LI-7500A 75H-2667 | 2019-11-14 | Lab none | Installed | West Pond Gas Analyzer CO2/H2O | Installed at West Pond to replace sn 75H-2182 for calibration. |
LI-7500A 75H-2182 | 2019-11-14 | West Pond Gas Analyzer CO2/H2O | Removed | Lab none | Removed from West Pond and brought back to the lab for calibration. Replaced by sn 75H-2667. |
HFP01 6499 | 2017-09-26 | East End-Temp2 Heat Flux Plate A | Removed | Lab none | Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2. |
KD140SX-UFBS 11ZV0M0856 | 2017-06-01 | Lab none | Installed | East End-Temp1 Solar Panel B | Installed at the East End Temporary Tower #1 Open Water location with the initial set up of this site. |
KD140SX-UFBS 11ZV0M0829 | 2017-06-01 | Twitchell Rice 0 | Removed | East End-Temp2 none | Removed from the Twitchell Rice tower location (the tower was removed previously) for use at the East End temporary towers. This panel was left at the Temp 2 (closed vegetation) site to be installed later. |
KD140SX-UFBS 11ZV0M0828 | 2017-06-01 | Twitchell Rice 0 | Removed | East End-Temp2 none | Removed from the Twitchell Rice tower location (the tower was removed previously) for use at the East End temporary towers. This panel was left at the Temp 2 (closed vegetation) site to be installed later. |
PAR-LITE 040412 | 2017-06-14 | East End-Temp1 PAR Sensor Incoming | Moved | East End-Temp1 PAR Sensor Reflected | After having been initially installed in the upward looking, incoming position on the berm, this sensor was moved to a area near open water and mounted in the downward looking, reflected position. |
CR10X 31308 | 2017-07-14 | Lab none | Installed | Bouldin Corn Data Logger Diffuse PAR | Installed to do the high frequency measurements for diffuse PAR and report the CR1000 met data logger. |
PA-II-SD 84f3ebd5ffb1 | 2019-12-17 | Lab none | Installed | Bouldin Corn Air Quality Sensor | Installed at Bouldin Corn after the tower was rebuilt after being knocked down by the wind. The older Purple Air is still in the lab drying out and needs to be put back together. |
CS547A 6494 | 2019-12-17 | Lab none | Installed | East End Conductivity Sensor | Reinstalled at East End after being removed after getting wet ans possibly causing a short circuit. Tested fine in the lab. |
A547 3585 | 2019-12-17 | Lab none | Installed | East End Interface A547 | Reinstalled at East End after being removed after getting wet ans possibly causing a short circuit. Tested fine in the lab. |
LI-7500A 0042 | 2019-12-11 | Lab none | Tested | Lab none | Tested for 2+ days in the lab connected to the lab 7550, and numbers looked ok. |
LI-7700 TG1-0155 | 2019-12-11 | Lab none | Maintenance | Lab none | Joe painted on white outdoor-furniture paint on the spar with cracked paint. We ran the sensor for 2 days in the lab connected to the lab 7550, and the data seemed ok, so it seems ready to deploy in the field again. |
TE525MM 65548-615 | 2019-12-13 | Bouldin Corn Precipitation | Tested | Bouldin Corn Precipitation | Plugged tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket which should result in 100 tips. Got 102 tips and did not adjust the bucket. |
LI-7550 AIU-0614 | 2019-12-13 | Lab none | Installed | Bouldin Corn Interface LI-7550 | Re-installed at Bouldin Corn after re-assembling tower. |
LI-7500A 0042 | 2019-12-13 | Lab none | Installed | Bouldin Corn Gas Analyzer CO2/H2O | Re-installed at Bouldin Corn after re-assembling tower. |
LI-7700 TG1-0155 | 2019-12-13 | Lab none | Installed | Bouldin Corn Gas Analyzer CH4 | Re-installed at Bouldin Corn after re-assembling tower. |
WM 1590 153805 | 2019-12-13 | Lab none | Installed | Bouldin Corn Anemometer | Installed at Bouldin Corn after re-assembling tower. This unit replaces sn 074417 that lost a transducer when the tower fell. |
AM25T 7908 | 2019-12-13 | Lab none | Installed | Bouldin Corn Multiplexer Thermocouple | Re-installed at Bouldin Corn after re-assembling tower. |
AM16/32B 21690 | 2019-12-13 | Lab none | Installed | Bouldin Corn Multiplexer Met | Re-installed at Bouldin Corn after re-assembling tower. |
CR10X 31308 | 2019-12-13 | Lab none | Installed | Bouldin Corn Data Logger Diffuse PAR | Re-installed at Bouldin Corn after re-assembling tower. |
CR1000 66251 | 2019-12-13 | Lab none | Installed | Bouldin Corn Data Logger Met | Re-installed at Bouldin Corn after re-assembling tower. |
HMP60 L1740395 | 2019-12-13 | Lab none | Installed | Bouldin Corn RH & Temp | Re-installed at Bouldin Corn after re-assembling tower. |
PTB110 L2340502 | 2019-12-13 | Lab none | Installed | Bouldin Corn Barometer | Re-installed at Bouldin Corn after re-assembling tower. |
TE525MM 65548-615 | 2019-12-13 | Lab none | Installed | Bouldin Corn Precipitation | Re-installed at Bouldin Corn after re-assembling tower. |
CS451 70011234 | 2019-12-13 | Lab none | Installed | Bouldin Corn Water Level | Re-installed at Bouldin Corn after re-assembling tower. Sensor is 91cm below ground level.
Top of well to ground = 47 cm Top of well to sensor = 138 cm Top of well to water level = 131 cm (4.3’) Sensor psi reading = 0.12 psi ~ 8 cm (consistent with physical measurement) |
HFP01 10861 | 2019-12-13 | Lab none | Installed | Bouldin Corn Heat Flux Plate C | Re-installed at Bouldin Corn after re-assembling tower. Need to check if this in A/B/C location. |
HFP01 10867 | 2019-12-13 | Lab none | Installed | Bouldin Corn Heat Flux Plate B | Re-installed at Bouldin Corn after re-assembling tower. Need to check if this in A/B/C location. |
HFP01 11326 | 2019-12-13 | Lab none | Installed | Bouldin Corn Heat Flux Plate A | Re-installed at Bouldin Corn after re-assembling tower. Need to check if this in A/B/C location. |
CS655 17128 | 2019-12-13 | Lab none | Installed | Bouldin Corn Soil Moisture Probe B | Re-installed at Bouldin Corn after re-assembling tower. Need to check if this in A/B/C location. |
CS655 17129 | 2019-12-13 | Lab none | Installed | Bouldin Corn Soil Moisture Probe A | Re-installed at Bouldin Corn after re-assembling tower. Need to check if this in A/B/C location. |
CS655 17133 | 2019-12-13 | Lab none | Installed | Bouldin Corn Soil Moisture Probe C | Re-installed at Bouldin Corn after re-assembling tower. Need to check if this in A/B/C location. |
ML3 M003370 | 2019-12-13 | Lab none | Installed | Bouldin Corn Soil Moisture Probe Theta | Re-installed at Bouldin Corn after re-assembling tower. |
PQS 1 140456 | 2019-12-13 | Lab none | Installed | Bouldin Corn PAR Sensor Diffuse | Re-installed at Bouldin Corn after re-assembling tower as part of a diffuse PAR sensor. |
PQS 1 140453 | 2019-12-13 | Lab none | Installed | Bouldin Corn PAR Sensor Incoming | Re-installed at Bouldin Corn in incoming position after re-assembling tower. |
PQS 1 140441 | 2019-12-13 | Lab none | Installed | Bouldin Corn PAR Sensor Reflected | Re-installed at Bouldin Corn in outgoing position after re-assembling tower. |
SD130BN 0030F4D21491 | 2019-12-13 | Lab none | Installed | Bouldin Corn Camera | Re-installed at Bouldin Corn after re-assembling tower. |
SRS-Pr PRI 930103952 | 2019-12-13 | Lab none | Installed | Bouldin Corn Narrow Band PRI Reflected | Re-installed at Bouldin Corn after re-assembling tower. This field-stop PRI sensor is in the outgoing position. |
SRS-Pi PRI 913003004 | 2019-12-13 | Lab none | Installed | Bouldin Corn Narrow Band PRI Incoming | Re-installed at Bouldin Corn after re-assembling tower. This hemispherical PRI sensor was installed in the incoming position. |
SRS-Ni NDVI 2046503357 | 2019-12-13 | Lab none | Installed | Bouldin Corn Narrow Band NDVI Reflected | Re-installed at Bouldin Corn in the outgoing position after re-assembling tower. |
SRS-Ni NDVI 882103020 | 2019-12-13 | Lab none | Installed | Bouldin Corn Narrow Band NDVI Incoming | Re-installed at Bouldin Corn in the incoming position after re-assembling tower. |
Mark I A001359 | 2019-12-13 | Lab none | Installed | Bouldin Corn Weather Station B | Re-installed at Bouldin Corn after re-assembling tower. |
HMP45AC T2130005 | 2019-12-16 | Lab none | Installed | Sherman Barn RH & Temp | Installed at SB to replace sn sn U3030041, which had been removed and brought back to the lab for testing. |
WM 1590 074417 | 2019-12-02 | Bouldin Corn Anemometer | Removed | Lab none | One traducer missing. Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
PA-II-SD 84f3ebd5ffb1 | 2019-12-06 | Loaned | Returned | Lab none | Returned by Tomo Yoshino and friends from their undergraduate project looking at air quality in BART stations. |
CS547A 6494 | 2019-12-05 | East End Conductivity Sensor | Removed | Lab none | Removed from EE and brought back to the lab for drying and testing. The datalogger door was open when we arrived and the conductivity sensor interface was wet. |
CS547A 6494 | 2017-09-26 | Lab none | Installed | East End Conductivity Sensor | Installed at EE, after I'm assuming it tested fine in the lab. |
LI-7550 AIU-0614 | 2019-12-02 | Bouldin Corn Interface LI-7550 | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
LI-7500A 0042 | 2019-12-02 | Bouldin Corn Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
LI-7700 TG1-0155 | 2019-12-02 | Bouldin Corn Gas Analyzer CH4 | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
