Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 [31] 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68
Ser No.DateLoc FromTitleLoc ToNotes
TS01 TS01-001 2019-05-30 Lab none Installed Tonzi Installed near the Floor solar panels looking at the grass to the north. Footprint of about 0.7m x 0.35m diameter
LI-7550 AIU-1866 2019-05-31 Factory Received Lab none Received back from Licor. They replaced the RF chip in the power supply.
HMP45AC T2130005 2019-05-22 Lab none Tested Lab none I removed the power cord and soldered on an extension cord. Tested in the lab with radiation shields and fans, and compared well with two HMP60s sn L1740398 (Campbell has OK'd this one) and sn G4540023. This sensor is probably ok, but maybe we can test it with a reference AirT/RH sensor from AMP.
CS451 70011234 2018-12-10 Lab none Tested Lab none Tested in both air and water in the lab and saw an offset of about +0.975 cm. See RT report "Pressure transducer comparison 12-2018."
CS451 70011229 2018-12-10 Lab none Tested Lab none Tested in both air and water in the lab, but saw diurnal fluctuations of only 0.1 cm, not the 5-6 cm we saw in the field. See RT report "Pressure transducer comparison 12-2018."
CR1000 47108 2019-05-28 Sherman Barn Data Logger Met Notes Sherman Barn Data Logger Met VX2 may be unreliable. NR01 PRT was drifting constantly until it was re-wired from VX2 to VX3, but that could also have been a wiring issue with another channel.
AM16/32B 11662 2019-05-28 Sherman Barn Multiplexer Met Notes Sherman Barn Multiplexer Met Channel 28 and 32 are probably bad (see field notes from SB 2019-04-11).
LI-7550 AIU-1866 2019-05-09 Lab none Shipped Factory Shipped back to Licor for them to inspect.
LI-7500A 75H-2514 2019-05-07 Lab none Installed Tonzi-Floor Gas Analyzer CO2/H2O Installed on the Floor tower to replace the one that was moved to Tonzi Tower last week.
LI-7550 AIU-0666 2019-05-07 Lab none Installed Tonzi-Floor Interface LI-7500 Installed on the Floor tower to replace the one that was moved to Tonzi Tower last week.
LI-7500A 0035 2019-05-02 Lab none Installed East End Gas Analyzer CO2/H2O Installed at EE to replace sn 75H-2667, which was brought back to the lab for calibration.
LI-7500A 75H-2667 2019-05-02 East End Gas Analyzer CO2/H2O Removed Lab none Removed from EE and brought back to the lab for calibration. it was replaced by sn 0035.
LI-7500A 0042 2019-04-30 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from the Tower for calibration. Replaced by sn 1594. The Tower 7550 box died and was replaced by the Floor 7550 box. No reason to believe the Head was damaged by the box crash but might be worth checking.
LI-7500A 0041 2019-04-30 Tonzi-Floor Gas Analyzer CO2/H2O Removed Lab none Removed from Floor tower for calibration. Not replaced because the Floor 7550 box was moved to the Tower to replace the 7550 box there that died.
LI-7500A 1594 2019-04-30 Lab none Installed Tonzi-Tower Installed on the Tower to replace sn 0042 for calibration. Using Floor 7550 box because Tower 7550 box died
LI-7550 AIU-1869 2019-04-30 Tonzi-Floor Interface LI-7500 Moved Tonzi-Tower Gas Analyzer CO2/H2O Moved to the Tower to replace sn AIU-1866 which had some kind of power problem and failed.
LI-7550 AIU-1866 2019-04-30 Tonzi-Tower Interface LI-7550 Removed Lab Interface LI-7550 Removed from Tonzi Tower due to power issues. Power light was flickering with audible relay click when a head was attached. Seemed to boot without head. Might have caused 5A Tower load breaker to pop on 4-23. Brought in for testing - replaced with Floor box sn AIU-1869
LI-7500A 75H-2668 2019-04-30 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace sn 74H-2176 for calibration.
LI-7500A 75H-2176 2019-04-30 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 75H-2668.
LI-7550 AIU-1866 2019-05-01 Lab none Tested Lab none Plugged it in to the 24V wall power and it was dead. Will e-mail Licor.
RF401 35754 2019-03-11 Lab none Installed Tonzi Installed in the OSU SP3 box to connect to the Tower computer.
RF401 34362 2019-03-11 Lab none Installed Tonzi Installed in the Tower computer box to connect the computer to the remote data loggers.
miniDOT 1219 2016-10-11 Mayberry DO Sensor E Moved Mayberry DO Sensor D Moved from middle position to upper position (top sensor could not be removed so it was moved to lower position; bottom/lower sensor was removed).
miniDOT 1220 2018-10-04 Mayberry DO Sensor E Notes Mayberry DO Sensor E Measured at -45 cm.
miniDOT 1220 2017-10-13 Lab none Installed Mayberry DO Sensor E After the barnacle was removed in the lab, reinstalled at MB on the floating boom, lower position.
miniDOT 1220 2017-08-16 Mayberry DO Sensor E Removed Lab none Removed from lower position (-62 cm) on floating boom and brought in to the lab for barnacle removal.
miniDOT 1221 2018-10-04 Mayberry DO Sensor D Notes Mayberry DO Sensor D Measured at -15 cm.
miniDOT 1221 2017-05-25 Mayberry DO Sensor D Moved Mayberry DO Sensor D Moved from static boom (upper position, -10 cm based on water level on 2017-05-10) to floating boom with sn 1219 (upper position, -17 cm).
miniDOT 1219 2017-05-25 Mayberry DO Sensor D Moved Mayberry DO Sensor D Moved from static boom (upper position, -10 cm based on water level on 2017-05-10) to floating boom with sn 1221 (upper position, -17 cm).
miniDOT 1220 2017-05-25 Mayberry DO Sensor D Moved Mayberry DO Sensor E Moved from static boom (upper position, -10 cm based on water level on 2017-05-10) to floating boom (lower position, -62 cm).
miniDOT 1219 2015-04-23 Mayberry DO Sensor E Moved Mayberry DO Sensor E Sensor was moved up 10 cm on fence post: from -65 cm (based on water level 2014-10-09) to -50 cm (based on water level on 2015-04-23).
miniDOT 1221 2015-04-23 Mayberry DO Sensor F Moved Mayberry DO Sensor F Sensor was moved up 15 cm on the fence post: from -109 cm (based on water level 2014-10-09) to -92 cm (based on water level on 2015-04-23).
miniDOT 1221 2017-05-10 Mayberry DO Sensor D Mayberry DO Sensor D sn 1221 moved slightly in the upper position from -8cm (based on water level on 2017-02-23) to -10cm (based on water level on 2017-05-10) with the other 2 miniDOTs.
miniDOT 1220 2017-05-10 Mayberry DO Sensor E Mayberry DO Sensor D sn 1220 moved from middle position (-38 cm based on water level on 2017-02-23) to upper position (-10cm based on water level on 2017-05-10) with the other 2 miniDOTs.
miniDOT 1219 2017-05-10 Mayberry DO Sensor F Moved Mayberry DO Sensor D sn 1219 moved from lower position (-68 cm based on water level on 2017-02-23) to upper position (-10cm based on water level on 2017-05-10). All 3 miniDOT sensors (sn 1219, sn 1220, sn 1221) are at this depth for now.
HMP45AC U3030046 2019-04-29 Lab none Tested Lab none Continued testing in lab and RH is consistently bad. Do not use.
LI-7500A 0042 2004-04-07 Lab none Calibration Lab none Calibrated in the lab after being recently returned from Licor. There is a change in the power wiring. Red/white is positive, green is negative. They also changed the delay to 240 ms.
LI-7500A 0035 2004-04-07 Lab none Calibration Lab none Calibrated in the lab after being recently returned from Licor. There is a change in the power wiring. Red/white is positive, green is negative. They also changed the delay to 240 ms.
URM06 URM06-002 2019-04-25 Lab none Installed Mayberry Installed on the Mayberry Tower on a short arm off the center-east side about 68cm above the water level.
LI-7700 TG1-0223 2019-04-25 Mayberry Gas Analyzer CH4 Upgrade Mayberry Gas Analyzer CH4 The thermocouple module was broken so we swapped it for a new one. The LI-7700 firmware was upgraded from v1.0.21 to v1.0.27 mostly minor bug fixes see change log https://www.licor.com/env/support/LI-7700/topics/change-log.html
LI-7500A 0418 2019-04-25 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed on West Pond to replace sn 75H-2668 for calibartion
LI-7500A 75H-2668 2019-04-25 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from West Pond for calibration. Replaced by sn 0418
LI-7700 TG1-0224 2019-04-25 West Pond Gas Analyzer CH4 Upgrade West Pond Gas Analyzer CH4 The LI-7700 firmware was upgraded from v1.0.21 to v1.0.27 mostly minor bug fixes see change log https://www.licor.com/env/support/LI-7700/topics/change-log.html
URM06 URM06-001 2019-04-25 Lab none Installed Bouldin Alfalfa Installed on the Boldin Alfalfa radiation boom about 262cm above the ground to measure veg height.
LI-7500A 75H-2180 2019-04-25 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed on Bouldin Alfalfa to replace sn 0035 for calibration
LI-7500A 0035 2019-04-25 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Alfalfa for calibration. Replaced by sn 75H-2180.
LI-7500A 1594 2019-04-26 Lab none Calibration Lab none Last calibration: 2019-01-30
AGC=99.5 SDM= DAC1= DAC2=
Ambient P=100.4kPa, Tair=24.1C, Tdew=10.2C
Zero Air at 1050psi, AMP CA4396 408.97 at 700 Tdewspan=16.00
CO2 zero 0.8643 -> .8646 Offset: 1.33ppm
CO2 span 1.0157 -> 1.0178 Offset: -1.16ppm
H2O zero 0.8891 -> .8868 Offset: -12.6ppm
H2O span 1.0043 -> 1.0026 Offset: .04ppm
Notes:
LI-7500A 0035 2019-04-26 Lab none Calibration Lab none Last calibration: 2019-01-28
AGC=99.2 SDM= DAC1= DAC2=
Ambient P=100.4kPa, Tair=22.6C, Tdew=10.7C
Zero Air at 1100psi, AMP CA4396 408.97 at 700 Tdewspan=16.00
CO2 zero 0.9375 -> .9371 Offset: -1.79ppm
CO2 span 1.0193 -> 1.0189 Offset: .25ppm
H2O zero 0.8969 -> .8989 Offset: 12.3ppm
H2O span 1.0023 -> 1.0186 Offset: -.34ppm
Notes:
LI-7500A 0073 2019-04-25 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed on Bouldin Corn to replace sn 1594 for calibration
LI-7500A 1594 2019-04-25 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 0073.
LI-7700 TG1-0155 2019-04-25 Bouldin Corn Gas Analyzer CH4 Upgrade Bouldin Corn Gas Analyzer CH4 The LI-7700 firmware was upgraded from v1.0.21 to v1.0.27 mostly minor bug fixes see change log https://www.licor.com/env/support/LI-7700/topics/change-log.html
LI-7500A 75H-2668 2019-04-26 Lab none Calibration Lab none Last calibration: 2019-02-11
AGC=100.0 SDM= DAC1= DAC2=
Ambient P=100.4kPa, Tair=24.5C, Tdew=11.1C
Zero Air at 1050psi, AMP CA4396 408.97 at 700 Tdewspan=16.00
CO2 zero 0.939 -> .9391 Offset: .22ppm
CO2 span 0.018 -> 1.0169 Offset: .59ppm
H2O zero 0.9073 -> .9069 Offset: -2.2ppm
H2O span 1.0117 -> 1.0139 Offset: -.05ppm
Notes:
LI-7700 TG1-0223 2019-04-25 Mayberry Gas Analyzer CH4 Calibration Mayberry Gas Analyzer CH4 Calibration Location: MB
Last calibration: 2018-10-26
RSSI=80%, Optical RH=12%
Ambient P=100.9kPa, Tair=26C
Zero Air at 1600psi, AMP CA3923 3.121 at 1600
CH4 Zero Offset: 0.013ppm
CH4 Span Offset: 0.015ppm
Notes: TC module swapped before calibration; breezy day
HMP45AC T2130005 2019-04-24 Lab none Tested Lab none Tested overnight in the lab. This sensor has an extension cable soldered on. Didn't see any spikes, and the numbers seem reasonable. However, the control wire seems extraneous--the AirT and RH numbers seem ok even when the control wire is not connected to the datalogger, even though the sensor still has the power control board. More testing needed before we deploy it.
HMP45AC U3030046 2019-04-24 Lab none Tested Lab none Tested overnight in the lab. This sensor has an extension cable soldered on. The RH was consistently 5-7% lower than the other 3 sensors (L1740398, T2130005, G4540023). I removed the extension cable, and the numbers were still low. I removed the humidity chip, and it read 150%. When I replaced the humidity chip, the RH was within 2-4% of the other 3 sensors. I added the same extension cable back on, and the numbers became low. This sensor doesn't look consistent and needs more testing before it can be deployed.
CR10X 47772 2019-04-16 Tonzi Removed Lab none Removed with the wooden cal panel for the FLIR because cows chewed the TC wires and the new cal panel is better
SI-131 7385 2019-04-16 Tonzi Removed Lab none Removed with the wooden cal panel for the FLIR because cows chewed the TC wires and the new cal panel is better
LI-7500A 75H-2180 2019-04-23 Lab none Calibration Lab none Last calibration: 2019-01-25
AGC=102 SDM= DAC1= DAC2=
Ambient P=100.6kPa, Tair=24.7C, Tdew=10.9C
Zero Air at 1100psi, AMP Air 408.97 at 700 Tdewspan=16.00
CO2 zero 0.9137 -> .9136 Offset: -.34ppm
CO2 span 1.0262 -> 1.0266 Offset: -.2ppm
H2O zero 0.8891 -> .8886 Offset: -3.0ppm
H2O span 1.0199 -> 1.0236 Offset: -.07ppm
Notes:
LI-7500A 0073 2019-04-23 Lab none Calibration Lab none Last calibration: 2019-01-07
AGC=98.6 SDM= DAC1= DAC2=
Ambient P=100.4kPa, Tair=26.0C, Tdew=11.8C
Zero Air at 1100psi, AMP Air 408.97 at 700 Tdewspan=16.00
CO2 zero 0.956 -> .9563 Offset: .91ppm
CO2 span 1.0081 -> 1.0094 Offset: -0.78ppm
H2O zero 0.7316 -> .7336 Offset: 12.1ppm
H2O span 1.0023 -> 1.0153 Offset: -.26ppm
Notes:
LI-7500A 0418 2019-04-22 Lab none Calibration Lab none Last calibration: 2019-01-07
AGC=95 SDM= DAC1= DAC2=
Ambient P=100.6kPa, Tair=24.2C, Tdew=11.4C
Zero Air at 1100psi, AMP Air 408.97 at 700 Tdewspan=16.00
CO2 zero 0.8992 -> .8992 Offset: -.04ppm
CO2 span 1.0033 -> 1.0011 Offset: 1.21ppm
H2O zero 0.8465 -> .8464 Offset: -.4ppm
H2O span 0.9832 -> .9866 Offset: -.07ppm
Notes:
LI-7500A 75H-2514 2019-04-22 Lab none Calibration Lab none Last calibration: 2019-01-07
AGC=102 SDM= DAC1= DAC2=
Ambient P=100.6kPa, Tair=24.7C, Tdew=10.9C
Zero Air at 1100psi, AMP Air 408.97 at 700 Tdewspan=16.00
CO2 zero 0.9051 -> .905 Offset: -.69ppm
CO2 span 1.0167 -> 1.01 Offset: 3.39ppm
H2O zero 0.8397 -> .8389 Offset: -5.5ppm
H2O span 1.0177 -> 1.0155 Offset: .05ppm
Notes:
LI-7500A 75H-2669 2019-04-18 Lab none Installed Sherman Barn Gas Analyzer CO2/H2O Installed at SB to replace sn 75H-2180, which was brought back to the lab for calibration.
LI-7500A 75H-2180 2019-04-18 Sherman Barn Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from SB and brought back to the lab for calibration. It was replaced by sn 75H-2669.
PA-II-SD 68c63acc07a9 2019-04-18 Lab none Installed Bouldin Corn Air Quality Sensor Installed again at BC after checking the new microSD card and the sensor were ok.
CS451 70011234 2019-04-11 Lab none Installed Bouldin Corn Water Level Installed in a well underneath the BC tower. Sensor is 92cm below ground level. From field notes:
Top of pipe to ground = 25cm
Top of pipe to sensor = 117cm
Top of pipe to water = 85cm
Sensor reading = 0.46 psi ~ 32cm (matches manual reading)
PA-II-SD 68c63acc07a9 2019-04-11 Bouldin Corn Air Quality Sensor Removed Lab none Removed from BC and brought back to the lab for testing. The microSD card was cracked, but the sensor itself was fine.
LI-7500A 75H-2150 2019-04-18 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed at SW to replace sn 0418, which was brought back to the lab for calibration.
LI-7500A 0418 2019-04-18 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed from SW and brought back to the lab for calibration. It was replaced by sn 75H-2150.
LI-7500A 0065 2019-04-18 Lab none Installed East Pond Gas Analyzer CO2/H2O Installed at EP to replace sn 0073, which was brought back to the lab for calibration.
LI-7500A 0073 2019-04-18 East Pond Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from EP and brought back to the lab for calibration. It was replaced by sn 0065.
LI-7500A 75H-2182 2019-04-18 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at MB to replace sn 75H-2514, which was brought back to the lab for calibration.
LI-7500A 75H-2514 2019-04-18 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from MB and brought back to the lab for calibration. It was replaced by sn 75H-2182.
RF401 35785 2019-04-16 Lab none Installed Tonzi Install in the OSU SP1 box near the soil co2 system to connect the CR3000 to the Tower computer. Using Pakbus mode at 38400
WMP 1352 000022 2019-04-17 Loaned Returned Lab none Returned from David R. Cook at Argonne National Laboratory by way of LBL Ameriflux. Apparently David forgot who loaned them the sonic. Also it was reported that the sonic wasn't working. I had left a "serial output problem" note on its side but it seems to be spitting 10Hz binary data at 9600 baud.
LI-7500A 0065 2019-04-15 Lab none Calibration Lab none Last calibration: 2019-02-11
AGC=99.2 SDM=0 DAC1=- DAC2=-
Ambient P=100.16kPa, Tair=27.11C, Tdew=6.82C
Zero Air 369068 at 1100psi, AMP CA04396 408.97 at 750 Tdewspan=15.99
CO2 zero 0.892 -> 0.8911 Offset: -4.4ppm
CO2 span 1.0073 -> 1.0171 Offset: -5.12ppm
H2O zero 0.7263 -> 0.7260 Offset: -2.4ppm
H2O span 1.0104 -> 1.0118 Offset: -0.02ppm
Notes: First lab calibration since factory calibration. I manually typed in the new coefficients and zero/span values. I saved a config file then opened it in a text editor and changed the calibration dates, standard gas values and WX coefficient. I saved and uploaded this config. This IRGA now has the SD* temperature coefficients and its signal strength was reset at the factory. Our zeros were lower than the factory calibration so I calibrated and saved a new lab calibration config file.
