Equipment and Calibration History

View Equipment |

View history entry 4153 for Licor LI-7500A sn: 0035

Date: 2019-11-18
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2019-08-01 AGC=100.1 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=26.4C, Tdew=10.0C Zero Air at 950psi, AMP CA04396 408.97 at 600 Tdewspan=20.0 CO2 zero 0.9374 -> 0.9374 Offset: 0.0ppm CO2 span 1.0207 -> 1.0170 Offset: 1.93ppm H2O zero 0.8984 -> 0.8980 Offset: -2.6ppm H2O span 1.0323 -> 1.0314 Offset: 0.03ppm Notes: by Joe
Entered by: