Equipment and Calibration History

View Equipment |

View history entry 4500 for Campbell Sci. CS547A sn: 5688

Date: 2020-05-26
From: Mayberry - Conductivity Sensor A
To: Mayberry - Conductivity Sensor A
Title: Calibration
Notes: Cleaned and calibrated in the field with DI water (0 mS) and a YSI solution (10,000 uS). Need to double check units -- are these the temperature-corrected mScm readings from the datalogger, where the temperature correction occurs in the datalogger program? In DI water: 0.54 (units?) In 10,000 uS calibration soln: 10.15 If unit is mScm, Calibration: (Corrected mScm) = 1.04 * (Sensor mScm) - 0.562
Entered by: Daphne Szutu