Equipment and CalibrationView History
Equipment Entry
Instrument: | CO2 Probe |
Manufacture: | Forerunner-Eosense |
Model No: | GP-001 |
Serial No: | GP20160096 |
Tag No: | |
Acquisition Date: | 2017-09-20 |
Project: | LBNL |
Location on 2020-10-08: | Factory, |
Notes: |
0-1% range. Received from LBNL in exchange from probes given to Jennifer Soong on 2017-09-08 |
|
|
Documents
Calibration: cal_FR-Eosense_GP002_snGP20160096_20170915.pdf
Report: Eosense CO2 Probe Compare - 2018-04-23
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
4858 |
2020-10-08 |
Lab none |
Shipped |
Factory none |
Sent to Eosense for inspection |
4852 |
2020-10-05 |
East End CO2 probe |
Removed |
Lab none |
Removed from EE because it still rails high in air and in water. Will send it back to the factory |
4850 |
2019-10-02 |
Lab none |
Installed |
East End CO2 probe |
Installed at EE off the north side of the tower. |
4111 |
2019-09-20 |
Lab none |
Tested |
Lab none |
Tested in the lab, and seems to be responding fine against another 1% probe. Not sure what the issue was in the field. However, the connection between the sensor and the screw connector is loose, so it should be sealed before re-deployment. |
4108 |
2019-09-18 |
East End CO2 probe |
Removed |
Lab none |
Removed from East Pond because it was railed high even in air. |
4097 |
2019-09-04 |
Lab none |
Installed |
East End CO2 probe |
The CO2 probe attached to the plastic chain to hang next to the miniDot. |
|
|
|