Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Forerunner-Eosense
Model No:GP-001
Serial No:GP20160096
Tag No:
Acquisition Date:2017-09-20
Project:LBNL
Location on 2020-10-08:Factory,
Notes: 0-1% range. Received from LBNL in exchange from probes given to Jennifer Soong on 2017-09-08

Documents

Calibration: cal_FR-Eosense_GP002_snGP20160096_20170915.pdf
Report: Eosense CO2 Probe Compare - 2018-04-23

History

Rec No.DateLoc FromTitleLoc ToNotes
4858 2020-10-08 Lab none Shipped Factory none Sent to Eosense for inspection
4852 2020-10-05 East End CO2 probe Removed Lab none Removed from EE because it still rails high in air and in water. Will send it back to the factory
4850 2019-10-02 Lab none Installed East End CO2 probe Installed at EE off the north side of the tower.
4111 2019-09-20 Lab none Tested Lab none Tested in the lab, and seems to be responding fine against another 1% probe. Not sure what the issue was in the field. However, the connection between the sensor and the screw connector is loose, so it should be sealed before re-deployment.
4108 2019-09-18 East End CO2 probe Removed Lab none Removed from East Pond because it was railed high even in air.
4097 2019-09-04 Lab none Installed East End CO2 probe The CO2 probe attached to the plastic chain to hang next to the miniDot.
 
 