LI-7500A 75H-2150 | 2019-08-01 | Lab none | Calibration | Lab none | Last calibration: 2019-04-15
AGC=97.7 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=23.4C, Tdew=13.3C Zero Air at 975psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9014 -> .9022 Offset: 3.41ppm CO2 span 1.003 -> 1.0086 Offset: -2.05ppm H2O zero 0.8671 -> .8673 Offset: 1.2ppm H2O span 1.0077 -> 10.114 Offset: -.08ppm Notes: When SS was 97.7, I reset it to 100. I uploaded the the calibration file from 2018-10-24 and copied in the manual coefficients for CO2 span, CO2 zero, H2O span, and H2O zero from the database's record of the most recent calibration on 2019-04-15. |
LI-7500A 75H-2669 | 2019-08-01 | Lab none | Calibration | Lab none | Last calibration: 2019-04-15
AGC=102.6 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=22.7C, Tdew=14.1C Zero Air at 975psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9186 -> .9178 Offset: -3.12ppm CO2 span 1.0134 -> 1.0125 Offset: .53ppm H2O zero 0.8991 -> .9019 Offset: 17.8ppm H2O span 1.0017 -> 1.015 Offset: -.3ppm Notes: When SS was 102.6, I reset it to 100. I uploaded the the calibration file from 2018-10-24 and copied in the manual coefficients for CO2 span, CO2 zero, H2O span, and H2O zero from the database's record of the most recent calibration on 2019-04-15. |
LI-7500A 0035 | 2019-08-01 | Lab none | Calibration | Lab none | Last calibration: 2019-04-26
AGC=99.2 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=24.0C, Tdew=13.8C Zero Air at 975psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9371 -> .9374 Offset: 1.23ppm CO2 span 1.0189 -> 1.0207 Offset: -.96ppm H2O zero 0.8989 -> .8984 Offset: -2.9ppm H2O span 1.0186 -> 1.0323 Offset: -.29ppm Notes: When the SS was 99.2, I reset it to 100. I uploaded the the calibration file from 2018-10-24 and copied in the manual coefficients for CO2 span, CO2 zero, H2O span, and H2O zero from the database's record of the most recent calibration on 2019-04-15. |
LI-7500A 0041 | 2019-07-25 | Lab none | Installed | Sherman Wetland Gas Analyzer CO2/H2O | Installed at SW to replace sn 75H-2150, which was brought back to the lab for calibration. |
LI-7500A 75H-2150 | 2019-07-25 | Sherman Wetland Gas Analyzer CO2/H2O | Removed | Lab none | Removed from SW and brought back to the lab for calibration. It was replaced by sn 0041. |
LI-7500A 75H-2667 | 2019-07-25 | Lab none | Installed | Sherman Barn Gas Analyzer CO2/H2O | Installed at SB to replace sn 75H-2669, which was brought back to the lab for calibration. |
LI-7500A 75H-2669 | 2019-07-25 | Sherman Barn Gas Analyzer CO2/H2O | Removed | Lab none | Removed from SB and brought back to the lab for calibration. it was replaced by sn 75H-2667. |
LI-7500A 0042 | 2019-07-25 | Lab none | Installed | Mayberry Gas Analyzer CO2/H2O | Installed at MB to replace 75H-2182, which was brought back to the lab for calibration. |
LI-7500A 75H-2182 | 2019-07-25 | Mayberry Gas Analyzer CO2/H2O | Removed | Lab none | Removed from MB and brought back to the lab for calibration. It was replaced by sn 0042. |
PAR-LITE 010161 | 2019-07-25 | Lab none | Installed | Mayberry PAR Sensor Reflected Water | Installed at MB (PAR reflected water) to replace sn 040412, which was brought back to the lab for testing. |
PAR-LITE 040412 | 2019-07-25 | Mayberry PAR Sensor Reflected Water | Removed | Lab none | Removed from MB (PAR reflected water) and brought back to the lab for testing (spiky). It was replaced by sn 010161. |
NR-LITE 990351 | 2019-07-25 | Lab none | Installed | Mayberry Net Radiometer Water | Installed at MB over the water to replace sn 990350, which was brought back to the lab for repair. |
NR-LITE 990350 | 2019-07-25 | Mayberry Net Radiometer Water | Removed | Lab none | Removed from MB (Rnet_water) and brought back to the lab for repair (broken cable). It was replaced by sn 990351. |
LI-7500A 75H-2176 | 2019-07-25 | Lab none | Installed | East End Gas Analyzer CO2/H2O | Installed at EE to replace sn 0035, which was brought back to the lab for calibration. |
LI-7500A 0035 | 2019-07-25 | East End Gas Analyzer CO2/H2O | Removed | Lab none | Removed from EE and brought back to the lab for calibration. it was replaced by 75H-2176. |
PAR-LITE 10162 | 2019-07-25 | Lab none | Installed | East End PAR Sensor Incoming | Installed at EE to replace sn 110340, which was removed for spiky behavior. |
PQS 1 110340 | 2019-07-25 | East End PAR Sensor Incoming | Removed | Lab none | Removed from EE because it had been spiky in the last month. It was replaced sn 10162. |
LI-7500A 75H-2667 | 2019-07-24 | Lab none | Calibration | Lab none | Last calibration: 2019-02-11 AGC=104.1 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=26.7C, Tdew=11.7C Zero Air at 1000psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9425 -> .9413 Offset: -4.47ppm CO2 span 1.0143 -> 1.0158 Offset: -.77ppm H2O zero 0.8349 -> .8363 Offset: 8.7ppm H2O span 1.0222 -> 1.0222 Offset: -.02ppm Notes: When the SS was 104.1, I reset it to 100.0 |
LI-7500A 0042 | 2019-07-24 | Lab none | Calibration | Lab none | Last calibration: 2019-01-02
AGC=97.7 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=27.3C, Tdew=14.4C Zero Air at 1000psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9489 -> .9481 Offset: -3.43ppm CO2 span 1.0187 -> 1.0049 Offset: 7.19ppm H2O zero 0.8434 -> .8433 Offset: -1.0ppm H2O span 1.0334 -> 1.0179 Offset: .35ppm Notes: When SS was 97.7, I reset it to 100.0 |
LI-7500A 0041 | 2019-07-24 | Lab none | Calibration | Lab none | Last calibration: 2019-01-02
AGC=97.8 SDM= DAC1= DAC2= Ambient P=100.5kPa, Tair=27.2C, Tdew=12.7C Zero Air at 1000psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9953 -> .9970 Offset: 5.75ppm CO2 span 1.0194 -> 1.0163 Offset: -1.89ppm H2O zero 0.8248 -> .8201 Offset: -26.5ppm H2O span 1.0312 -> 1.0243 Offset: .16ppm Notes: When SS was 97.8, I reset it to 100.0. Larger H2O zero offset than I expected, but H2O span was ok. Last calibration, the H2O zero offset was a similar magnitude and positive. |
LI-7500A 75H-2176 | 2019-07-24 | Lab none | Calibration | Lab none | Last calibration: 2019-01-02
AGC=102.0 SDM= DAC1= DAC2= Ambient P=100.5kPa, Tair=26.1C, Tdew=14.5C Zero Air at 1000psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.8676 -> .8672 Offset: -2.14ppm CO2 span 1.0114 -> 1.0061 Offset: 2.7ppm H2O zero 0.8525 -> .8521 Offset: -2.4ppm H2O span 0.8206 -> .9182 Offset: .05ppm Notes: When SS was 102.0, I reset it to 100. |
GP-001 GP20150101 | 2019-07-10 | East End CO2 probe | Removed | Lab none | Removed from EE and brought back to the lab for further testing (it was maxing out, even when out of the water). |
GP-001 GP20150101 | 2019-01-18 | Lab none | Installed | East End CO2 probe | Installed at EE to replace sn GP20140019, which was brought back to the lab for testing. |
LI-7500 0976 | 2009-01-08 | Lab none | Installed | Tonzi-Floor | Installed on the Tonzi Floor to replace sn 0042 for calibration |
LI-7500A 0042 | 2009-01-08 | Tonzi-Floor | Removed | Lab none | Removed from the Tonzi Floor for calibration. Replaced by sn 0976. |
LI-7500A 0418 | 2009-01-08 | Lab Gas Analyzer CO2/H2O | Installed | Tonzi-Floor | Installed on the Tonzi Tower to replace sn 0065 for calibration |
LI-7500A 0065 | 2009-01-08 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab Gas Analyzer CO2/H2O | Removed from the Tonzi Tower for calibration. Replaced by sn 0418. |
LI-7500A 0041 | 2009-01-08 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed on the Vaira tower to replace sn 0073 for calibration |
LI-7500A 0073 | 2009-01-08 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from the Vaira Tower for calibration. Replaced by sn 0041. |
