Equipment and Calibration History

View Equipment |

View history entry 4154 for Licor LI-7500A sn: 75H-2182

Date: 2019-11-18
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2019-08-01 AGC=100.0 SDM=0 DAC1=- DAC2=- Ambient P=100.3kPa, Tair=27.5C, Tdew=9.64C Zero Air at 950psi, AMP CA04396 408.97 at 600 Tdewspan=20.0 CO2 zero 0.8846 -> 0.8846 Offset: -0.05ppm CO2 span 1.0226 -> 1.0217 Offset: 0.33ppm H2O zero 0.8456 -> 0.8462 Offset: 3.6ppm H2O span 1.0214 -> 1.0265 Offset: -0.1ppm Notes:
Entered by: