Equipment and CalibrationView History
Equipment Entry
Instrument: | CO2 Probe |
Manufacture: | Vaisala |
Model No: | GMP221 |
Serial No: | H0960006 |
Tag No: | |
Acquisition Date: | 2012-05-23 |
Project: | USDA |
Location on 2013-01-04: | Disposed, |
Notes: |
Came with cricuit board GMB220 sn H1410158 |
|
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
914 |
2013-01-04 |
Lab none |
Tested |
Disposed |
Upon inspection in the lab. this sensor was found to be severely corroded. However, at the time of inspection the sensor was not damp. |
913 |
2013-01-03 |
Mayberry CO2 Probe B |
Removed |
Lab none |
Removed from the tules near the tule tower due to bad values. |
659 |
2012-07-26 |
Lab none |
Installed |
Mayberry CO2 Probe B |
Installed on the Tule Tower to measure water CO2 concentration. |
1418 |
2012-06-27 |
Lab none |
Calibration |
Lab none |
Trueppm = Probeppm*1.003 + 133; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 3056ppm, and 10000ppm. Ta=25.9C, P=101.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
|
|
|