LI-7500A 75H-2182 2019-04-15 Lab none Calibration Lab none Last calibration: 2018-10-24
AGC=87.6 SDM=0 DAC1=- DAC2=-
Ambient P=100.19kPa, Tair=26.94C, Tdew=7.42C
Zero Air 369068 at 1100psi, AMP CA04396 408.97 at 750 Tdewspan=15.99
CO2 zero 0.8844 -> 0.8842 Offset: -1.10ppm
CO2 span 1.0188 -> 1.0156 Offset: 0.08ppm
H2O zero 0.8449 -> 0.8459 Offset: 6.3ppm
H2O span 1.0143 -> 1.0189 Offset: -0.1ppm
Notes: by Joe
LI-7500A 75H-2150 2019-04-15 Lab none Calibration Lab none Last calibration: 2018-10-24
AGC=97.2 SDM=0 DAC1=- DAC2=-
Ambient P=100.19kPa, Tair=26.84C, Tdew=7.32C
Zero Air 369068 at 1100psi, AMP CA04396 408.97 at 750 Tdewspan=15.99
CO2 zero 0.9021 -> 0.9014 Offset: -3.16ppm
CO2 span 1.005 -> 1.0030 Offset: 0.86ppm
H2O zero 0.866 -> 0.8671 Offset: 8.0ppm
H2O span 1.0069 -> 1.0077 Offset: -0.02ppm
Notes: By Joe
LI-7500A 75H-2669 2019-04-15 Lab none Calibration Lab none Last calibration: 2018-10-10
AGC=103.8 SDM=0 DAC1=- DAC2=-
Ambient P=100.16kPa, Tair=24.85C, Tdew=6.62C
Zero Air 369068 at 1150psi, AMP CA04396 408.97 at 800 Tdewspan=15.99
CO2 zero 0.9188 -> 0.9186 Offset: -1.10ppm
CO2 span 1.0106 -> 1.0134 Offset: -1.52ppm
H2O zero 0.8992 -> 0.8991 Offset: -0.3ppm
H2O span 1.0034 -> 1.0017 Offset: 0.03ppm
Notes: Very stable - Joe
LI-7700 TG1-0421 2019-04-11 Bouldin Alfalfa Gas Analyzer CH4 Upgrade Bouldin Alfalfa Gas Analyzer CH4 Upgraded LI-7700 sn TG1-0421 firmware from 1.0.23 to 1.0.27. To connect, Joe had to change the laptop network subnet from 192.168.2.5 to 192.168.1.5 (to match the LI-7700 IP address of 192.168.1.77), even though the subnet mask was 255.255.255.0. We then restarted the instrument and calibrated LI-7700 sn TG1-0421.
USB2000-VIS-NIR USB2G4461 2016-10-11 East End Calibration East End Calibration Location: East End
Last calibration: 2016-04-13
RSSI=82%, Optical RH=21%
Ambient P=101.6kPa, Tair=21.1C
zero air at 800psi, AMP CH4 4.03 at 1400
CH4 Zero Offset: 0.02ppm
CH4 Span Offset: 0.01ppm
Notes: Breezy
LI-7700 TG1-0357 2019-04-11 Sherman Barn Gas Analyzer CH4 Calibration Sherman Barn Gas Analyzer CH4 Calibration Location: Sherman Barn
Last calibration: 2018-07-10
RSSI=70%, Optical RH=3%
Ambient P=101.9kPa, Tair=16.5C
Zero Air M4002 at 1800psi, AMP CA03923 3.121 at 1800
CH4 Zero Offset: 0.035ppm
CH4 Span Offset: -0.002ppm
Notes: Windy 3m/s to 7.5m/s
LI-7700 TG1-0211 2019-04-11 East Pond Gas Analyzer CH4 Calibration East Pond Gas Analyzer CH4 Calibration Location: East Pond
Last calibration: 2018-10-26
RSSI=70%, Optical RH=14%
Ambient P=101.9kPa, Tair=17.1C
Zero Air M4002 at 1600psi, AMP CA03923 3.121 at 1800
CH4 Zero Offset: 0.014ppm
CH4 Span Offset: 0.003ppm
Notes: Windy day
LI-7700 TG1-0155 2019-04-11 Bouldin Corn Gas Analyzer CH4 Calibration Bouldin Corn Gas Analyzer CH4 Calibration Location: Bouldin Corn
Last calibration: 2018-10-25
RSSI=87%, Optical RH=22%
Ambient P=101.5kPa, Tair=19.1C
Zero Air M4002 at 1600psi, AMP C03923 3.121 at 1600
CH4 Zero Offset: 0.024ppm
CH4 Span Offset: -0.007ppm
Notes: Windy
LI-7700 TG1-0424 2019-04-11 Sherman Wetland Gas Analyzer CH4 Calibration Sherman Wetland Gas Analyzer CH4 Calibration Location: Sherman Wetland
Last calibration: 2018-10-24
RSSI=77%, Optical RH=13%
Ambient P=102.0kPa, Tair=15.4C
Zero Air M4002 at 1800psi, AMP CA03923 3.121 at 1900
CH4 Zero Offset: 0.051ppm
CH4 Span Offset: 0.024ppm
Notes: Windy day 3m/s to 7m/s
LI-7700 TG1-0423 2019-04-11 East End Gas Analyzer CH4 Calibration East End Gas Analyzer CH4 Calibration Location: East End
Last calibration: 2018-10-30
RSSI=73%, Optical RH=11%
Ambient P=101.9kPa, Tair=16.5C
Zero Air M4002 at 1700psi, AMP CA03923 3.121 at 1800
CH4 Zero Offset: 0.036ppm
CH4 Span Offset: 0.013ppm
Notes: Windy day
LI-7700 TG1-0421 2019-04-11 Bouldin Alfalfa Gas Analyzer CH4 Calibration Bouldin Alfalfa Gas Analyzer CH4 Calibration Location: Bouldin Alfalfa
Last calibration: 2018-11-06
RSSI=83%, Optical RH=17%
Ambient P=101.3kPa, Tair=19.2C
Zero Air M4002 at 1600psi, AMP CA03923 3.121 at 1600
CH4 Zero Offset: 0.037ppm
CH4 Span Offset: 0.007ppm
Notes: Breezy day, upgraded firmware from v1.0.23 to v1.0.27 before calibration
LI-7700 TG1-0224 2019-04-11 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration: 2018-10-25
RSSI=71%, Optical RH=24%
Ambient P=101.8kPa, Tair=17.2C
Zero Air M4002 at 1600psi, AMP CA03923 3.1210 at 1700
CH4 Zero Offset: -0.005ppm
CH4 Span Offset: 0.008ppm
Notes: Windy day. Optical RH high - needs new desiccant.
PRB-K Type #8 00000128B7B90 2019-03-29 Vaira Moisture Point Gate Removed Lab none Removed from the location near the gate. This probe has two broken pins top and bottom and is no longer making readings.
LI-6400 PSC-0785 2019-03-25 Lab none Calibration Lab none Camilo Rey-Sanchez and Daphne Szutu

Camilo cleaned the light-source leaf chamber (it wasn't too dusty) and attached it to this console for calibration. We connected the reference and sample inlet to the cal gases with a y-junction. Camilo had checked the zeros using the chemicals last week and found them unsatisfactory (we didn't write them down). We set the zero and the span using cal gas and dewpoint generator.

Zero:
Cal Gas (zero air), flow at 1SCFH:
Before: CO2_S:~-15 CO2_R: ~-15 H2O_S: ~-2 H2O_R: ~-2
After: *All fluctuating around zero.