Mark I A001911 | 2019-07-02 | Lab none | Installed | Sherman Wetland Weather Station | Installed at the midpoint of the western railing. Should be remounted on the radiation boom. |
Mark I A000251 | 2019-07-02 | Lab none | Installed | Tonzi | Installed at the top of the tower east of the southeast corner. Should be remounted on the radiation boom. |
URM06 URM06-001 | 2019-06-20 | Bouldin Alfalfa Range Finder | Moved | Bouldin Corn Range Finder | Moved from BA to BC to look at corn height. It was installed on the BC radiation boom. |
miniDOT 1220 | 2019-06-13 | Lab none | Tested | Lab none | Tested but we probably can't use this sensor any more. We put in new AA lithium batteries, and the miniDOT was able to power up, but the LED light continuously blinks 2x, "card error." We tried re-formatting the card, using another 2GB card previously used in a miniDOT (could have been a bad card), and then with a good SD card at EE. In all cases, the LED status light still indicated a "card error." |
miniDOT 1220 | 2019-05-23 | Mayberry DO Sensor E | Removed | Lab none | Brought back to the lab for testing. The o-ring was warped and it was damp inside. There was some corrosion on the board and the batteries were dead. |
PTB110 E1330045 | 2019-05-23 | Mayberry Barrometer | Removed | Lab none | Reading dropped down to 50, so removed and brought back to the lab for inspection. |
miniDOT 1220 | 2018-11-28 | Mayberry DO Sensor E | Maintenance | Mayberry DO Sensor E | Swapped 2x AA lithium batteries |
miniDOT 1221 | 2018-08-23 | Mayberry DO Sensor D | Maintenance | Mayberry DO Sensor D | Swapped 2x AA lithium batteries |
2300-000 2300565 | 2019-06-04 | Lab none | Installed | Mayberry Fluorometer | Installed at MB after a thorough cleaning in the lab. |
2300-000 2300565 | 2019-02-28 | Lab none | Installed | Mayberry Fluorometer | Reinstalled after a thorough cleaning and getting new brushes in the lab. |
2300-000 2300565 | 2019-05-31 | Mayberry Fluorometer | Removed | Lab none | Removed from the tower for a thorough cleaning in the lab. |
SI-131 7385 | 2019-05-30 | Lab none | Installed | Tonzi | Installed near the Floor solar panels looking at the grass to the north. Footprint of about 1.23m x 2.5m diameter |
TS01 TS01-002 | 2019-05-30 | Lab none | Installed | Tonzi | Installed near an oak tree about 25m north of the Floor solar panels. Measuring trunk temperature from about 110cm distance. |
SI-131 7638 | 2019-05-30 | Lab none | Installed | Tonzi | Installed near an oak tree about 25m north of the Floor solar panels. Measuring trunk temperature from about 40cm distance. |
CR10X 47772 | 2019-05-30 | Lab none | Installed | Tonzi | Installed near an oak tree about 25m north of the Floor solar panels to support two IRTs |
TS01 TS01-001 | 2019-05-30 | Lab none | Installed | Tonzi | Installed near the Floor solar panels looking at the grass to the north. Footprint of about 0.7m x 0.35m diameter |
LI-7550 AIU-1866 | 2019-05-31 | Factory | Received | Lab none | Received back from Licor. They replaced the RF chip in the power supply. |
HMP45AC T2130005 | 2019-05-22 | Lab none | Tested | Lab none | I removed the power cord and soldered on an extension cord. Tested in the lab with radiation shields and fans, and compared well with two HMP60s sn L1740398 (Campbell has OK'd this one) and sn G4540023. This sensor is probably ok, but maybe we can test it with a reference AirT/RH sensor from AMP. |
CS451 70011234 | 2018-12-10 | Lab none | Tested | Lab none | Tested in both air and water in the lab and saw an offset of about +0.975 cm. See RT report "Pressure transducer comparison 12-2018." |
CS451 70011229 | 2018-12-10 | Lab none | Tested | Lab none | Tested in both air and water in the lab, but saw diurnal fluctuations of only 0.1 cm, not the 5-6 cm we saw in the field. See RT report "Pressure transducer comparison 12-2018." |
CR1000 47108 | 2019-05-28 | Sherman Barn Data Logger Met | Notes | Sherman Barn Data Logger Met | VX2 may be unreliable. NR01 PRT was drifting constantly until it was re-wired from VX2 to VX3, but that could also have been a wiring issue with another channel. |
AM16/32B 11662 | 2019-05-28 | Sherman Barn Multiplexer Met | Notes | Sherman Barn Multiplexer Met | Channel 28 and 32 are probably bad (see field notes from SB 2019-04-11). |
LI-7550 AIU-1866 | 2019-05-09 | Lab none | Shipped | Factory | Shipped back to Licor for them to inspect. |
LI-7500A 75H-2514 | 2019-05-07 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed on the Floor tower to replace the one that was moved to Tonzi Tower last week. |
LI-7550 AIU-0666 | 2019-05-07 | Lab none | Installed | Tonzi-Floor Interface LI-7500 | Installed on the Floor tower to replace the one that was moved to Tonzi Tower last week. |
LI-7500A 0035 | 2019-05-02 | Lab none | Installed | East End Gas Analyzer CO2/H2O | Installed at EE to replace sn 75H-2667, which was brought back to the lab for calibration. |
LI-7500A 75H-2667 | 2019-05-02 | East End Gas Analyzer CO2/H2O | Removed | Lab none | Removed from EE and brought back to the lab for calibration. it was replaced by sn 0035. |
LI-7500A 0042 | 2019-04-30 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from the Tower for calibration. Replaced by sn 1594. The Tower 7550 box died and was replaced by the Floor 7550 box. No reason to believe the Head was damaged by the box crash but might be worth checking. |
LI-7500A 0041 | 2019-04-30 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Floor tower for calibration. Not replaced because the Floor 7550 box was moved to the Tower to replace the 7550 box there that died. |
LI-7500A 1594 | 2019-04-30 | Lab none | Installed | Tonzi-Tower | Installed on the Tower to replace sn 0042 for calibration. Using Floor 7550 box because Tower 7550 box died |
LI-7550 AIU-1869 | 2019-04-30 | Tonzi-Floor Interface LI-7500 | Moved | Tonzi-Tower Gas Analyzer CO2/H2O | Moved to the Tower to replace sn AIU-1866 which had some kind of power problem and failed. |
LI-7550 AIU-1866 | 2019-04-30 | Tonzi-Tower Interface LI-7550 | Removed | Lab Interface LI-7550 | Removed from Tonzi Tower due to power issues. Power light was flickering with audible relay click when a head was attached. Seemed to boot without head. Might have caused 5A Tower load breaker to pop on 4-23. Brought in for testing - replaced with Floor box sn AIU-1869 |
LI-7500A 75H-2668 | 2019-04-30 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed on the Vaira tower to replace sn 74H-2176 for calibration. |
LI-7500A 75H-2176 | 2019-04-30 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 75H-2668. |
LI-7550 AIU-1866 | 2019-05-01 | Lab none | Tested | Lab none | Plugged it in to the 24V wall power and it was dead. Will e-mail Licor. |
RF401 35754 | 2019-03-11 | Lab none | Installed | Tonzi | Installed in the OSU SP3 box to connect to the Tower computer. |
RF401 34362 | 2019-03-11 | Lab none | Installed | Tonzi | Installed in the Tower computer box to connect the computer to the remote data loggers. |
miniDOT 1219 | 2016-10-11 | Mayberry DO Sensor E | Moved | Mayberry DO Sensor D | Moved from middle position to upper position (top sensor could not be removed so it was moved to lower position; bottom/lower sensor was removed). |
miniDOT 1220 | 2018-10-04 | Mayberry DO Sensor E | Notes | Mayberry DO Sensor E | Measured at -45 cm. |
miniDOT 1220 | 2017-10-13 | Lab none | Installed | Mayberry DO Sensor E | After the barnacle was removed in the lab, reinstalled at MB on the floating boom, lower position. |
miniDOT 1220 | 2017-08-16 | Mayberry DO Sensor E | Removed | Lab none | Removed from lower position (-62 cm) on floating boom and brought in to the lab for barnacle removal. |
miniDOT 1221 | 2018-10-04 | Mayberry DO Sensor D | Notes | Mayberry DO Sensor D | Measured at -15 cm. |
miniDOT 1221 | 2017-05-25 | Mayberry DO Sensor D | Moved | Mayberry DO Sensor D | Moved from static boom (upper position, -10 cm based on water level on 2017-05-10) to floating boom with sn 1219 (upper position, -17 cm). |
miniDOT 1219 | 2017-05-25 | Mayberry DO Sensor D | Moved | Mayberry DO Sensor D | Moved from static boom (upper position, -10 cm based on water level on 2017-05-10) to floating boom with sn 1221 (upper position, -17 cm). |
miniDOT 1220 | 2017-05-25 | Mayberry DO Sensor D | Moved | Mayberry DO Sensor E | Moved from static boom (upper position, -10 cm based on water level on 2017-05-10) to floating boom (lower position, -62 cm). |
miniDOT 1219 | 2015-04-23 | Mayberry DO Sensor E | Moved | Mayberry DO Sensor E | Sensor was moved up 10 cm on fence post: from -65 cm (based on water level 2014-10-09) to -50 cm (based on water level on 2015-04-23). |
miniDOT 1221 | 2015-04-23 | Mayberry DO Sensor F | Moved | Mayberry DO Sensor F | Sensor was moved up 15 cm on the fence post: from -109 cm (based on water level 2014-10-09) to -92 cm (based on water level on 2015-04-23). |
miniDOT 1221 | 2017-05-10 | Mayberry DO Sensor D | Mayberry DO Sensor D | sn 1221 moved slightly in the upper position from -8cm (based on water level on 2017-02-23) to -10cm (based on water level on 2017-05-10) with the other 2 miniDOTs. | |
miniDOT 1220 | 2017-05-10 | Mayberry DO Sensor E | Mayberry DO Sensor D | sn 1220 moved from middle position (-38 cm based on water level on 2017-02-23) to upper position (-10cm based on water level on 2017-05-10) with the other 2 miniDOTs. | |
miniDOT 1219 | 2017-05-10 | Mayberry DO Sensor F | Moved | Mayberry DO Sensor D | sn 1219 moved from lower position (-68 cm based on water level on 2017-02-23) to upper position (-10cm based on water level on 2017-05-10). All 3 miniDOT sensors (sn 1219, sn 1220, sn 1221) are at this depth for now. |
HMP45AC U3030046 | 2019-04-29 | Lab none | Tested | Lab none | Continued testing in lab and RH is consistently bad. Do not use. |
LI-7500A 0042 | 2004-04-07 | Lab none | Calibration | Lab none | Calibrated in the lab after being recently returned from Licor. There is a change in the power wiring. Red/white is positive, green is negative. They also changed the delay to 240 ms. |
LI-7500A 0035 | 2004-04-07 | Lab none | Calibration | Lab none | Calibrated in the lab after being recently returned from Licor. There is a change in the power wiring. Red/white is positive, green is negative. They also changed the delay to 240 ms. |
URM06 URM06-002 | 2019-04-25 | Lab none | Installed | Mayberry | Installed on the Mayberry Tower on a short arm off the center-east side about 68cm above the water level. |
LI-7700 TG1-0223 | 2019-04-25 | Mayberry Gas Analyzer CH4 | Upgrade | Mayberry Gas Analyzer CH4 | The thermocouple module was broken so we swapped it for a new one. The LI-7700 firmware was upgraded from v1.0.21 to v1.0.27 mostly minor bug fixes see change log https://www.licor.com/env/support/LI-7700/topics/change-log.html |
LI-7500A 0418 | 2019-04-25 | Lab none | Installed | West Pond Gas Analyzer CO2/H2O | Installed on West Pond to replace sn 75H-2668 for calibartion |
LI-7500A 75H-2668 | 2019-04-25 | West Pond Gas Analyzer CO2/H2O | Removed | Lab none | Removed from West Pond for calibration. Replaced by sn 0418 |
LI-7700 TG1-0224 | 2019-04-25 | West Pond Gas Analyzer CH4 | Upgrade | West Pond Gas Analyzer CH4 | The LI-7700 firmware was upgraded from v1.0.21 to v1.0.27 mostly minor bug fixes see change log https://www.licor.com/env/support/LI-7700/topics/change-log.html |
URM06 URM06-001 | 2019-04-25 | Lab none | Installed | Bouldin Alfalfa | Installed on the Boldin Alfalfa radiation boom about 262cm above the ground to measure veg height. |
LI-7500A 75H-2180 | 2019-04-25 | Lab none | Installed | Bouldin Alfalfa Gas Analyzer CO2/H2O | Installed on Bouldin Alfalfa to replace sn 0035 for calibration |
LI-7500A 0035 | 2019-04-25 | Bouldin Alfalfa Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Alfalfa for calibration. Replaced by sn 75H-2180. |
LI-7500A 1594 | 2019-04-26 | Lab none | Calibration | Lab none | Last calibration: 2019-01-30 AGC=99.5 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=24.1C, Tdew=10.2C Zero Air at 1050psi, AMP CA4396 408.97 at 700 Tdewspan=16.00 CO2 zero 0.8643 -> .8646 Offset: 1.33ppm CO2 span 1.0157 -> 1.0178 Offset: -1.16ppm H2O zero 0.8891 -> .8868 Offset: -12.6ppm H2O span 1.0043 -> 1.0026 Offset: .04ppm Notes: |
LI-7500A 0035 | 2019-04-26 | Lab none | Calibration | Lab none | Last calibration: 2019-01-28 AGC=99.2 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=22.6C, Tdew=10.7C Zero Air at 1100psi, AMP CA4396 408.97 at 700 Tdewspan=16.00 CO2 zero 0.9375 -> .9371 Offset: -1.79ppm CO2 span 1.0193 -> 1.0189 Offset: .25ppm H2O zero 0.8969 -> .8989 Offset: 12.3ppm H2O span 1.0023 -> 1.0186 Offset: -.34ppm Notes: |
LI-7500A 0073 | 2019-04-25 | Lab none | Installed | Bouldin Corn Gas Analyzer CO2/H2O | Installed on Bouldin Corn to replace sn 1594 for calibration |
LI-7500A 1594 | 2019-04-25 | Bouldin Corn Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Corn for calibration. Replaced by sn 0073. |
LI-7700 TG1-0155 | 2019-04-25 | Bouldin Corn Gas Analyzer CH4 | Upgrade | Bouldin Corn Gas Analyzer CH4 | The LI-7700 firmware was upgraded from v1.0.21 to v1.0.27 mostly minor bug fixes see change log https://www.licor.com/env/support/LI-7700/topics/change-log.html |
LI-7500A 75H-2668 | 2019-04-26 | Lab none | Calibration | Lab none | Last calibration: 2019-02-11 AGC=100.0 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=24.5C, Tdew=11.1C Zero Air at 1050psi, AMP CA4396 408.97 at 700 Tdewspan=16.00 CO2 zero 0.939 -> .9391 Offset: .22ppm CO2 span 0.018 -> 1.0169 Offset: .59ppm H2O zero 0.9073 -> .9069 Offset: -2.2ppm H2O span 1.0117 -> 1.0139 Offset: -.05ppm Notes: |
LI-7700 TG1-0223 | 2019-04-25 | Mayberry Gas Analyzer CH4 | Calibration | Mayberry Gas Analyzer CH4 | Calibration Location: MB
Last calibration: 2018-10-26 RSSI=80%, Optical RH=12% Ambient P=100.9kPa, Tair=26C Zero Air at 1600psi, AMP CA3923 3.121 at 1600 CH4 Zero Offset: 0.013ppm CH4 Span Offset: 0.015ppm Notes: TC module swapped before calibration; breezy day |
HMP45AC T2130005 | 2019-04-24 | Lab none | Tested | Lab none | Tested overnight in the lab. This sensor has an extension cable soldered on. Didn't see any spikes, and the numbers seem reasonable. However, the control wire seems extraneous--the AirT and RH numbers seem ok even when the control wire is not connected to the datalogger, even though the sensor still has the power control board. More testing needed before we deploy it. |
HMP45AC U3030046 | 2019-04-24 | Lab none | Tested | Lab none | Tested overnight in the lab. This sensor has an extension cable soldered on. The RH was consistently 5-7% lower than the other 3 sensors (L1740398, T2130005, G4540023). I removed the extension cable, and the numbers were still low. I removed the humidity chip, and it read 150%. When I replaced the humidity chip, the RH was within 2-4% of the other 3 sensors. I added the same extension cable back on, and the numbers became low. This sensor doesn't look consistent and needs more testing before it can be deployed. |
CR10X 47772 | 2019-04-16 | Tonzi | Removed | Lab none | Removed with the wooden cal panel for the FLIR because cows chewed the TC wires and the new cal panel is better |