AM25T 7908 | 2019-12-02 | Bouldin Corn Multiplexer Thermocouple | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
AM16/32B 21690 | 2019-12-02 | Bouldin Corn Multiplexer Met | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
CR10X 31308 | 2019-12-02 | Bouldin Corn Data Logger Diffuse PAR | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
CR1000 66251 | 2019-12-02 | Bouldin Corn Data Logger Met | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
HMP60 L1740395 | 2019-12-02 | Bouldin Corn RH & Temp | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
PTB110 L2340502 | 2019-12-02 | Bouldin Corn Barometer | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
TE525MM 65548-615 | 2019-12-02 | Bouldin Corn Precipitation | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
URM06 URM06-001 | 2019-12-02 | Bouldin Corn Range Finder | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
CS451 70011234 | 2019-12-02 | Bouldin Corn Water Level | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
HFP01 10861 | 2019-12-02 | Bouldin Corn Heat Flux Plate C | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
HFP01 10867 | 2019-12-02 | Bouldin Corn Heat Flux Plate B | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
HFP01 11326 | 2019-12-02 | Bouldin Corn Heat Flux Plate A | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
CS655 17128 | 2019-12-02 | Bouldin Corn Soil Moisture Probe B | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
CS655 17129 | 2019-12-02 | Bouldin Corn Soil Moisture Probe A | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
CS655 17133 | 2019-12-02 | Bouldin Corn Soil Moisture Probe C | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
ML3 M003370 | 2019-12-02 | Bouldin Corn Soil Moisture Probe Theta | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
PQS 1 140456 | 2019-12-02 | Bouldin Corn PAR Sensor Diffuse | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
PQS 1 140453 | 2019-12-02 | Bouldin Corn PAR Sensor Incoming | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
PQS 1 140441 | 2019-12-02 | Bouldin Corn PAR Sensor Reflected | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
SD130BN 0030F4D21491 | 2019-12-02 | Bouldin Corn Camera | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
SRS-Pi PRI 913003004 | 2019-12-02 | Bouldin Corn Narrow Band PRI Incoming | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
SRS-Pi PRI 897103002 | 2019-12-02 | Bouldin Corn Narrow Band PRI Reflected | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
SRS-Ni NDVI 2046503357 | 2019-12-02 | Bouldin Corn Narrow Band NDVI Reflected | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
PRB-K Type #8 00000A6D72F5 | 2005-01-27 | Lab none | Installed | Tonzi | Installed at Tonzi by minirhizotron transect west of soil CO2 profiles |
PRB-K Type #8 00000A3D04E8 | 2005-01-27 | Lab none | Installed | Tonzi | Installed at Tonzi by minirhizotron transect west of soil CO2 profiles |
PRB-K Type #8 00000A6D882B | 2005-01-27 | Lab none | Installed | Tonzi | Installed at Tonzi by minirhizotron transect west of soil CO2 profiles |
PRB-K Type #8 0000097A4AAA | 2003-04-16 | Lab none | Installed | Tonzi | Installed at Tonzi closest to big tree by Liukang |
PRB-K Type #8 0000097A3B40 | 2003-04-16 | Lab none | Installed | Tonzi | Installed at Tonzi second closest to big tree by Liukang |
PRB-K Type #8 0000097A3C6E | 2003-04-16 | Lab none | Installed | Tonzi | Installed at Tonzi third closets to big tree by Liukang |
PRB-K Type #8 0000097A432E | 2003-04-16 | Lab none | Installed | Tonzi | Installed at Tonzi furthest from big tree by Liukang |
PRB-K Type #8 000007933FDE | 2001-11-10 | Lab none | Installed | Vaira Moisture Point Tower | Installed at Vaira near the eddy tower by Dennis and Liukang |
PRB-K Type #8 000000573FAA | 2001-11-10 | Lab none | Installed | Vaira Moisture Point Solar | Installed at Vaira near the solar panels by Dennis and Liukang |
PRB-K Type #8 000000574632 | 2001-11-10 | Lab none | Installed | Tonzi | Installed near the soil co2 measurements tower by Dennis and Liukang |
PRB-K Type #8 000000402FF3 | 2001-11-10 | Lab none | Installed | Tonzi | Installed near the water potential table and original sap flow measurements by Dennis and Liukang |
PRB-K Type #8 000000574B4B | 2001-11-10 | Lab none | Installed | Tonzi | Installed near the Floor solar panels by Dennis and Liukang |
PRB-K Type #8 000000402165 | 2001-11-10 | Lab none | Installed | Tonzi | Installed near the understory tower by Dennis and Liukang |
PRB-K Type #8 000000574B4B | 2019-12-03 | Tonzi | Removed | Lab none | Removed from Tonzi near the Floor solar panels. This probe has three broken pins and has not been measured in a long time but this was the first time the soil was wet enough to remove it by hand. |
LI-7500A 0042 | 2019-12-04 | Lab none | Calibration | Lab none | Last calibration: 2019-11-12 AGC=100.5 SDM= DAC1= DAC2= Ambient P=100.1kPa, Tair=20.3C, Tdew=10.11C Zero Air at 850psi, AMP CA4396 408.97 at 550 Tdewspan=20.00 CO2 zero 0.9481 -> .9490 Offset: 3.19ppm CO2 span 1.0091 -> 1.0182 Offset: -4.69ppm H2O zero 0.8449 -> .8441 Offset: -4.7ppm H2O span 1.0139 -> 1.0341 Offset: -.45ppm Notes: This sensor was on the BC tower when it fell down because of a wind/rain storm on 2019-11-26. First calibration since it was brought back to the lab. The CO2 span is larger than I expected. More testing needed before putting it back up in the field. |
SRS-Pr PRI 930103952 | 2019-12-04 | Lab none | Notes | Lab none | This sensor was not on the Bouldin Corn tower when we disassembled it on 2019-12-02. Although the equipment history says it was installed on the BC tower during initial set-up, the field note from the initial set-up on 2017-04-26 lists 2 other PRI sensors. I think this sensor was never put up on the BC tower and has remained in the lab. |
LI-7700 TG1-0155 | 2019-12-04 | Lab none | Calibration | Lab none | Calibration Location: Lab Last calibration: 2019-10-22 RSSI=87%, Optical RH=7% Ambient P=99.9kPa, Tair=20.2C Zero Air at 900psi, AMP CA4396 1.9497 at 650 CH4 Zero Offset: -0.091ppm CH4 Span Offset: 0.011ppm Notes: Testing this sensor after it fell down with the BC tower. First calibration. Offsets seem reasonable. May be a little less stable , or I was just impatient waiting for the span gas to equilibrate. |
SRS-Ni NDVI 882103020 | 2019-12-02 | Bouldin Corn Narrow Band NDVI Incoming | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
PA-II-SD 68c63acc07a9 | 2019-12-02 | Bouldin Corn Air Quality Sensor | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
Mark I A001359 | 2019-12-02 | Bouldin Corn Weather Station B | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
Mark II C002445 | 2019-12-02 | Bouldin Corn Weather Station C | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
WMP 1352 000019 | 2019-11-25 | Loaned | Returned | Lab none | Returned by David R. Cook at Argonne National Laboratory. |
NR-LITE 041556 | 2019-10-01 | Lab none | Calibration | Lab none | New calibration factor of 13.4 uV/(W/m2). Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 09/2018" for details. https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=27 |
NR-LITE 990354 | 2019-10-01 | Lab none | Calibration | Lab none | New calibration factor of 13.4 uV/(W/m2).
Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 09/2018" for details. https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=27 |
NR-LITE 990353 | 2019-10-01 | Lab none | Calibration | Lab none | New calibration factor of 12.6 uV/(W/m2).
Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 09/2018" for details. https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=27 I'm not sure why the calibration factor is 5% different from the previous calibration in the summer, and which is the "better" calibration to use. The summer had a wider range of Rnet values, but the sensors were more clean in the fall. I think the fall calibration is better. |
NR-LITE 990352 | 2019-10-01 | Lab none | Calibration | Lab none | New calibration factor of 14.5 uV/(W/m2).
Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 09/2018" for details. https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=27 I'm not sure why the calibration factor is 5% different from the previous calibration in the summer, and which is the "better" calibration to use. The summer had a wider range of Rnet values, but the sensors were more clean in the fall. I think the fall calibration is better. |
NR-LITE 990350 | 1999-07-29 | Factory | Calibration | Factory | Initial Factory calibration 13.5uV/W/m2 |