Equipment and Calibration

View History

Equipment Entry

Instrument:Anemometer
Manufacture:Gill
Model No:WM 1590
Serial No:120908
Tag No:NULL
Acquisition Date:2012-03-13
Project:DWR-Wetland
Location on 2024-02-28:Lab, NULL
Notes: 4x14bit analog out, 20Hz, Firmware version 2329-116-01

Documents

PackingSlip: ps_Dynamax_20120313.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
6590 2024-02-28 West Pond Anemometer Removed Lab none Removed from WP as part of its decomissioning
2581 2017-01-17 Lab none Installed West Pond Anemometer Installed on West Pond tower to replace sn 120907.
2572 2017-01-06 Lab none Upgrade Lab none Upgraded firmware to correct for W bug.

Old SN: Y120908
New SN: X120908
Upgrade file: Y074417-Y124215-m.gcu, Y012345-s.gcu, X120908 X124215-m.gcu
Old firmware: 2329-114
New firmware: 2329-116-01

All Checksum fields report PASS.

There are multiple upgrade .gcu files. During the initial upgrade using file "Y074417-Y124215-m.gcu," the sensor was set at 57,600 baud rate. This baud rate caused an error in the upgrade, so the SN reset to Y012345. Gill sent me the second upgrade file, "Y-12345-s.gcu," which reset the sensor to SN X120908. Finally, I used the third upgrade file, "X120908 X124215-m.gcu," to upgrade the firmware.

Log file: WMUpgradeLog_X120908_170106_122011.txt

Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting.

-------

THE FOLLOWING FIELDS HAVE CHANGED

Baud Rate : 19200
Baud Rate : 57600
Output rate : P1 1 Hz
Output rate : P8 20 Hz

THE NEW CONFIGURATION IN FULL

==== DETAILS OF DETECTED ANEMOMETER========================

Serial Number : X120908

---- SW VERSION--------------------------------------------
Anemometer Firmware Version: 2329-116

---- HW DETAILS OF ANEMOMETER------------------------------
Detected Anemometer Class : WindMaster 20Mhz
Number of AXIS on this unit : 3
DAC resolution of attached unit : 14
Number of DAC channels : 4
Maximum sampling rate : (8) P8 20 Hz

---- USER CONFIGURATION OF ANEMOMETER----------------------
Baud Rate : 19200
Baud Rate : 57600
Output Message Format : M1 UV(W)
Output Message Velocity Units : m/s
Output Message Terminator :
Retries : ON
Instantaneous Sampling : OFF
Factory Calibration : ON
Physical Communications : Auto
Resolution settings : High resolution
Alignment settings : North to spar
Averaging settings : Averaging inactive
Output rate : P1 1 Hz
Output rate : P8 20 Hz
Polar angle minimum magnitude: 0.050 m/s
Power-On Message : Display Power On Message
ASCII Format : Comma Separated (CSV)
SOS and Sonic Temperature display : SOS
Analogue output mode : -5V to 5V
Unit Identifier : 'Q'
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)
Output Channel 4 : Speed of Sound
2147 2016-01-12 Twitchell Alfalfa Anemometer Removed Lab none Removed from field for calibration
1006 2013-05-24 Lab none Installed Twitchell Alfalfa Anemometer Installed on the northwest corner of the scaffolding at the Alfalfa site.
855 2012-10-30 Mayberry-Portable Removed Lab none Removed from the portable tower and returned to the lab.
854 2012-05-10 Lab none Installed Mayberry-Portable Initial installation on the Protable Tower at Mayberry.