Skip to main content
rice field banner biometlab logo
   
   
 
   
   
   
   
   
 
 

Equipment and Calibration

View History

Equipment Entry

Instrument:Anemometer
Manufacture:ATI
Model No:SATI-3K
Serial No:000603
Tag No:
Acquisition Date:2012-09-21
Project:Borrowed
Location on 2013-01-22:Returned,
Notes: Borrowed from Robin in Allen Glodsteins lab for an inter-comparison

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
898 2013-01-22 Lab none Returned Returned Returned to Robin in Goldstein lab
897 2012-12-11 Tonzi-Tower Removed Lab none Removed from the Tonzi tower - end of inter comparison
896 2012-11-15 Lab none Installed Tonzi-Tower Installed on the Tonzi tower just to the east of the Gill sonic. This will be an inter-comparison of a few weeks. Data is being collected in separate tables on the CR1000
847 2012-11-08 West Pond Anemometer ATI Removed Lab none Removed from West Pond after a three week inter-comparison.
826 2012-10-18 Sherman Island Anemometer ATI Installed West Pond Anemometer ATI Removed from the Sherman tower and installed on the West Pond. The Sherman system was returned to configuration before installing the ATI. At West Pond, the ATI was installed northeast of the Gill WMP just west (and upwind) of the LI7500. The data is recorded by a separate CR1000 in the box that Robin supplied with the ATI.
825 2012-09-27 Lab none Installed Sherman Island Installed just to the south of the existing Gill WMP for an inter-comparison. The serial data was added to the CR1000 on the RS-232 port and the program was changed to add a third table to record the ATI data.
2131 2012-09-21 Borrowed Borrowed Lab none Borrowed from Robin in Allen Goldstein's lab for an inter-comparison
 
 

Login