View history entry 4127 for Licor LI-7500A sn: 1594

Date: 2019-10-30
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2019-04-26 AGC=99.6 SDM=0 DAC1=- DAC2=- Ambient P=100.9kPa, Tair=24.6C, Tdew=-11.9C Zero Air at 900psi, AMP CA04396 408.97 at 800 Tdewspan=20.0 CO2 zero 0.8646 -> 0.8646 Offset: -0.3ppm CO2 span 1.0178 -> 1.0166 Offset: 0.53ppm H2O zero 0.8868 -> 0.8895 Offset: -1.7ppm H2O span 1.0026 -> 0.9958 Offset: 0.18ppm Notes: Ambient air was very dry. Joe