Equipment and Calibration History

View Equipment |

View history entry 4982 for Licor LI-7700 sn: TG1-0421

Date: 2020-11-19
From: West Pond -
To: West Pond -
Title: Calibration
Notes: Calibration Location: West Pond Last calibration: 2020-05-19 RSSI=80%, Optical RH=8% Ambient P=102.0kPa, Tair=17.0C Zero Air at 500psi, AMP CA3923 3.121 at 500 CH4 Zero Offset: 0.005ppm CH4 Span Offset: 0.002ppm Notes: Still day. This sensor was originally at BA before we moved it to WP to calibrate. The sensor usually at WP (sn TG1-0224) was removed and brought back to the lab because its serial number was not recognized in the 7700 or 7550 software.
Entered by:
Field notes