Equipment and Calibration History

View Equipment |

View history entry 4480 for Licor LI-7500A sn: 75H-2182

Date: 2020-05-20
From: Lab -
To: Lab - none
Title: Calibration
Notes: Last calibration: 2020-02-19 AGC=100.0 SDM=0 DAC1=- DAC2=- Ambient P=100.9kPa, Tair=21.1C, Tdew=10.5C Zero Air at 800psi, AMP CA04396 408.97 at 600 Tdewspan=61.0 CO2 zero 0.885 -> 0.8849 Offset: -0.07ppm CO2 span 1.0232 -> 1.0267 Offset: -1.72ppm H2O zero 0.8459 -> 0.8471 Offset: 7.5ppm H2O span 1.0249 -> 1.0159 Offset: 0.18ppm Notes:
Entered by: