiment at BA
5812 2022-04-22 Lab none Maintenance Lab none The internal lithium battery was low ~0.9V. I replaced it with a "slightly used" Lithium AA battery that was reading 3.2V on the voltmeter.
3312 2018-01-31 Factory Returned Lab none Returned from Campbell where it was evaluated, updated and returned free of charge.
3311 2018-01-22 Factory Maintenance Factory Campbell says: Tested unit on bench, unit functioned upon receipt. No problems found, but unit is running OS Rev3. This Rev is suspected to cause random freezes that require disconnection of the lithium battery. The issue has been resolved in the latest CR6 OSes. Updated the logger to latest Rev 7.02. Performed calibration, tested unit in the environmental chamber, and performed final bench test. Replaced the internal lithium battery and included a new USB cable.
3300 2018-01-19 Lab none Shipped Factory This CR6 became unresponsive for no know reason - only a green light on the RJ45 jack. It was sent to Campbell on RMA 27680 for evaluation and repair.
3299 2018-01-15 Lab none Damaged Lab none I tried to install this CR6 at Tonzi for the new co2 profile on the Tower but it was non-functional. It tested fine in the lab this morning, but was a brick in the field. The green light on the RJ45 jack comes on but no other indication that it has power.
3172 2017-10-19 East End-Temp1 Data Logger Met Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
2855 2017-06-01 Lab none Installed East End-Temp1 Data Logger Met Installed at the East End Temporary Tower #1 Open Water location with the initial set up of this site.
1963 2015-08-21 Factory Calibration Factory Initial factory calibration. Recommended interval: 3 years of service.
 
 