Equipment and CalibrationView History
Equipment Entry
Instrument: | CO2 Probe |
Manufacture: | Forerunner-Eosense |
Model No: | GP-001 |
Serial No: | GP20140009 |
Tag No: | |
Acquisition Date: | 2014-06-13 |
Project: | USDA |
Location on 2017-05-31: | Disposed, |
Notes: |
0 to 2% range. Replacement for previous Forerunner probes that were not water proof. |
|
|
Documents
Calibration: cal_FR-Eosense_GP001_snGP20140009_20140609.pdf
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
2808 |
2017-05-31 |
Factory |
Disposed |
Disposed |
Disposed of at Eosense factory because it was not repairable |
2666 |
2017-03-20 |
Lab none |
Shipped |
Factory |
Sent to Eosense for evaluation. Bad readings after a total 433 days in use. |
2228 |
2015-09-30 |
Twitchell Alfalfa |
Removed |
Lab none |
This 2% probe was removed from the 2cm location at Alfalfa due to |
1626 |
2014-10-30 |
Lab none |
Installed |
Twitchell Alfalfa |
Installed at -2cm at Alfalfa |
1585 |
2014-10-02 |
West Pond CO2 Probe B |
Removed |
Lab none |
Removed from West Pond due to over range readings. It was replaced by a 0 to 4% probe sn GP20140021. |
1584 |
2014-06-26 |
Lab none |
Installed |
West Pond CO2 Probe B |
This 0 to 2% probe was installed at West Pond to hang off scaffold leg 5 cm below surface. No program change, using existing SE channels from last Forerunner probe. |
|
|
|