Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 [63] 64 65 66 67 68
Ser No.DateLoc FromTitleLoc ToNotes
LI-7500A 0041 2012-06-07 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Swapped off the Sherman tower for calibration. Replaced by sn 0418.
LI-7500A 0418 2012-06-06 Lab none Calibration Lab none Last calibration: 2011-12-15
AGC=51 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100kPa, Tair=24-57C, Tdew=6.06C
zero air at 2000psi, 436.53 at 1600 Tdewspan=
CO2 zero 0.9017 -> 0.9011 Offset: -2.40ppm
CO2 span 1.0066 -> 1.0032 Offset: 1.89ppm
H2O zero 0.8493 -> 0.8456 Offset: -20.4ppm
H2O span 0.9719 -> 0.9835 Offset: -0.26ppm
Notes:
HFP01SC 00244 2012-05-17 Twitchell Rice Heat Flux Plate A Removed Lab none I found this heat flux plate in the field. It had not been removed with the other ground sensors and got cut off by the tractor.
LI-7500A 0035 2012-06-04 Lab none Calibration Lab none Last calibration: 2012-05-17
AGC=50 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.44kPa, Tair=21.9C, Tdew=11.6C
dry air at 1900psi, CO2 436.53 at 1500 Tdewspan=14.48
CO2 zero 0.9387 -> 0.9389 Offset: 1.21ppm
CO2 span 0.901 -> 1.0191 Offset: 63ppm
H2O zero 1.0064 -> 0.8992 Offset: 2.58ppm
H2O span 0.9558 -> 1.0371 Offset: 0.7ppm
Notes:
LI-7500A 0065 2011-03-30 Vaira Gas Analyzer CO2/H2O Removed Lab none Swapped off the Vaira tower for calibration. Replaced by sn 0418.
LI-7500A 0418 2011-03-30 Lab none Installed Vaira Gas Analyzer CO2/H2O Swapped on to the Vaira tower to replace sn 0065 for calibration.
LI-7500A 1594 2012-01-17 Lab none Installed Mayberry Gas Analyzer CO2/H2O Swapped on to the Mayberry tower. sn 0035 removed for calibration.
LI-7500A 1594 2012-01-04 Vaira Gas Analyzer CO2/H2O Removed Lab none Swapped off of Vaira tower for calibration. Replaced with serial number ??.
LI-7500 0976 2011-03-30 Tonzi-Tower Removed Lab none Swapped off on the Tonzi Tower for calibration. Replaced by sn 0041.
LI-7500A 0035 2012-01-17 Mayberry Gas Analyzer CO2/H2O Removed Lab none Swapped off the Mayberry tower for calibration. Replaced by sn 1594.
LI-7700 TG1-0155 2012-05-29 Lab none Shipped Factory Sent to Licor for repair on RMA 169766 - probably loose spars maybe covered by recall.
LI-7700 TG1-0155 2012-05-25 Mayberry-Portable Removed Lab none Removed from portable tower at Mayberry. RSSI is about zero probably because loose spars allowed it to get misaligned.
CS450 20010770 2012-05-21 Lab none Installed DWR Corn Installed at the edge of the field at the DWR Corn site
ML2x 377/009 2012-05-21 Lab none Installed DWR Corn Installed at the edge of the field at the DWR Corn site
ML2x 377/010 2012-05-21 Lab none Installed DWR Corn Installed at the edge of the field at the DWR Corn site
ML2x 377/008 2012-05-21 Lab none Installed DWR Corn Installed at the edge of the field at the DWR Corn site
TR-525M 50229-112 2012-05-21 Lab none Installed DWR Corn Installed at the edge of the field at the DWR Corn site
PQS 1 110339 2012-05-21 Lab none Installed DWR Corn Installed on a temporary boom at the top northwest corner of the scaffolding at the DWR Corn site. This sensor is facing down collecting Reflected PAR
PQS 1 110338 2012-05-21 Lab none Installed DWR Corn Installed on a temporary boom at the top northwest corner of the scaffolding at the DWR Corn site. The sensor is facing upward collecting incoming PAR.
