View history entry 4530 for Licor LI-7500A sn: 75H-2669

Date: 2020-05-22
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2020-02-04 AGC=100.7 SDM=0 DAC1=- DAC2=- Ambient P=100.2kPa, Tair=23.5C, Tdew=8.3C Zero Air at 750psi, AMP CA04396 408.97 at 600 Tdewspan=16.0 CO2 zero 0.9184 -> 0.9185 Offset: 0.44ppm CO2 span 1.016 -> 1.0124 Offset: 1.78ppm H2O zero 0.9006 -> 0.8999 Offset: -4.3ppm H2O span 1.0101 -> 1.0094 Offset: 0.01ppm Notes:
Entered by: