Equipment and CalibrationView History
Equipment Entry
Instrument: | Heat Flux Plate |
Manufacture: | TE Technology |
Model No: | HP-127-1.0-1.3-71P |
Serial No: | P04 |
Tag No: | |
Acquisition Date: | 2012-11-28 |
Project: | DOE-TCP |
Location on 2015-12-02: | Vaira, |
Notes: |
Potted (water proof) peltier chip |
|
|
Documents
Calibration: Cal_PeltierHFP_20130107_snP04.pdf
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
2104 |
2015-12-02 |
Lab none |
Installed |
Vaira |
Reinstalled at Vaira at -16cm. This was spliced onto the existing cable in the field. |
2113 |
2015-12-01 |
Lab none |
Repair |
Lab none |
The red wire had broken right where it met the plate. I was able solder on a new wire. I covered the wires with heat shrink and painted the connections to the plate. |
2111 |
2015-11-20 |
Vaira |
Removed |
Lab none |
Removed from Vaira at -16cm. Looks like the insulation had cracked and the wire rotted. |
918 |
2013-02-06 |
Lab none |
Installed |
Vaira |
Installed in a profile at -16cm to the east of the Vaira radiation tower |
923 |
2013-01-16 |
Lab none |
Calibration |
Lab none |
Calibrated by Sebastian at University of British Columbia: Multiplier = 106.2(W/m2)/mV |
|
|
|