Skip to main content
rice field banner biometlab logo
   
   
 
   
   
   
   
   
 
 

Equipment and Calibration

View History

Equipment Entry

Instrument:Anemometer
Manufacture:Gill
Model No:WMP 1352
Serial No:000019
Tag No:991003713
Acquisition Date:0000-00-00
Project:
Location on 2021-07-06:Lab,
Notes:

Documents

No documents found

Report: Tonzi Tower Eddy System Compare - 2017-05-11

History

Rec No.DateLoc FromTitleLoc ToNotes
5361 2021-07-06 Borrowed none Returned Lab none Returned from FLEXLAB at LBL by Hershow Al-Barazi - Cable to be returned later hopefully.
4474 2020-05-22 Lab none Loaned Loaned none Loaned to Rod Mobini <rmobini@lbl.gov> at LBL FLEXLAB for an experiment on the indoor spread of droplets related to COVID-19 over the summer. The sonic was cleaned and tested okay in the lab. The settings were standardized to: B3 = 9600 baud
E1 = Full duplex
F1 = 8 data, no parity
G1 = No averaging
J2 = Two differential inputs
J5 = Zero single ended inputs
L1 = <cr><lf>
M1 = UVW ASCII message
NA = Name 'A'
O1 = comma separated
P6 = 10Hz
U1 = m/s
Z1 = Not used
X1 = align to axis
4198 2019-11-25 Loaned Returned Lab none Returned by David R. Cook at Argonne National Laboratory.
3484 2018-06-19 Lab none Loaned Loaned Loaned to David R. Cook at Argonne National Laboratory for a year or so. Two other sonics sn 000022, sn 000023 and a PSU sn 000193 went too
2786 2017-05-24 Tonzi-Tower Removed Lab none Removed from the Tonzi Tower. Replaced by the new Licor GHG system from Ameriflux.
1801 2015-04-29 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower to replace sn 000021 that has been having trouble with the serial stream dropping out.
551 2012-04-24 Mayberry Anemometer Removed Lab none Removed from the Mayberry eddy tower. Replaced with WM sonic sn 074417
1755 2011-04-26 Lab none Installed Mayberry Anemometer Installed on the Mayberry tower to replaced sn 000022 that had stopped sending data.
4761 2010-07-01 Sherman Island-Levee Anemometer Removed Lab none Removed from levee site due to theft of everything else.
4758 2010-04-06 Sherman Island Anemometer 2ndTower Moved Sherman Island-Levee Anemometer Moved along with the secondary tower from the Sherman cow pasture to the Sherman Levee location.
4757 2009-06-26 Lab none Installed Sherman Island Anemometer 2ndTower Installed the the secondary tower to replace sn 074417 which has a firmware issue that reduces the response rate of the SOS/T channel.
4912 2003-05-30 Tonzi-Floor Anemometer Floor Removed Lab none Removed from Tonzi-Floor due to no data. On power up it would send some data and then quit.
4893 2003-01-07 Lab none Installed Tonzi-Floor Anemometer Floor Installed at Tonzi-Floor to replace sn 000021.
2098 2001-01-05 Vaira Anemometer Removed Lab none Removed from the Vaira Tower due to going off scale at midday. Replaced by sn 000024
2097 2000-10-22 Lab none Installed Vaira Anemometer Installed on the Vaira Tower with the initial setup of the site.
 
 

Login