Equipment and Calibration History

View Equipment |

View history entry 3705 for Licor LI-7500A sn: 75H-2514

Date: 2018-09-26
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-04-18 AGC=100.1 SDM=0 DAC1=0 DAC2=0 Ambient P=100.3kPa, Tair=27.3C, Tdew=12.7C Zero Air at 1600psi, AMP Air 408.97 at 1200 Tdewspan=16.00 CO2 zero 0.906 -> .9045 Offset: -6.04ppm CO2 span 1.0106 -> 1.0036 Offset: 3.52ppm H2O zero 0.8368 -> .8370 Offset: 1.3ppm H2O span 1.016 -> 1.0165 Offset: -.01ppm Notes:
Entered by: