Equipment and Calibration
View History
Equipment Entry
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6292 | 2023-05-09 | Bouldin Corn Forced Diffusion C | Moved | Bouldin Alfalfa Forced Diffusion A | Moved from BC to BA to replace sn FD20160001 which broke. We should have soil CO2 flux data from Whendee's chamber measurements at BC. |
6246 | 2023-03-08 | Lab none | Installed | Bouldin Corn Forced Diffusion C | Installed under BC tower to replace sn FD20180067 which got flooded |
6198 | 2023-02-28 | Factory none | Returned | Lab none | Received after repair at Eosense: CO2 sensor replaced, relay board replaced, pump replaced, concentration calibration, flux calibration |
6199 | 2022-08-23 | Lab none | Shipped | Factory none | Shipped to Eosense on RMA220809 for repair |
5444 | 2021-09-01 | Tonzi-SoilCO2 Forced Diffusion Open | Removed | Lab none | Removed from Tonzi-SoilCO2 due to consistent offset between soil and atmosphere chambers. This resulted in small fluxes when they should be pretty much zero. Replaced by sn FD20190027. |
5407 | 2021-08-11 | Tonzi-SoilCO2 Forced Diffusion Open | Maintenance | Tonzi-SoilCO2 Forced Diffusion Open | Performance check done. This was a double check of the last check since the CO2 concentrations were so far from ambient. The two chambers were still offset: Average Flux: 1.3, 0.37 STDEV Average ATM: 343.7ppm, 41.5C Average Soil: 408.5ppm, 41.5C Need to ask Eosense again. |
5370 | 2021-07-13 | Tonzi-SoilCO2 Forced Diffusion Open | Maintenance | Tonzi-SoilCO2 Forced Diffusion Open | Performance check done. Referenced to 400ppm. Large offsets need to be double checked: Average Flux: 0.8, 0.21 STDEV Average ATM: 107.4ppm, 40.2C Average Soil: 146.1ppm, 40.3C |
4174 | 2019-11-19 | Lab none | Installed | Tonzi Forced Diffusion Open | Installed in the open at the Tonzi soil co2 profiles |