Equipment and Calibration History
Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 [23] 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70Ser No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
LI-7700 TG1-0211 | 2020-11-19 | Lab none | Installed | Bouldin Alfalfa Gas Analyzer CH4 | Installed at Bouldin Alfalfa to replace sn TG1-0421 for calibration. |
LI-7700 TG1-0421 | 2020-11-19 | Bouldin Alfalfa Gas Analyzer CH4 | Moved | West Pond Gas Analyzer CH4 | Removed from Bouldin Alfalfa for calibration. Replaced by sn TG1-0211. Later we discovered that the 7700 at WP was behaving weirdly so we swapped this instrument in there are calibrated it there. |
TE525MM 65548-615 | 2020-11-19 | Bouldin Corn Precipitation | Bouldin Corn Precipitation | Plugged tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket which should result in 100 tips. Got 116 tips. Did not adjust - will retest with better water metering system. | |
CS547A 5688 | 2014-06-13 | Lab none | Installed | Mayberry Conductivity Sensor A | Installed at about 10cm below the water surface at Mayberry with CR10x, home-made interface sn 547-01 and 5 level TC profile. |
GP-001 GP20150111 | 2020-11-12 | Sherman Wetland-Temp1 CO2 Probe | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
CS450 20010770 | 2020-11-12 | Sherman Wetland-Temp1 Water Level | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
HMP60 G4540021 | 2020-11-12 | Sherman Wetland-Temp1 RH & Temp | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
PAR-LITE 050597 | 2020-11-12 | Sherman Wetland-Temp1 PAR Sensor Incoming | Notes | Sherman Wetland-Temp1 PAR Sensor Incoming | Tried to remove from boom with decommissioning of Sherman Wetland Temp tower, but it was stuck. Need vice grips and/or drill. |
PAR-LITE 050596 | 2020-11-12 | Sherman Wetland-Temp1 PAR Sensor Reflected | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
CNR1 990181 | 2020-11-12 | Sherman Wetland-Temp1 Net Radiometer 4-way | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
SRS-Ni NDVI 2046503402 | 2020-11-12 | Sherman Wetland-Temp1 Narrow Band NDVI Reflected | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
SRS-Ni NDVI 2046503368 | 2020-11-12 | Sherman Wetland-Temp1 Narrow Band NDVI Incoming | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
HFP01 10809 | 2020-11-12 | Sherman Wetland-Temp1 Heat Flux Plate B | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
HFP01SC 00245 | 2020-11-12 | Sherman Wetland-Temp1 Heat Flux Plate A | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
AM25T 2856 | 2020-11-12 | Sherman Wetland-Temp1 Multiplexer Thermocouple | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
AM16/32A 1164 | 2020-11-12 | Sherman Wetland-Temp1 Multiplexer Met | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
CR1000 50889 | 2020-11-12 | Sherman Wetland-Temp1 Data Logger Met | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
PTB110 H0160093 | 2020-11-12 | Sherman Wetland-Temp1 Barometer | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
LI-7550 AIU-0650 | 2020-11-12 | Sherman Wetland-Temp1 Interface LI-7550 | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
LI-7500A 75H-2180 | 2020-11-12 | Sherman Wetland-Temp1 Gas Analyzer CO2/H2O | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
LI-7700 TG1-0211 | 2020-11-12 | Sherman Wetland-Temp1 Gas Analyzer CH4 | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
WM 1590 153803 | 2020-11-12 | Sherman Wetland-Temp1 Anemometer | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
CR1000 4190 | 2020-11-12 | Sherman Wetland | Removed | Lab none | Removed from SW and brought back to the lab because the comparison has ended. This datalogger will be moved to WP next. |
HMP155 R0360206 | 2020-11-12 | Sherman Wetland | Removed | Lab none | Removed from SW and brought back to the lab because the comparison has ended. This reference sensor will be moved to WP next. |
TE525MM 65552-615 | 2020-11-12 | Sherman Wetland Precipitation | Tested | Sherman Wetland Precipitation | Dripped 475ml of water through the bucket which should result in 100tips. We got 92 tips. We turned its set screws two full rotations CW. We tested it again with another 475ml of water and got 96 tips. We left it as is. This needs to be tested again. One tipper also seems faster than the other. |
LI-7700 TG1-0211 | 2020-11-17 | Lab | Calibration | Lab | Calibration Location: Lab Last calibration: 2020-05-22 RSSI=72%, Optical RH=22% Ambient P=100.2kPa, Tair=20.5C Zero Air at 575psi, AMP CA4396 1.9497 at 400 CH4 Zero Offset: -.064ppm CH4 Span Offset: .011ppm Notes: Swapped optics drierite after calibration and flushed headspace with zero air. New optics RH was 11%. |
LI-7700 TG1-0223 | 2020-11-12 | Mayberry Gas Analyzer CH4 | Calibration | Mayberry Gas Analyzer CH4 | Calibration Location: Mayberry
Last calibration: 2020-05-22 RSSI=78%, Optical RH=26% Ambient P=101.6kPa, Tair=11.6C Zero Air at 700psi, AMP CA3923 3.121 at 700 CH4 Zero Offset: 0.011ppm CH4 Span Offset: 0.002ppm Notes: Still day. Swapped out pink silica gel and put in fresh drierite in the optics chem bottle. Used zero air to flush head space down to 9% RH. Optics RH was 12% when we left the site. |
LI-7700 TG1-0424 | 2020-11-12 | Sherman Wetland Gas Analyzer CH4 | Calibration | Sherman Wetland Gas Analyzer CH4 | Calibration Location: Sherman Wetland
Last calibration: 2020-05-21 RSSI=82%, Optical RH=12% Ambient P=101.8kPa, Tair=14.7C Zero Air at 700psi, AMP CA3923 3.121 at 700 CH4 Zero Offset: -0.084ppm CH4 Span Offset: 0.002ppm Notes: Still day |
CS547A 7157 | 2020-11-03 | Sherman Wetland Conductivity Sensor | Moved | Mayberry none | Moved from SW to MB for a temporary intercomparison because there was no water at the SW tower and sn 5688 at MB was giving high temp number |
PAR-LITE 060831 | 2020-01-10 | Factory none | Calibration | Factory none | y=mx+b 131.80 0 Calibrated by Chad Hanson at OSU as part of the Ameriflux standard program 7.59 uV/umol/m2/s |
PQS 1 161073 | 2020-11-03 | Lab none | Returned | Returned none | Returned to Stephen Chan at Ameriflux in exchange for a freshly calibrated PAR sensor. |
RPI0W RPI0W-004 | 2020-10-22 | Lab none | Installed | Sherman Wetland-Temp1 Camera | Installed at Sherman Wetland Temp Tower |
2300-000 2300565 | 2020-10-22 | Mayberry Fluorometer | Maintenance | Mayberry Fluorometer | Replaced all 3 brushes. One brush had splayed needles that blocked one of the lenses (the lens with the visible red light) |
PA-II-SD 84f3ebd5ffb1 | 2020-10-22 | Lab none | Installed | Bouldin Corn Air Quality Sensor | Installed at BC to replace sn 68c63acc07a9, which was brought back to the lab. |
PA-II-SD 68c63acc07a9 | 2020-10-22 | Bouldin Corn Air Quality Sensor | Removed | Lab none | Removed from BC and brought back to the lab. The microSD card is hard to seat properly into its slot. It was replaced by sn 84f3ebd5ffb1. |
PAR-LITE 990031 | 2020-10-26 | Lab none | Repair | Lab none | The cable insulation on this sensor was rotted. I cut the cable off near the sensor and spliced on a new cable. I also added an aluminum bracket to provide strain relief and used liquid electrical tape to seal the hole where the cable enters the sensor. |
PAR-LITE 990025 | 2020-10-26 | Lab none | Repair | Lab none | The cable insulation on this sensor was rotted. I cut the cable off near the sensor and spliced on a new cable. I also added an aluminum bracket to provide strain relief and used liquid electrical tape to seal the hole where the cable enters the sensor. The white wire (+) had a nick in its insulation right at the point where it entered the sensor that I covered with heat shrink. |
SRS-Nr NDVI 943203017 | 2020-10-22 | Tonzi | Removed | Lab none | Removed from the Tonzi floor met because it was no longer being used - tree NDVI pointing to the sky doesn't work |
eosFD FD20180048 | 2020-10-22 | Lab none | Installed | Tonzi-SoilCO2 Forced Diffusion Tree 1 | Installed at the under tree location at Tonzi to replace sn FD20180068 which has been reading zero CO2 concentration. |
eosFD FD20180068 | 2020-10-22 | Tonzi-SoilCO2 Forced Diffusion Tree 1 | Removed | Lab none | Removed from the under tree location at Tonzi because its co2 sensor has been reading zero. Replaced by sn FD20180048 |
Line Quantum LQ1570 | 2001-05-07 | Vaira | Moved | Tonzi none | Moved from Vaira to Tonzi |
BB-Fab BB-TF | 2003-12-19 | Tonzi-Floor Battery Bank | Upgrade | Tonzi-Floor Battery Bank | Removed four batteries from the floor system and install eoght new Excide NC27 95Ahr batteries. Treated all terminals with antiseize compound. The four batteries removed will be used at Vaira. |
BB-Fab BB-VR | 2003-12-19 | Vaira Battery Bank Main | Upgrade | Vaira Battery Bank Main | Cleaned battery terminals and added four batteries from the Floor site for a total of 12 at Vaira. |
LI-7500A 0035 | 2003-10-02 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0073 for calibration. |
LI-7500A 0073 | 2003-10-02 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0035. |
LI-7500A 0418 | 2003-10-02 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0041 for calibration. |
LI-7500A 0041 | 2003-10-02 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0418. |
LI-7500A 0042 | 2003-10-02 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0065 for calibration. |
LI-7500A 0065 | 2003-10-02 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0042. |
RPI0W RPI0W-000 | 2020-10-20 | Lab none | Tested | Lab none | Tested in the lab using spare parts from sn RPI0W-004. The board overall seems fine, but the board cannot connect to the camera, so the board is not going to be useful to run cameras in the future. The camera and camera cable seem good, so we have a spare camera and camera cable here. Running updates did not seem to change anything. I also changed the clock +7 min. |
RPI0W RPI0W-000 | 2020-10-05 | Sherman Wetland-Temp1 Camera | Removed | Lab none | Removed from SW Temp Tower and brought back to the lab because the camera was not connecting. |
RPI0W RPI0W-000 | 2020-01-30 | East Pond Camera | Moved | Sherman Wetland-Temp1 Camera | Removed with the decommissioning of the EP tower and installed at the new SW Temporary site 1250m upwind of the Permanent Tower. |
LI-7500A 0042 | 2003-06-20 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0065 for calibration. |
LI-7500A 0065 | 2003-06-20 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0042. |
LI-7500A 0035 | 2003-06-20 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0041 for calibration. |
LI-7500A 0041 | 2003-06-20 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0035. |
LI-7500A 0418 | 2003-06-20 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0073 for calibration. |
LI-7500A 0073 | 2003-06-20 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0418. |
WMP 1352 000024 | 2003-05-30 | Lab none | Installed | Tonzi-Floor Anemometer Floor | Installed on Tonzi-Floor to replace sn 000019 that was having problems - not sending data. |
WMP 1352 000019 | 2003-05-30 | Tonzi-Floor Anemometer Floor | Removed | Lab none | Removed from Tonzi-Floor due to no data. On power up it would send some data and then quit. |
LI-7500A 0073 | 2003-05-30 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira due to bad power connector on the box. |
LI-7500A 0065 | 2003-05-30 | Tonzi-Floor Gas Analyzer CO2/H2O | Moved | Vaira Gas Analyzer CO2/H2O | Moved from Tonzi-Floor to Vaira because the Floor sonic was dead and the Vaira 7500 box had a bad power connector |
LI-7500A 0073 | 2003-05-09 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0042 for calibration. |
LI-7500A 0042 | 2003-05-09 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0073. |
LI-7500A 0418 | 2003-05-09 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0065 for calibration. |
LI-7500A 0065 | 2003-05-09 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0418. |
LI-7500A 0035 | 2003-05-09 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0041 for calibration. |
LI-7500A 0041 | 2003-05-09 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0035. |
LI-7500A 0042 | 2003-04-06 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0418 for calibration. |
LI-7500A 0418 | 2003-04-06 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0042. |
LI-7500A 0065 | 2003-03-17 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0073 for calibration. |
LI-7500A 0073 | 2003-03-17 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0065. |
LI-7500A 0041 | 2003-03-17 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0035 for calibration. |
LI-7500A 0035 | 2003-03-17 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0041. |
BB-Fab BB-TT | 2003-02-13 | Tonzi-Tower Battery Bank | Upgrade | Tonzi-Tower Battery Bank | Four existing batteries removed for use at Vaira. 12 new batteries from Ione Autoparts installed |
BB-Fab BB-VR | 2003-02-13 | Vaira Battery Bank Main | Upgrade | Vaira Battery Bank Main | Added four batteries removed from Tonzi-Tower. Now a total of eight batteries. |
HMP45AC W0510026 | 2003-01-21 | Lab none | Installed | Tonzi none | Installed at Tonzi to replace sn U3030032 for calibration. |
HMP45AC U3030032 | 2003-01-21 | Tonzi none | Removed | Lab none | Removed from Tonzi. Replaced by sn W0510026 |
WMP 1352 000019 | 2003-01-07 | Lab none | Installed | Tonzi-Floor Anemometer Floor | Installed at Tonzi-Floor to replace sn 000021. |
WMP 1352 000021 | 2003-01-07 | Tonzi-Floor Anemometer Floor | Removed | Lab none | Removed from Tonzi-Floor. Replaced by sn 000019. |
BB-Fab BB-TF | 2001-01-03 | Lab none | Installed | Tonzi-Floor Battery Bank | Initial installation with four solar panels and three deep cycle batteries. |
LI-7500A 0042 | 2000-10-12 | Lab none | Calibration | Lab none | Zero with N2 was 2.0, after running several weeks in the lab.Cal gas 402.25 read 411.1 ppm at P=101.31 kPa and T 23.17 C. Licor Dewpoint sensor read 14.57 Tdew. Vaisala standard read 13.9 C. We are using the Vaisala as the standard for the experiment. Rezeroed and spanned D/A #1, H2O mmol m-3 0v=0 5 v=1000 D/A #2, CO2 mmol m-3 0v=13 mmol m-3 5v=20 mmol m-3 |
ML2x 54/020 | 2000-10-12 | Lab none | Calibration | Lab none | Water: 1.1367V vs air (zero): -0.00101V |
ML2x 54/019 | 2000-10-12 | Lab none | Calibration | Lab none | Water: 1.1317V vs air (zero): -0.000034V |
ML2x 54/017 | 2000-10-12 | Lab none | Calibration | Lab none | Water: 1.1294V vs air (zero): -0.00101V |
ML2x 54/016 | 2000-10-12 | Lab none | Calibration | Lab none | Water: 1.1403V vs air (zero): -0.00067V |
ML2x 54/015 | 2000-10-12 | Lab none | Calibration | Lab none | Water: 1.1381V vs air (zero): 0.00318V |
LI-7500A 0041 | 2002-06-06 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0035. |
LI-7500A 0042 | 2002-12-23 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0035 for calibration. |
LI-7500A 0035 | 2002-12-23 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0042. |
LI-7500A 0041 | 2002-12-23 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0073 for calibration. |
LI-7500A 0073 | 2002-12-23 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0041. |
LI-7500A 0065 | 2002-12-23 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0418 for calibration. |
LI-7500A 0418 | 2002-12-23 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0065. |
LI-7500A 0042 | 2002-09-20 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0418 for calibration. |
LI-7500A 0418 | 2002-09-20 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0042. |
LI-7500A 0065 | 2002-09-20 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0073 for calibration. |
LI-7500A 0073 | 2002-09-20 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0065. |
LI-7500A 0041 | 2002-09-20 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0035 for calibration. |
LI-7500A 0035 | 2002-09-20 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0041. |