NR01 1570 2012-05-21 Lab none Installed DWR Corn Installed on boom at the top northwest corner of the scaffolding at the DWR Corn site
PTB110 H0160069 2012-05-21 Lab none Installed DWR Corn Installed in data logger box on northwest corner of the scaffolding at the DWR Corn site
HMP60 G4540023 2012-05-21 Lab none Installed DWR Corn Installed on the top southwest corner of the scaffolding at the DWR Corn site
HFP01 6343 2012-05-21 Lab none Installed DWR Corn Installed at the edge of the field at the DWR Corn site
HFP01 6491 2012-05-21 Lab none Installed DWR Corn Installed at the edge of the field at the DWR Corn site
HFP01 6492 2012-05-21 Lab none Installed DWR Corn Installed at the edge of the field at the DWR Corn site
LI-7500A 1594 2012-05-18 Lab none Installed Tonzi-Tower Swapped onto the to tower to replace sn 1594 for calibration
LI-7500A 0418 2012-05-18 Tonzi-Tower Removed Lab none Swapped off of the Tonzi tower for calibration, replaced by sn 1594
LI-7500A 0073 2012-05-18 Vaira Gas Analyzer CO2/H2O Removed Lab none Swapped off the tower for calibration, replaced by sn 0042
LI-7500A 0042 2012-05-18 Lab none Installed Vaira Gas Analyzer CO2/H2O Swapped onto the to tower to replace sn 0073 for calibration
PS-30M 11400072 2012-05-10 Lab none Installed Mayberry-Portable Installedin a box in the truck bed to power the portable tower at Mayberry.
KD140SX-UFBS 11ZV0M0861 2012-05-10 Lab none Installed Mayberry-Portable Installed on the truck to power the portable tower at Mayberry.
CFM100 4993 2012-05-07 Twitchell Rice Removed Lab none Removed from the Twitchell tower. No longer needed - data being collected by Li7550 box
CR1000 23865 2012-05-07 Twitchell Rice Data Logger Eddy Removed Lab none Removed from the tower. No longer needed - data being collected by Li7550 box
LI-7500A 0065 2012-05-07 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed from the tower.  Replaced by the new DWR Li7500A (sn 2150).
WMP 1352 000024 2012-05-07 Twitchell Rice Anemometer Removed Lab none Removed from the tower. Replaced by the new DWR WM (sn ) with analog outs.
HMP45AC U3030046 2012-05-02 Factory Repair Lab none The RH sensor and probe were replaced. A new handle was ordered. The unit was calibrated.
GMP222 X3830001 2012-05-02 Factory Disposed Disposed Filter was cracked, probe completely replace - new probe is serial number H0630070
GMP221 X1520002 2012-05-02 Factory Disposed Disposed Filter was cracked, probe completely replace - new probe is serial number H1310020
GMP221 Y4110008 2012-05-02 Factory Disposed Disposed Filter was cracked, probe completely replace - new probe is serial number H1310008
N940-5APE.W 03387648 2012-05-02 Lab none Installed Sherman Island Installed in the box at Sherman with the LGR sn 08-0114
FMA RMT-200 08-0114 2012-05-02 Lab none Installed Sherman Island Installed in the box at sherman to replace the LI7700. Running on the diaphragm pump.
LI-7500A 75H-2176 2012-05-02 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed on folding boom at the top of the tower.
LI-7550 AIU-0650 2012-05-02 Lab none Installed West Pond Interface LI-7550 Installed on the West Pond tower at mid-level on the scaffolding.
WM 1590 120907 2012-05-02 Lab none Installed West Pond Anemometer Installed on folding boom at the top of the tower.
PS-30M 11400077 2012-05-02 Lab none Installed West Pond Solar Controller Installed to power the site at West Ponds. The charge controller is on a plastic box with three batteries on one level of scaffolding just inside the edge of the east pond.
KD140SX-UFBS 11ZV0M0855 2012-05-02 Lab none Installed West Pond Solar Panel C Installed to power the site at West Ponds. The solar panels are on top of one level of scaffolding just inside the edge of the east pond.
KD140SX-UFBS 11ZV0M0856 2012-05-02 Lab none Installed West Pond Solar Panel B Installed to power the site at West Ponds. The solar panels are on top of one level of scaffolding just inside the edge of the east pond.
KD140SX-UFBS 11ZV0M0835 2012-04-24 Lab none Installed Twitchell Rice Installed to power the site. The solar panels are on the the wooden base to the north of the box. Old panels are disconnected
KD140SX-UFBS 11ZV0M0829 2012-04-24 Lab none Installed Twitchell Rice Installed to power the site. The solar panels are on the the wooden base to the north of the box. Old panels are disconnected
KD140SX-UFBS 11ZV0M0828 2012-04-24 Lab none Installed Twitchell Rice Installed to power the site. The solar panels are on the the wooden base to the north of the box. Old panels are disconnected
HMP45AC U3030046 2012-03-28 Lab none Shipped Factory Sent to Vaisala for repair. Unreasonable analog readings, some corrosion on circuit board.
