View history entry 5595 for Licor LI-7500A sn: 1594

Date: 2021-11-30
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-05-25 AGC=97.5 SDM=0 DAC1=- DAC2=- Ambient P=100.8kPa, Tair=27.6C, Tdew=12.7C Zero Air at 1350psi, AMP CA04119 524.77 at 1700 Tdewspan=20.0 CO2 zero 0.8648 -> 0.8645 Offset: -1.4ppm CO2 span 1.0136 -> 1.0128 Offset: 0.53ppm H2O zero 0.8861 -> 0.8877 Offset: 8.7ppm H2O span 1.005 -> 1.0045 Offset: 0.01ppm Notes:
Entered by: