Equipment and Calibration History
Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 [17] 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70Ser No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
HMP155 T2930114 | 2021-11-05 | Bouldin Alfalfa | Moved | Bouldin Alfalfa none | Removed from main tower (lower position) at the conclusion of this year's advection experiment and installed on the main tower for an HMP intercomparison |
S2-411 1050 | 2021-11-05 | Cal Kit Narrow Band NDVI Reflected | Removed | Lab none | Removed from BA tower at the conclusion of the rad calibration and brought back to the lab. |
S2-411 1049 | 2021-11-05 | Cal Kit Narrow Band NDVI Incoming | Removed | Lab none | Removed from BA tower at the conclusion of the rad calibration and brought back to the lab. |
PAR-LITE 060831 | 2021-11-05 | Cal Kit PAR Sensor Incoming | Removed | Lab none | Removed from BA tower at the conclusion of the rad calibration and brought back to the lab. This sensor will go back to AMP. |
CNR4 121038 | 2021-11-05 | Cal Kit Net Radiometer 4-way | Removed | Lab none | Removed from BA tower at the conclusion of the rad calibration and brought back to the lab. This sensor will go back to AMP. |
WM 1590 152601 | 2021-11-05 | Bouldin Alfalfa-ADEX Anemometer Main Lower | Removed | Lab none | Removed from main tower (lower position) and brought back to the lab at the conclusion of this year's advection experiment |
HMP155 T2820183 | 2021-11-05 | Bouldin Alfalfa none | Moved | Bouldin Alfalfa none | Removed from bouncer tower (lower position) at the conclusion of this year's advection experiment and installed on the main tower for an HMP intercomparison |
WM 1590 153805 | 2021-11-05 | Bouldin Alfalfa-ADEX Anemometer Bouncer Lower | Removed | Lab none | Removed from bouncer tower (lower position) and brought back to the lab at the conclusion of this year's advection experiment |
HMP155 T2820180 | 2021-11-05 | Bouldin Alfalfa-ADEX RH & Temp Bouncer Upper | Moved | Cal Kit RH & Temp | Removed from bouncer tower (upper position) at the conclusion of this year's advection experiment and installed on the main tower for an HMP intercomparison |
WM 1590 152603 | 2021-11-05 | Bouldin Alfalfa-ADEX Anemometer Bouncer Upper | Removed | Lab none | Removed from bouncer tower (upper position) and brought back to the lab at the conclusion of this year's advection experiment |
CNR4 120788 | 2021-11-08 | Borrowed none | Received | Lab none | Sebastien of AMP dropped this off today as an exchange for an older 4-way radiometer reference sensor, sn 121038. Came with a mounting rod and cables only. |
PQS 1 212225 | 2021-11-08 | Borrowed none | Received | Lab none | Sebastien of AMP dropped this off in exchange for an older reference PAR, sn 060831. It came in its own box and a rubber cap over the sensor. |
CNR4 121038 | 2021-11-08 | Lab none | Returned | Returned none | Returned to Sebastien of AMP in exchange for a freshly calibrated 4-way radiometer CNR4 sn 120788 |
PAR-LITE 060831 | 2021-11-08 | Lab none | Returned | Returned none | Returned to Sebastien of AMP in exchange for a freshly calibrated reference PAR sn 212225 |
TR-5251 40415-1006 | 2021-11-04 | Mayberry Precipitation | Adjusted | Mayberry Precipitation | Tested tipping bucket - got 53/50 tips. Adjusted screws 1/2 turn out (CCW). |
GP-001 GP20160127 | 2021-11-04 | Lab none | Installed | Hill Slough CO2 Probe | Installed near the miniDot on the SW corner of the platform. This sensor floats so that the sensor window is about 10cm below the water surface. There is enough slack in the cable so it can go down some more with a lower tide, but hard to say where it will be with higher tides. |
miniDOT-usb 931528 | 2021-11-04 | Lab none | Installed | Hill Slough DO Sensor | Installed on a yellow chain near the SW corner of the platform. I think the sensor window is about 10cm above the ground |
TE525 48130-611 | 2021-11-04 | Hill Slough Precipitation | Adjusted | Hill Slough Precipitation | The screws were adjusted 1/2 turn in (CW) before testing. The test got 51/50 tips - okay. |
LI-190/R Q100420 | 2021-10-21 | Lab none | Installed | Mayberry PAR Sensor Diffuse PAR | Installed as part of the rotating shadow band for diffuse PAR - not working until 11-04 |
CR10X 35474 | 2021-10-21 | Lab none | Installed | Mayberry Data Logger Diffuse PAR | Installed to measure diffuse PAR from rotating shadow band. |
LI-7500A 0418 | 2021-10-21 | Mayberry Gas Analyzer CO2/H2O | Adjusted | Mayberry Gas Analyzer CO2/H2O | Rest SS from 104 to 100 after cleaning |
LI-7500A 0073 | 2021-09-08 | Lab none | Shipped | Factory none | Shipped to Licor for renovation. The signal is noisy and I have asked for its chopper, detector and light source to be replaced as these parts are about 20 years old. |
LI-7500A 0073 | 2021-11-03 | Factory none | Returned | Lab none | Returned to the lab from revamp and calibration at Licor. |
LI-7500A 0065 | 2021-10-21 | Gilbert Tract Gas Analyzer CO2/H2O | Adjusted | Gilbert Tract Gas Analyzer CO2/H2O | Reset signal strength from 108 to 100 |
SP75 020854E-6-03-02-06862 | 2021-10-26 | Vaira Solar Panel SoilCO2-B | Removed | Lab none | Removed from Vaira - not in use, was for old soil co2 system. |
SP75 020854E-6-03-02-06905 | 2021-10-26 | Vaira Solar Panel SoilCO2-A | Removed | Lab none | Removed from Vaira - not in use, was for old soil co2 system. |
SP75 020854E-6-03-02-06862 | 2012-02-16 | Lab none | Installed | Vaira Solar Panel SoilCO2-B | Installed to provide 24V power for soil CO2 system but mistakenly wired for 12V - fixed on 3/19 |
SP75 020854E-6-03-02-06905 | 2012-02-16 | Lab none | Installed | Vaira Solar Panel SoilCO2-A | Installed to provide 24V power for soil CO2 system but mistakenly wired for 12V - fixed on 3/19 |
SP75 020854E-6-03-02-06862 | 2012-02-09 | Tonzi Solar Panel Floor H | Removed | Lab none | Removed from the Floor solar system. This panel has not been connected to the system since at least August 2008 |
SP75 020854E-6-03-02-06905 | 2012-02-09 | Tonzi Solar Panel Floor G | Removed | Lab none | Removed from the Floor solar system. This panel has not been connected to the system since at least August 2008 |
TE525MM 65560-615 | 2021-10-21 | Bouldin Alfalfa Precipitation | Tested | Bouldin Alfalfa Precipitation | We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 54 tips and adjusted its set screws 1.5 turns CCW (1/2 turn of both screws is a 2-3% adjustment). |
TE525MM 65548-615 | 2021-10-21 | Bouldin Corn Precipitation | Tested | Bouldin Corn Precipitation | We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 54 tips and adjusted its set screws 1.5 turns CCW (1/2 turn of both screws is a 2-3% adjustment). |
TR-525M 50229-112 | 2021-10-21 | East End Precipitation | Tested | East End Precipitation | We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 54 tips and adjusted its set screws 1.5 turns CCW (1/2 turn of both screws is a 2-3% adjustment). |
TR-5251 40415-1006 | 2021-10-21 | Mayberry Precipitation | Tested | Mayberry Precipitation | We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 54 tips and adjusted its set screws 1.5 turns CCW (1/2 turn of both screws is a 2-3% adjustment). |
S2-411 1050 | 2021-10-21 | Cal Kit Narrow Band NDVI Reflected | Moved | Cal Kit Narrow Band NDVI Reflected | Moved from GT to BA as part of the radiometer cal kit to begin the calibration at BA |
S2-411 1049 | 2021-10-21 | Cal Kit Narrow Band NDVI Incoming | Moved | Cal Kit Narrow Band NDVI Incoming | Moved from GT to BA as part of the radiometer cal kit to begin the calibration at BA |
PAR-LITE 060831 | 2021-10-21 | Cal Kit PAR Sensor Reflected | Moved | Cal Kit PAR Sensor Incoming | Moved from GT Reflected to BA Incoming as part of the radiometer cal kit to begin the calibration at BA |
CNR4 121038 | 2021-10-21 | Cal Kit Net Radiometer 4-way | Moved | Cal Kit Net Radiometer 4-way | Moved from GT to BA as part of the radiometer cal kit to begin the calibration at BA |
CR1000 23865 | 2021-10-21 | Cal Kit Data Logger | Moved | Cal Kit Data Logger | Moved from GT to BA as part of the radiometer cal kit to begin the calibration at BA |
miniDOT 1221 | 2021-10-20 | Lab none | Tested | Lab none | This sensor was covered in barnacles, the copper screen was falling apart and there are nicks in the coating on the window. I cleaned up the sensor and tested it side by side with miniDot sn 931528. Both sensors were suspended in a bucket of water with a bubbler in the bottom. The sensors were out of the stream of bubbles. Ice was added to the water bringing its temperature down from 20C to 0C. Then the dissolved O2 was recorded by both sensor over about 24 hours as the water returned to room temperature. The two miniDots matched each other by better then 2%. And they matched the theoretical O2 saturation of the water at about 4%. This sensor had a quailty flag of about 0.84 whereas the newer, uncompromised sensor (sn 931528) had a quality flag on 1.03. I think I'll send a photo of the window coating and these notes to PME to get their take, but I think it's okay to return sn 1221 to service. |
miniDOT-usb 899502 | 2021-09-02 | Lab none | Installed | Mayberry DO Sensor D | Installed at Mayberry to replace sn 1221 that was covered with barnacles. |
miniDOT 1221 | 2021-09-02 | Mayberry DO Sensor D | Removed | Lab none | Removed from Mayberry because it had a bunch of barnacles growing on it, the copper screen was disintegrating and the coating on the window looked damage. Replaced by sn 899502 |
ML2x 139/052 | 2021-03-17 | Lab none | Calibration | Lab none | After looking at the data recorded from 2021-03-17 to 2021-10-06 it seemed that this probe was reading too low in the spring and too high in the fall. Most years the soil gets completely saturated in the spring and completely dry in the fall. If you take the max and min values from all years they are very similar. Min: 473mV STDEV=6.8mV, Max: 913mV STDEV 24.8. Then we can use these to scale this sensor's min and max with this year's data and we get Corrected (mV) = raw (mV) * 1.272 -187.73 And using the manual's default conversion values for mineral soil (0.05 and -5), the total conversion is: %m = mV * 0.636 + 4.39 |
CNR4 160848 | 2019-10-21 | Lab none | Returned | Returned none | Robert returned it to Stephen around 2019-10-21 when Robert went to LBL for the weekly NGEE Tropics meeting |
LI-7500A 0065 | 2021-09-15 | Lab none | Installed | Gilbert Tract Gas Analyzer CO2/H2O | Installed at Gilbert Tract to replace sn 0041 for calibration. |
LI-7500A 0041 | 2021-09-15 | Gilbert Tract Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Gilbert Tract for calibration. Replaced by sn 0065. |
GP-001 GP20200089 | 2021-09-09 | Lab none | Installed | Gilbert Tract CO2 Probe | Installed at Gilbert Tract but the cable was too short - not hooked up until 2021-09-15 |
miniDOT-usb 906606 | 2021-09-09 | Lab none | Installed | Gilbert Tract DO Sensor | Installed at Gilbert Tract on a yellow chain that just barely meets the water. When the site becomes tidal we will have to have a longer chain or more likely a boardwalk to reach deeper water. |
HFP01 11329 | 2021-09-09 | Lab none | Installed | Gilbert Tract Heat Flux Plate C | Installed at Gilbert Tract at -2cm to the west of the tower |
HFP01 11328 | 2021-09-09 | Lab none | Installed | Gilbert Tract Heat Flux Plate B | Installed at Gilbert Tract at -2cm to the northwest of the tower |
HFP01 11325 | 2021-09-09 | Lab none | Installed | Gilbert Tract Heat Flux Plate A | Installed at Gilbert Tract at -2cm to the southwest of the tower |
TE525MM 65552-615 | 2021-09-15 | Lab none | Installed | Gilbert Tract Precipitation | Installed at Gilbert Tract on the northeast corn of the scaffolding. |
miniDOT-usb 906606 | 2021-09-02 | Sherman Wetland DO Sensor | Removed | Lab none | Removed from Sherman Wetland with the decommissioning of the site. |
HFP01 11329 | 2021-09-02 | Sherman Wetland Heat Flux Plate B | Removed | Lab none | Removed from Sherman Wetland with the decommissioning of the site. The aluminum conduit was rotting and needed to be removed. |
HFP01 11328 | 2021-09-02 | Sherman Wetland Heat Flux Plate A | Removed | Lab none | Removed from Sherman Wetland with the decommissioning of the site. The aluminum conduit was rotting and needed to be removed. |
HFP01 11325 | 2021-09-02 | Sherman Wetland Heat Flux Plate C | Removed | Lab none | Removed from Sherman Wetland with the decommissioning of the site. The aluminum conduit was rotting and needed to be removed. |
TE525MM 65552-615 | 2021-09-02 | Sherman Wetland Precipitation | Removed | Lab none | Removed from Sherman Wetland with the decommissioning of the site. |
CS450 20010026 | 2021-09-02 | Sherman Wetland Water Level | Moved | Gilbert Tract Water Level | Moved from SW to GT with the decomissioning of SW and the beginning of GT. Not installed until 09/09 |
KU265-6MCA 158APW4M0000190 | 2021-09-02 | Sherman Wetland Solar Panel A | Moved | Gilbert Tract Solar Panel A | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
KU265-6MCA 158APW4M0000193 | 2021-09-02 | Sherman Wetland Solar Panel B | Moved | Gilbert Tract Solar Panel B | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
BB-Fab BB-SW | 2021-09-02 | Sherman Wetland Battery Bank | Disposed | Disposed none | With the decommisioning of SW this battery bank becomes BB-GT |
MNKID-B KB15214 | 2021-09-02 | Sherman Wetland Solar Controller | Moved | Gilbert Tract Solar Controller | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
SD130BN 0030F4D214A5 | 2021-09-02 | Sherman Wetland Camera | Moved | Gilbert Tract Camera | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
HMP45AC W0730036 | 2021-09-02 | Sherman Wetland RH & Temp | Moved | Gilbert Tract RH & Temp | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
PTB110 L2510774 | 2021-09-02 | Sherman Wetland Barometer | Moved | Gilbert Tract Barometer | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
SRS-Ni NDVI 2046503404 | 2021-09-02 | Sherman Wetland Narrow Band NDVI Reflected | Moved | Gilbert Tract Narrow Band NDVI Reflected | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
SRS-Ni NDVI 882103059 | 2021-09-02 | Sherman Wetland Narrow Band NDVI Incoming | Moved | Gilbert Tract Narrow Band NDVI Incoming | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
SRS-Pr PRI 930203026 | 2021-09-02 | Sherman Wetland Narrow Band PRI Reflected | Moved | Gilbert Tract Narrow Band PRI Reflected | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
SRS-Pi PRI 912903006 | 2021-09-02 | Sherman Wetland Narrow Band PRI Incoming | Moved | Gilbert Tract Narrow Band PRI Incoming | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
NR01 2466 | 2021-09-02 | Sherman Wetland Net Radiometer 4-way | Moved | Gilbert Tract Net Radiometer 4-way | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
PQS 1 140450 | 2021-09-02 | Sherman Wetland PAR Sensor Reflected | Moved | Gilbert Tract PAR Sensor Reflected | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
PQS 1 140454 | 2021-09-02 | Sherman Wetland PAR Sensor Incoming | Moved | Gilbert Tract PAR Sensor Incoming | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
PQS 1 140457 | 2021-09-02 | Sherman Wetland PAR Sensor Diffuse | Moved | Gilbert Tract PAR Sensor Diffuse | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
CR10X 25799 | 2021-09-02 | Sherman Wetland Data Logger Diffuse PAR | Moved | Gilbert Tract Data Logger Diffuse PAR | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
CS547A 5688 | 2021-09-02 | Sherman Wetland Conductivity Sensor | Moved | Gilbert Tract Conductivity Sensor | Moved from SW to GT with the decomissioning of SW and the beginning of GT. Not installed until 09/09 |
NL115 13054 | 2021-09-02 | Sherman Wetland Interface NL115 | Moved | Gilbert Tract Interface NL115 | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
AM16/32B 21689 | 2021-09-02 | Sherman Wetland Multiplexer Met | Moved | Gilbert Tract Multiplexer Met | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
AM25T 7909 | 2021-09-02 | Sherman Wetland Multiplexer Thermocouple | Moved | Gilbert Tract Multiplexer Thermocouple | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
CR1000 38211 | 2021-09-02 | Sherman Wetland Data Logger Met | Moved | Gilbert Tract Data Logger Met | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
WM 1590 120906 | 2021-09-02 | Sherman Wetland Anemometer | Moved | Gilbert Tract Anemometer | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
LI-7500A 0041 | 2021-09-02 | Sherman Wetland Interface LI-7550 | Moved | Gilbert Tract Interface LI-7550 | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
LI-7550 AIU-1569 | 2021-09-02 | Sherman Wetland Gas Analyzer CO2/H2O | Moved | Gilbert Tract Gas Analyzer CO2/H2O | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
LI-7700 TG1-0424 | 2021-09-02 | Sherman Wetland Gas Analyzer CH4 | Moved | Gilbert Tract Gas Analyzer CH4 | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
LI-7500A 75H-2667 | 2021-10-07 | Lab none | Installed | East End Gas Analyzer CO2/H2O | Installed at East End to replace sn 1594 for calibration. |
LI-7500A 1594 | 2021-10-07 | East End Gas Analyzer CO2/H2O | Removed | Lab none | Removed from East End for calibration. Replaced by sn 75H-2667. |
LI-7500A 75H-2150 | 2021-10-07 | Lab none | Installed | West Pond Gas Analyzer CO2/H2O | Installed at West Pond to replace sn 75H-2182 for calibration. |
LI-7500A 75H-2182 | 2021-10-07 | West Pond Gas Analyzer CO2/H2O | Removed | Lab none | Removed from West Pond for calibration. Replaced by sn 75H-2150. |
LI-7500A 0418 | 2021-10-07 | Lab none | Installed | Mayberry Gas Analyzer CO2/H2O | Installed at Mayberry to replace sn 75H-2180 for calibration. |
LI-7500A 75H-2180 | 2021-10-07 | Mayberry Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Mayberry for calibration. Replaced by sn 0418. |
CS450 20010026 | 2020-08-28 | Lab none | Installed | Sherman Wetland Water Level | Installed after repair at Campbell -90cm from top of pipe -45cm from ground surface. |
TR-525I-R1 24875-999 | 2021-10-06 | Vaira Precipitation | Calibration | Vaira Precipitation | Tested using our new method with the peristaltic pump and 237ml of water which should result in 50 tips. I got 54 tips. I adjusted the screws two full turns CCW which should decrease the number of tips by 8-12% (4-6 tips). |
BB-Fab BB-TT | 2021-10-06 | Tonzi-Tower Battery Bank | Maintenance | Tonzi-Tower Battery Bank | I cleaned corrosion off the positive lugs and sprayed them with sealer |
PQS 1 110340 | 2021-10-06 | Lab none | Installed | Tonzi PAR Sensor Incoming | Installed at Tonzi to replace sn 010164 due to bad readings from a broken wire. |
PAR-LITE 010164 | 2021-10-06 | Tonzi PAR Sensor Incoming | Removed | Lab none | Removed from Tonzi-Tower due to bad readings from a broken wire. Replaced by sn 110340. |
TR-525I-R1 24872-999 | 2021-10-06 | Tonzi | Calibration | Tonzi none | Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 54 tips – I turned the screws two turns CCW |
TE525 48102-611 | 2021-10-06 | Tonzi | Calibration | Tonzi none | Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 50 tips – No adjustment |
TE525 48132-611 | 2021-10-06 | Tonzi | Calibration | Tonzi none | Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 52 tips – I turned the screws one turn CCW. |
LI-190/R Q100421 | 2015-05-08 | Factory none | Calibration | Factory none | Initial factory calibration: 6.63 uA/(1000umol/m2/s) Using a 1000 Ohm resistor 150.83 (umol/m2/s)/mV |
LI-7500A 0418 | 2021-09-29 | Lab | Calibration | Lab | Last calibration: 2021-09-21 AGC=100 SDM= DAC1= DAC2= Ambient P=100.5kPa, Tair=26.1C, Tdew=11.5C Zero Air at 1500psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero 0.8992 -> .8995 Offset: .96ppm CO2 span 0.9972 -> 1 Offset: -2.11ppm H2O zero 0.849 -> .849 Offset: .1ppm H2O span 0.9774 -> .9783 Offset: -.01ppm Notes: 2nd calibration since chemical swap. Not much change since the 1st calibration, which was the day after the chemical swap. CO2 span was the only one that changed significantly (although still a pretty small change). |
LI-7500A 75H-2150 | 2021-09-29 | Lab | Calibration | Lab | Last calibration: 2021-06-07 AGC=100 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=26.3C, Tdew=9.07C Zero Air at 1425psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero 0.9024 -> .9018 Offset: -2.57ppm CO2 span 1.0107 -> 1.0034 Offset: 5.1ppm H2O zero 0.8687 -> .8689 Offset: 1.1ppm H2O span 1.0111 -> 1.0112 Offset: -.01ppm Notes: |