Span with AMP 408.97ppm cal gas, flow at 1SCFH:
Before: CO2_R: 413 CO2_S: 423
After: *Fluctuating around 409

Span with 16.0C dew point (2019-03-26), flow at 500 CC/min:
Before: H2O_R: 17.50 C H2O_S: 18.52 C
After: H2O_R: 16.00 C H2O_S: 16.00 C
LI-6400 PSC-2474 2019-03-25 Lab none Calibration Lab none Camilo Rey-Sanchez and Daphne Szutu

Camilo attached a leaf chamber with a transparent top. He cleaned the chamber, but it wasn't as dusty as he suspected. I changed the time from 15:26 to 13:38. We connected the reference and sample inlet to the cal gases/dewpoint generator with a y-junction. We checked the zero using the chemicals and they seemed ok, but we decided to calibrate zero/spans using the the cal gases and dewpoint generator anyways.

Zero:
Checked using chemicals: CO2_S: 1.4 CO2_R: -0.1 H2O_S: 0.55 H2O_R: 1.4

Set zero using cal gas:
Cal gas = zero air, 1100 psi, 1 SCFH flow
Before: CO2_S:-0.9 CO2_R: 4.2 H2O_S: -0.043 H2O_R: -0.592
After: *All fluctuating around zero.

Span with 408.97ppm cal gas from AMP:
Cal gas = AMP cal gas at 408.97ppm, 800 psi, 1 SCFH flow
Before: CO2_R: 408.7 CO2_S: 407.3
After: CO2_R: Did not adjust CO2_S: 409.0

Span with 16.0C dew point:
Before: H2O_R: ?? C H2O_S:?? C
After: ??
CR1000 66252 2019-03-14 Bouldin Alfalfa Data Logger Met Maintenance Bouldin Alfalfa Data Logger Met The CR1000 packbus id was change from 1 to 52
2300-000 2300565 2019-02-22 Mayberry Fluorometer Removed Lab none Removed from MB and brought back to the lab to swap out its brushes and other maintenace.
LI-7500A 0065 2019-02-11 Lab none Calibration Lab none Last calibration: 2018-09-26
AGC= SDM= DAC1= DAC2=
Ambient P=kPa, Tair=C, Tdew=C
at psi, at Tdewspan=12.00
CO2 zero 0.8903 -> .8920 Offset: ppm
CO2 span 1.0184 -> 1.0073 Offset: ppm
H2O zero 0.7086 -> 07263 Offset: ppm
H2O span 1.0155 -> 1.0104 Offset: ppm
Notes: CX=34338.1, WX=34091.7. Calibrated at Licor factory after source/chopper replacement.
LI-7500A 0065 2019-02-22 Factory Received Lab none Received from Licor after repair and calibration. Service notes: Inspected the unit and found the source lamp intermittent. Replaced the defective source and the chopper motor per the customer's request. Subjected the unit to environmental tests and calibrated.
LI-7500A 0065 2019-01-21 Lab none Shipped Factory Shipped back to factory for repair. Received by Licor on 2019-01-21.
SRS-Ni NDVI 1234903795 2019-02-20 Lab none Installed Vaira Narrow Band Reflected Installed on the Vaira Tower in the reflected position.
SRS-Ni NDVI 1334903221 2019-02-20 Lab none Installed Vaira Narrow Band Incoming Installed on the Vaira Tower in the incoming position
PA-II-SD 68c63acc07a9 2019-02-12 Lab none Installed Bouldin Corn Air Quality Sensor Installed at BC. It is bolted to the bottom side of the upper boardwalk.
LI-7700 TG1-0421 2019-02-12 Bouldin Alfalfa Gas Analyzer CH4 Tested Bouldin Alfalfa Gas Analyzer CH4 Tried to upgrade firmware from 1.0.23 to 1.0.27, but there was an error, "Fatal: Network Error: Connection Timed out."
LI-7700 TG1-0424 2019-02-12 Sherman Wetland Gas Analyzer CH4 Upgrade Sherman Wetland Gas Analyzer CH4 Upgraded firmware (embedded software) from 1.0.23 to 1.0.27. Licor change log: "Expanded a buffer in the DSP to prevent phase changes in the DSP data caused when the buffer filled up. This fixes the synchronization loss and resultant data loss between the LI-7550 and LI-7700."