Equipment and Calibration History

View Equipment |

View history entry 4482 for Licor LI-7500A sn: 75H-2668

Date: 2020-05-20
From: Lab -
To: Lab - none
Title: Calibration
Notes: Last calibration: 2020-02-11 AGC=100.3 SDM=0 DAC1=- DAC2=- Ambient P=100.9kPa, Tair=20.4C, Tdew=9.7C Zero Air at 800psi, AMP CA04396 408.97 at 600 Tdewspan=16.0 CO2 zero 0.9393 -> 0.9390 Offset: -1.3ppm CO2 span 1.0171 -> 1.0198 Offset: -1.37ppm H2O zero 0.9072 -> 0.9082 Offset: 5.7ppm H2O span 1.0147 -> 1.0119 Offset: 0.05ppm Notes:
Entered by: