<--2019-10-31 09:20:00 | 2019-12-05 09:15:00-->Other sites visited today: Mayberry | Sherman Wetland | Bouldin Corn | Vaira | Tonzi | Tonzi
Phenocam link: Searching...
Sherman Barn: 2019-11-19 10:40:00 (DOY 323)
Author: Daphne Szutu
Others:
Summary: Regular data collection/cleaning, spectral, AirT reading bad - brought HMP back to lab, tested and adjusted tipping bucket
2019-11-19 Sherman Barn I arrived at 10:40. Ariane was at Mayberry collecting water samples. It was a windy day with clear skies. The grass around the tower is all brown and short. There were some cows north of the tower. I downloaded met, USB GHG, and game camera data. I forgot to get the phenocam data. I reset the game camera clock. I cleaned flux and rad sensors and refilled the wash reservoir from 1/3 full. The HMP45 (sn U3030041) AirT reading had been hovering around 2 degC for the past few days(RH looked ok). It wasn't an obvious power or wiring problem, so I brought it back to the lab for testing. This sensor also had AirT spikes a while back. I took spectral measurements. I tested the tipping bucket sn 49187-109. The funnel looked clean, so I connected it to a CR23x and dripped 475ml of water through the bucket, which should result in 100tips--got 104 tips. I turned both set screws ½ rotation CCW. It should be retested at the next site visit. LI-7500 readings: CO2 402ppm H2O 524ppm T 18.1C P 100.9kPa SS 94--> 100 after cleaning LI7700 readings: CH4 1.90ppm RSSI 17-->78 after cleaning I left at 11:40. |