Equipment and CalibrationView History
Equipment Entry
Instrument: | Precipitation |
Manufacture: | Texas Electronics |
Model No: | TE525 |
Serial No: | 48102-611 |
Tag No: | |
Acquisition Date: | 2018-06-26 |
Project: | OSU AirMOSS |
Location on 2021-10-06: | Tonzi, |
Notes: |
Acquired from the OSU AirMOSS project, Location SP03-23 |
|
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
5477 |
2021-10-06 |
Tonzi |
Calibration |
Tonzi none |
Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 50 tips – No adjustment |
4144 |
2019-11-07 |
Tonzi |
Calibration |
Tonzi |
Plug tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket which should result in 100tips. Got 103 tips and did not adjust the bucket. |
3617 |
2012-01-09 |
Lab none |
Installed |
Tonzi |
Installed on the ground at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ. |
|
|
|