GMP222 X3830001 2012-03-28 Lab none Shipped Factory Sent to Vaisala for repair. Reporting errors 10, 11 and 12 readings unreasonable.
GMP221 X1520002 2012-03-28 Lab none Shipped Factory Sent to Vaisala for repair. Reporting errors 10, 11 and 12 readings unreasonable.
GMP221 Y4110008 2012-03-28 Lab none Shipped Factory Sent to Vaisala for repair. Reporting errors 10, 11 and 12 readings unreasonable.
LI-7500A 0073 2011-06-21 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Swapped off of the tower for calibration, replaced by sn 1594
LI-7500A 0042 2011-06-21 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Swapped onto the to tower to replace sn 0073 for calibration
LI-7500A 1594 2012-05-17 Lab none Calibration Lab none Last calibration: 2012-01-16
AGC=51 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.19kPa, Tair=23.04C, Tdew=8.4C
dry air at 16psi, CO2 436.53 at 1500 Tdewspan=13.7
CO2 zero 0.8649 -> 0.8642 Offset: 0.5ppm
CO2 span 1.0139 -> 1.0143 Offset: 0.2ppm
H2O zero 0.8951 -> 0.8993 Offset: 2.6ppm
H2O span 1.0063 -> 0.9721 Offset: 0.79ppm
Notes: installed at Tonzi Tower on 5/18/2012 by Siyan
LI-7500A 0042 2012-05-17 Lab none Calibration Lab none Last calibration: 2012-05-16
AGC=54 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.19kPa, Tair=23.04C, Tdew=8.4C
dry air at 16psi, CO2 436.53 at 1500 Tdewspan=13.7
CO2 zero 0.9512 -> 0.9546 Offset: 13.62ppm
CO2 span 0.8486 -> 1.0182 Offset: 17.44ppm
H2O zero 1.0024 -> 0.8346 Offset: -1.89ppm
H2O span 0.999 -> 0.8284 Offset: 5.11ppm
Notes: calibrated in the lab; >10ppm offset although the IRGA was just fixed by Li-cor. Joe called Li-cor, but no clue.
LI-7500A 0035 2012-05-17 Lab none Calibration Lab none Last calibration: 2012-05-16
AGC=50 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.19kPa, Tair=23.04C, Tdew=8.4C
dry air at 16psi, CO2 436.53 at 1500 Tdewspan=13.7
CO2 zero 0.9419 -> 0.9546 Offset: 13.62ppm
CO2 span 0.8842 -> 1.0182 Offset: 17.44ppm
H2O zero 1.0055 -> 0.8346 Offset: -1.89ppm
H2O span 0.999 -> 0.8284 Offset: 5.11ppm
Notes: calibrated in the lab; >10ppm offset although the IRGA was just fixed by Li-cor. Joe called Li-cor, but no clue.
LI-7500A 0035 2012-05-16 Lab none Calibration Lab none Last calibration: 2011-09-14
AGC=50 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.19kPa, Tair=23.04C, Tdew=8.4C
dry air at 16psi, CO2 436.53 at 1500 Tdewspan=13.7
CO2 zero 0.9338 -> 0.9419 Offset: ppm
CO2 span 1.0108 -> 0.8842 Offset: ppm
H2O zero 0.7353 -> 1.0055 Offset: ppm
H2O span 0.9843 -> 0.9990 Offset: ppm
Notes: recently fixed by Li-cor; double check in the lab conditions; no calibration applied yet.
LI-7500A 0042 2012-05-16 Lab none Calibration Lab none Last calibration: 2011-12-13
AGC=54 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.19kPa, Tair=23.04C, Tdew=8.4C
dry air at 16psi, CO2 436.53 at 1500 Tdewspan=13.7
CO2 zero 0.9597 -> 0.9512 Offset: ppm
CO2 span 0.984 -> 0.8486 Offset: ppm
H2O zero 0.7828 -> 1.0024 Offset: ppm
H2O span 0.9341 -> 0.9990 Offset: ppm
Notes: recently fixed by Li-cor
double check in the lab conditions
no calibration applied yet
KD140SX-UFBS 11ZV0M0836 2012-05-02 Lab none Installed West Pond Solar Panel A Installed to power the site at West Ponds. The solar panels are on top of one level of scaffolding just inside the edge of the east pond.
LI-7500A 0035 2012-05-10 Factory Returned Lab none Returned from repair at Licor on RMA 168677. This head went in for a bad choper motor. Because of its age and time and cost of recalibration we had the detector and source replaced as well. New calibrations coefficients (2012-04-30): CO2 Cal: A = 1.56246E2, B = 6.54481E3, C = 4.79309E7, D = -1.55242E10, E = 2.03744E12, XS = 0.0010, Z = -1.00000E-4. H2O Cal: A = 5.38954E3, B = 4.64088E6, C = -5.68551E8, XS = -0.0006, Z = 1.17000E-2. Pressure Cal: A0 = 10.626, A1 = 26.036. Zero/Span: CO2 zero = 0.9419, CO2 span = 1.0053, H2O zero = 0.8842, H2O span = 0.9990
LI-7500A 0035 2012-03-27 Lab none Shipped Factory Shipped to Licor on RMA 168677. The chopper motor needed replacing.
LI-7500A 0042 2012-05-10 Factory Returned Lab none Returned from repair at Licor on RMA 168677. This head went in for a bad source. Because of its age and time and cost of recalibration we had the detector and chopper motor replaced as well. New calibrations coefficients (2012-04-25): CO2 Cal: A = 1.48568E2, B = 4.72425E3, C = 4.31507E7, D = -1.33645E10, E = 1.69762E12, XS = 0.0002, Z = 9.00000E-4. H2O Cal: A = 5.38386E3, B = 3.60338E6, C = 9.95283E7, XS = -0.0002, Z = 1.53000E-2. Pressure Cal: A0 = 10.485, A1 = 26.036. Zero/Span: CO2 zero = 0.9512, CO2 span = 1.0024, H2O zero = 0.8486, H2O span = 0.9990
LI-7500A 0042 2012-03-27 Lab none Shipped Factory Shipped to Licor for repair on RMA 168677. The source was intermittently bad.
Quantum Q29882 2012-05-10 Loaned Returned Lab none Returned
CR10X 20140 2012-05-02 Sherman Island Removed Lab none Removed from the soil CO2 box at Sherman. In trying to get a relay to cycle power on three Vaisala probes this data logger was swapped out. Turns out the problem was with the relay wiring not the data logger.
CR23X 2071 2012-04-27 Lab none Returned Returned Returned to the Goldstein lab after sitting idle in the lab for a month or so.
N940-5APE.W 03387648 2012-04-24 Twitchell Rice Removed Lab none Removed from the box at Twitchell, replced with LI-7700
FMA RMT-200 05-0010 2012-04-24 Twitchell Rice Removed Lab none Removed from the box at Twitchell, replced with LI-7700
278 4505052 2012-04-24 Mayberry Barrometer Rewired Mayberry Barrometer Output removed from CR1000 and swapped over to new LI-7700 sn TG1-0224 analog inputs
HMP45A U2120013 2012-04-24 Mayberry RH & Temp Rewired Mayberry RH & Temp Output removed from CR1000 and swapped over to new LI-7700 sn TG1-0224 analog inputs
NL115 6763 2012-04-24 Mayberry Interface NL115 Removed Lab none Removed from the Mayberry eddy tower. No longer needed
CR1000 38212 2012-04-24 Mayberry Data Logger Eddy Removed Lab none Removed from the Mayberry eddy tower. No longer needed
WMP 1352 000019 2012-04-24 Mayberry Anemometer Removed Lab none Removed from the Mayberry eddy tower. Replaced with WM sonic sn 074417
LI-7500A 1594 2012-04-24 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from the Mayberry eddy tower. Replaced with LI-7500A sn 75H-2182
LI-7700 TG1-0155 2012-04-24 Mayberry Gas Analyzer CH4 Removed Lab none Removed from the Mayberry eddy tower. Replaced with LI-7700 sn TG1-0224
WM 1590 074417 2012-04-24 Lab none Installed Mayberry Anemometer Installed on the Mayberry eddy tower the new Licor GHG system to swap out WMP sonic sn 000019. The old sonic did not have analog output, the new one does.
LI-7550 AIU-0666 2012-04-24 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed on the Mayberry eddy tower to swap out LI-7500 sn 1594
LI-7500A 75H-2182 2012-04-24 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed on the Mayberry eddy tower to swap out LI-7500 sn 1594
LI-7700 TG1-0224 2012-04-24 Lab none Installed Mayberry Gas Analyzer CH4 Installed on the Mayberry eddy tower to swap out LI-7700 sn TG1-0155
AM25T 6768 2012-04-17 Lab none Installed West Pond Multiplexer Thermocouples Installed in box at west pilot pond. Not running because of forgotten power connector.
AM16/32B 11661 2012-04-17 Lab none Installed West Pond Multiplexer Met Installed in box at west pilot pond. Not running because of forgotten power connector.
CR1000 47110 2012-05-21 Lab none Installed DWR Corn Installed with the rest of equipment on the scaffolding at the DWR Corn field. This data logger is collect the 30 minute met data.
A547 3586 2012-04-17 Lab none Installed West Pond Interface A547 Installed in box at west pilot pond. Not running because of forgotten power connector.
CS547A 6096 2012-04-17 Lab none Installed West Pond Conductivity Sensor Installed on post northwest of scaffolding at west pilot pond. Not running because of forgotten power connector.
CS450 20010769 2012-04-17 Lab none Installed West Pond Water Level Installed on post northwest of scaffolding at west pilot pond. Not running because of forgotten power connector.
DO6400/T DO-05 2012-04-17 Lab none Installed West Pond DO Sensor Upper Installed on post northwest of scaffolding at west pilot pond. Not running because of forgotten power connector.
DO6400/T DO-04 2012-04-17 Lab none Installed West Pond DO Sensor Lower Installed on post northwest of scaffolding at west pilot pond. Not running because of forgotten power connector.
PQS 1 110336 2012-04-17 Lab none Installed West Pond PAR Sensor Reflected Installed at top of scaffolding at west pilot pond. Not running because of forgotten power connector. This sensor is pointing down and measuring PARout.
PQS 1 110341 2012-04-17 Lab none Installed West Pond PAR Sensor Incoming Installed at top of scaffolding at west pilot pond. Not running because of forgotten power connector. This sensor is pointing up and measuring PARin.
NR01 1457 2012-04-17 Lab none Installed West Pond Net Radiometer Installed at top of scaffolding at west pilot pond. Not running because of forgotten power connector.
PTB110 H0160076 2012-04-17 Lab none Installed West Pond Barometer Installed in box at west pilot pond. Not running because of forgotten power connector.
HMP60 G4540025 2012-04-17 Lab none Installed West Pond RH & Temp Installed at top of scaffolding at west pilot pond. Not running because of forgotten power connector.
LI-7700 TG1-0155 2011-11-22 Mayberry Gas Analyzer CH4 Calibration Mayberry Gas Analyzer CH4 After cleaning the mirrors the RSSI was 69. We tried to check the calibration of the LI7700. I put the hood on the 7700 and ran zero air through it. I watched the value on the CR1000. It got down to about 10ppb - not as good as I expected. I tried to run a span gas but didn't have the correct regulator.
LI-7700 TG1-0155 2011-06-30 Mayberry Gas Analyzer CH4 Calibration Mayberry Gas Analyzer CH4 I removed the solar shield and installed the calibration hood. I was able to connect to the LI7700 through the LI7550 using Ethernet. I changed the wash cycle from 20sec spray rechecked every 15min to a 10sec spray rechecked every 3hr as suggested by Liukang. I flowed zero air from a small transfer tank at about 400psi. The LI7700 reported 0ppm CH4. I then flowed 1.976ppm secondary standard from the tank that used to be at Sherman. The LI7700 reported 1.976ppm CH4. Because the LI7700 seemed to be right on I did not to a formal calibration.
LI-7700 TG1-0155 2011-02-03 Lab none Installed Mayberry Gas Analyzer CH4 Installed on the Mayberry tower to replace the 7700 that was on loan to us from Licor. Collecting data on the CR1000 by direct ethernet to 7700.
LI-6262 IRG3-1079 2012-01-26 Lab none Calibration Lab none Calibrated in differential mode for check of Frank's new secondary standard. Ambient temperature was 21.349C, pressure was 101.69kPa. Using N2 and god bottle 402.69ppm CO2.
Flowing N2 in both sample and reference
CO2 okay.
H2O okay.
Flowing N2 in reference and god bottel in sample
CO2 = 400.70ppm adjusted to 402.69ppm.
Flowing zero air in reference and new secondary standard in sample, I wrote down about 60 readings and made an average. Frank's secondary standard has 444.44ppm CO2.
FMA RMT-200 08-0114 2012-01-26 Lab none Calibration Lab none Calibrated in the lab with god bottle 1.8978ppm CH4 at 300psi. Last calibration Nov 18, 2011. Before calibration while running the cal gas the LGR read:
12:07PST, 21.34C, 142.93Torr, 15.343usec, 1Hz, 1.889ppm
Afterward calibration it was right on. I check the concentration is Franks new tank: 1.948ppm